- Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 1
03002 R.M. Young Wind Sentry Set 03101 R.M. Young Wind Sentry Anemometer 03301 R.M. Young Wind Sentry Vane Revision: 9/13 Copyright © 1986-2013 Campbell Scientific, Inc. - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 2
- Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 3
") to be free from defects in materials and workmanship under normal use and service for twelve (12) months from date of shipment unless otherwise specified in the corresponding Campbell pricelist or product manual. Products not manufactured, but that are re-sold by Campbell, are warranted only - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 4
(435) 227-9000. After an application engineer determines the nature of the problem, an RMA number will be issued. Please write this number clearly on 's expense. Campbell Scientific reserves the right to refuse service on products that were exposed to contaminants that may cause health or safety - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 5
Anemometer 10 7.2.3 03301 Vane 11 7.3 Wiring 11 7.4 Programming 13 7.4.1 Wind Speed 13 7.4.2 Wind Direction 13 7.4.3 Wind Vector Processing Instruction 14 7.4.4 Example Programs 14 7.4.4.1 CR1000 Program 15 7.4.4.2 CR10X Program 15 7.4.5 Long Lead Lengths 17 8. Sensor Maintenance 17 i - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 6
Table of Contents 9. Troubleshooting 18 9.1 Wind Direction 18 9.2 Wind Speed 18 10. References 19 Appendices A. Wind Direction Sensor Orientation A-1 A.1 Determining True North and Sensor Orientation A-1 B. Wind Direction Measurement Theory B-1 B.1 BRHalf Instruction B-1 B.2 EX-DEL-SE (P4) - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 7
cable is Santoprene® rubber. This compound was chosen for its resistance to temperature extremes, moisture, and UV degradation. However, this jacket will support combustion in air. It is rated as slow burning when tested according to U.L. 94 H.B. and will pass FMVSS302. Local fire codes may preclude - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 8
R.M. Young Wind Sentry 3.1 Ships With List The 03002 Wind Sentry ships with: (1) 03002 Wind Sentry including 03102 anemometer 03302 vane crossarm band clamp (pn 4919) (1) 12" x 1.0" unthreaded aluminium pipe (pn 3659) (1) Allen wrench (pn 5201) The 03101 anemometer ships with: (1) 03101 anemometer - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 9
R.M. Young Wind Sentry 1. Install the cup wheel to the anemometer shaft using the Allen wrench provided with the sensor. 2. Mount a CM202, CM204, or CM206 crossarm to the tripod or tower. 3. Orient the crossarm North-South, with the CM220 mount or 17953 NU- RAIL on the north end. Appendix A contains - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 10
R.M. Young Wind Sentry Alumimum Pipe CM220 CM200-Series Crossarm FIGURE 4-2. 03002 mounted to CM200-Series Crossarm with CM220 Aluminum Pipe pn 17953 NU-RAIL CM200-Series Crossarm Cable Tie FIGURE 4-3. 03002 mounted to CM200-Series Crossarm with pn 17953 4 - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 11
R.M. Young Wind Sentry 4.2 Step 2 - Use SCWin Short Cut to Program Datalogger and Generate Wiring Diagram The simplest method for programming the datalogger to measure the 03002 is to use Campbell Scientific's SCWin Short Cut Program Generator. 1. Open Short Cut and click on New Program. 2. Select - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 12
R.M. Young Wind Sentry 3. Select 03002 Wind Speed & Direction Sensor and select the right arrow (in center of screen) to add it to the list of sensors to be measured then select Next. 4. Select Wind Vector for the output and then select Finish. 6 - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 13
These two sensors combined differ from the 03002 only by the absence of a junction box. The R.M. Young Instruction Manual includes additional information on the operating principles, installation, and maintenance of the sensor. Cable length for the Wind Sentry is specified when the sensor is - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 14
R.M. Young Wind Sentry The 03002's cable can terminate in: • Pigtails that connect directly to a Campbell Scientific datalogger (option -PT). • Connector that attaches to a prewired enclosure (option -PW). Refer to www.campbellsci.com/prewired-enclosures for more information. • Connector that - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 15
R.M. Young Wind Sentry Threshold: Transducer: Output: Output Frequency: Cup Wheel Diameter: Weight: 0.5 m s-1 (1.1 mph) Stationary coil, 1300 ohm nominal resistance AC sine wave signal induced by rotating magnet on cup wheel shaft 100 mV peak-to-peak at 60 rpm; 6 V peak-to-peak at 3600 rpm 1 cycle - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 16
R.M. Young Wind Sentry 7.2 Assembly and Mounting Tools required: • 5/64" Allen wrench • Allen wrench provided with sensor • 1/2" open end wrench • compass and declination angle for the site (see Appendix A) • small screw driver provided with datalogger • UV resistant cable ties • small pair of - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 17
R.M. Young Wind Sentry 03101 can also be mounted at the top of a tripod using the CM216 (see FIGURE 7-1). Mount the CM200-series crossarm to the tripod or tower. Screw the mounting pipe into the base of the 03101, and attach the sensor / mounting pipe to the crossarm. Route the sensor cable along - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 18
R.M. Young Wind Sentry TABLE 7-1. 03002-L Wiring Color Red Black Clear White Green Blue Description Wind Spd. Signal Wind Spd. Reference Shield Wind Dir. Reference Wind Dir. Signal Wind Dir. Excitation CR800 CR5000 CR3000 CR1000 Pulse SE Analog Excitation CR510 CR10(X) Pulse G G AG SE Analog - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 19
not need to read this section to use Short Cut. 7.4.1 Wind Speed Wind speed is measured with the Pulse Count instruction (PulseCount() in CRBasic and Instruction 3 in Edlog). Use the low level AC configuration. For dataloggers programmed with Edlog, specify configuration code 21 to output frequency - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 20
, 60 Hz, reverse excitation 2500 mV CR5000 CR3000 5000 mV, 60 Hz, reverse excitation 5000 mV 352 352 0 0 7.4.3 Wind Vector Processing Instruction The Wind Vector output instruction is used to process and store mean wind speed, unit vector mean wind direction, and standard deviation of the wind - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 21
R.M. Young Wind Sentry 7.4.4.1 CR1000 Program 'CR1000 'Declare Variables and Units Public Batt_Volt Public WS_ms Public WindDir Units Batt_Volt=Volts Units WS_ms=meters/second Units WindDir=Degrees 'Define Data Tables DataTable(Table1,True,-1) DataInterval(0,60,Min,10) WindVector (1,WS_ms,WindDir, - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 22
R.M. Young Wind Sentry ;If WS < starting threshold then set WS = 0 3: If (XF) (P89) 1: 2 X Loc [ WS_ms ] 2: 4 < 3: 0.21 F 4: 30 Then Do 4: Z=F x 10^n (P30) 1: 0 F 2: 0 n, Exponent of 10 3: 2 Z Loc [ WS_ms ] 5: End (P95) ;03002 or 03301 RM Young Wind Direction Sensor measurement - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 23
programmed with Edlog (and the CR200(X)), the EX-DEL-SE (P4) measurement instruction should be used. Enter a 2 in the P4 "Delay" parameter for a can be checked with a paper clip as described in the R.M. Young manual. The potentiometer has a life expectancy of fifty million revolutions. As it - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 24
R.M. Young Wind Sentry 9. Troubleshooting 9.1 Wind Direction Symptom: NAN, -9999, or no change in direction 1. Check that the sensor is wired to the excitation and single-ended channel specified by the measurement instruction. 2. Verify that the excitation voltage and range code are correct for the - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 25
of the American Association of State Climatologists: Height and Exposure Standards, for Sensors on Automated Weather Stations, vol. 9, No. 4. WMO, 1983: Guide to Meteorological Instruments and Methods of Observation, World Meteorological Organization, No. 8, 5th edition, Geneva, Switzerland. 19 - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 26
R.M. Young Wind Sentry 20 - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 27
number of degrees between True North and Magnetic North. The preferred method to obtain the magnetic declination for a specific site is to use a computer service offered by NOAA at www.ngdc.noaa.gov/geomag. Magnetic declination can also be obtained from a map or local airport. A general map showing - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 28
Appendix A. Wind Direction Sensor Orientation FIGURE A-1. Magnetic declination for the contiguous United States (2004) A-2 - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 29
Appendix A. Wind Direction Sensor Orientation FIGURE A-2. Declination angles east of True North are subtracted from 0 to get True North FIGURE A-3. Declination angles west of True North are added to 0 to get True North A-3 - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 30
Appendix A. Wind Direction Sensor Orientation A-4 - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 31
a different excitation voltage. See Section 13.5 in the CR10X manual, Section 4.3 in the CR1000 manual, or Section 3.5 in the CR3000 manual for more information on the bridge measurements. B.2 EX-DEL-SE (P4) Instruction Instruction 4 outputs a precise excitation voltage (Vx) and measures the voltage - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 32
degrees is 352 degrees / 2500 mV = 0.1408 for the CR10X, or, 352 degrees / 5000 mV = 0.0704 for the CR23X. See Section 13.5 in the datalogger manual from more information on the bridge measurements. B-2 - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 33
METEOROLOGICAL INSTRUMENTS INSTRUCTIONS WIND SENTRY MODEL 03002-5 R.M. YOUNG COMPANY 2801 AERO PARK DRIVE, TRAVERSE CITY, MICHIGAN 49686, USA TEL: (231) 946-3980 FAX: (231) 946-4772 WEB: www.youngusa.com P/N: 03002-5-90 REV: D110210 - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 34
wide temperature range high quality instrument oil. Standard bearings have light contacting seals to exclude contamination and help retain lubricant for longer service life. Cup wheel rotation produces an AC sine wave voltage signal with frequency directly proportional to wind speed. This AC signal - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 35
precision ball bearings and the azimuth potentiometer. Replacement of these components should only be performed by a qualified instrument technician. If service facilities are not available, return the instrument to the factory. Refer to the accompanying drawings to become familiar with part names - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 36
3. ALIGN VANE a) Connect excitation voltage and signal conditioning electronics to instrument according to wiring diagram. b) Loosen set screw in side of vane hub. c) Position instrument so crossarm is oriented north-south with vane on north side. Orient vane to a known angular reference. (See - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 37
CABLE & WIRING DIAGRAM 03002-5-90(D) Page 4 - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 38
GENERAL ASSEMBLY & REPLACEMENT PARTS Page 5 03002-5-90(D) - Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 39
- Campbell Scientific 03101 | 03002, 03101, and 03301 R. M. Young Wind Sentry Sens - Page 40