Fluke 1625 FE 1625 Users Manual - Page 24

RH with, Umeas = 48 V, RH with Umeas = 20 V, Resolution, electrode resistance R

Page 24 highlights

1625 Users Manual fluencing conditions (see diagramm), the display shows a warning symbol W and a notice that RS or RH are too high. F 1000% 100% 10% 1 % 3000 1000 RH/RE= 300 100 30 10 3 1 0,3 0 ,1% 1 0 100 1 k 10k 100k 1M RS edw005.eps Automatic switchover of measuring resolution in dependence to auxiliary earth electrode resistance RH: RH with Umeas = 48 V < 300 Ω < 6 kΩ < 60 kΩ < 600 kΩ RH with Umeas = 20 V < 250 Ω < 2,5 kΩ < 25 kΩ < 250 kΩ Resolution 1 mΩ 10 mΩ 100 mΩ 1 Ω 16

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90

1625
Users Manual
16
fluencing conditions (see diagramm), the display shows a warning symbol
W
and a notice that
RS or RH are too high.
%
0
0
0
1
F
%
0
0
1
%
0
1
%
1
%
1
,
0
M
1
k
0
0
1
k
0
1
k
1
0
0
1
0
1
3
,
0
1
3
0
1
0
3
0
0
1
=
R
/
R
E
H
0
0
3
0
0
0
1
0
0
0
3
R
S
edw005.eps
Automatic switchover of measuring resolution in dependence to auxiliary earth
electrode resistance R
H
:
RH with
Umeas = 48 V
RH with Umeas = 20 V
Resolution
< 300
< 250
1 m
<
6 k
<
2,5 k
10 m
<
60 k
<
25 k
100 m
< 600 k
< 250 k
1