Fluke 115 Fluke 115 Digital Multimeter Datasheet

Fluke 115 Manual

Fluke 115 manual content summary:

  • Fluke 115 | Fluke 115 Digital Multimeter Datasheet - Page 1
    Fluke 115 Multimeter Actual size R N10140 Technical Data Compact true-rms meter for field service technicians The Fluke 115 is the solution for a wide variety of electrical and electronic testing applications. This true-rms meter provides easy one-handed operation in a compact package. Features
  • Fluke 115 | Fluke 115 Digital Multimeter Datasheet - Page 2
    dc coupled, 45 Hz to 5 kHz for ac current.  Fluke Corporation 115 Multimeter Ordering information Fluke-115 Multimeter Included TL75 Test leads, holster, User's manual and 9 V battery (installed). Fluke. Keeping your world up and running. Fluke Corporation PO Box 9090, Everett, WA USA 98206
  • 1
  • 2

Technical Data
Fluke 115
Multimeter
Features include:
Resistance, continuity, frequency, capacitance, and diode test
Measures 20 A (30 seconds momentary; 10 A continuous)
Large white LED backlight to work in poorly lit areas
Compact ergonomic design for one-handed operation
Compatible with optional magnetic hanger (ToolPak
)
True-rms for accurate measurements on non-linear loads
Min/Max/Average with elapsed time to record signal
fluctuations
CAT III 600 V safety rated
Compact true-rms meter for
field service technicians
The Fluke 115 is the solution for a wide
variety of electrical and electronic test-
ing applications. This true-rms meter
provides easy one-handed operation in
a compact package.
General specifications
Accuracy is specified for 1 year after calibration, at operating
temperatures of 18 °C to 28 °C, with relative humidity at
0 % to 95 %.
The accuracy specifications take the form of:
± ( [ % of reading ] + [ counts ] )
Maximum voltage between any
terminal and earth ground
600 V
Surge protection
6 kV peak per IEC 61010-1 600 V
CAT III, Pollution Degree 2
Fuse for A input
11 A, 1000 V FAST Fuse
(Fluke PN 803293)
Display
Digital: 6,000 counts, updates 4/sec
Bar graph
33 segments, updates 32/sec
Operating temperature
-10 °C to + 50 °C
Storage temperature
-40 °C to + 60 °C
Battery
9 volt Alkaline, NEDA 1604A/
IEC 6LR61
Battery life
400 hours typical, without backlight
N10140
R
Actual size