Fluke 1625 FE 1625 Users Manual - Page 73

Principle of Operation, Purpose

Page 73 highlights

Dimensions Operating error*: Display range*: Measuring frequency: Measuring voltage: Measuring range*: Resolution: *when using: Stakeless Earth/Ground Resistance Testing Principle of Operation Total length of adaptercable 1,8 m, Length of shielded cable to clamp-on CT 1,5 m. ± (10% of m.v. + 5 digit). 0,010 Ω ... 130 Ω. 128 Hz. Um=48 VAC (primary). 0,020 Ω ... 100 Ω. .001... .1 ohm. CT EI-162AC to induce voltage and CT EI-162X to detect current. Recommended distance between CT's: 10 cm. Principle of Operation Purpose Testing the resistances of individual ground connections in systems with parallel ground connections (multi-grounded systems). 65

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90

Stakeless Earth/Ground Resistance Testing
Principle of Operation
65
Dimensions
Total length of adaptercable 1,8 m, Length
of shielded cable to clamp-on CT 1,5 m.
Operating error*:
± (10% of m.v. + 5 digit).
Display range*:
0,010
... 130
.
Measuring frequency:
128 Hz.
Measuring voltage:
U
m
=48 VAC (primary).
Measuring range*:
0,020
... 100
.
Resolution:
.001... .1 ohm.
*when using:
CT EI-162AC to induce voltage and
CT EI-162X to detect current.
Recommended distance between CT's:
10 cm.
Principle of Operation
Purpose
Testing the resistances of individual ground connections in systems with paral-
lel ground connections (multi-grounded systems).