Fluke 51 Fluke 51II, 52II, 53II, and 54II Thermometer Datasheet

Fluke 51 Manual

Fluke 51 manual content summary:

  • Fluke 51 | Fluke 51II, 52II, 53II, and 54II Thermometer Datasheet - Page 1
    50 Series II Thermometers Fluke 54 II Laboratory accuracy. Wherever you go. The Fluke 50 Series II contact thermometers offer fast response and laboratory accuracy (0.05% + 0.3°C) in a rugged handheld test tool. • Large backlit dual display shows any combination of T1, T2 (52 and 54 only), T1-T2
  • 1

C25
80PK-26
80PK-25
FVF-SC1
TPAK
50 Series II Thermometers
Laboratory accuracy. Wherever you go.
Additional features for the 53 and
54 Series II:
Data Logging up to 500 points of data
with user adjustable recording interval
Real time clock captures the exact time of
day when events occur
Recall function allows logged data to be
easily reviewed on the meter display
IR communication port allows data to be
exported to optional FlukeView
®
Temperature PC software
Features
Specifications
Recommended Accessories
Battery life:
1000 hours typical, AA
Size (HxWxD):
173 x 86 x 38 mm
Weight:
0.4 kg
Three Year Warranty
11080-eng
Rev. 01
Included Accessories
Impact absorbing holster
Two bead probe thermocouples 80PK-1
(54+52)
One bead probe thermocouple 80PK-1
(51+53)
Ordering Information
Fluke 51 II
Thermometer
Fluke 52 II
Thermometer
Fluke 53 II
Thermometer
Fluke 54 II
Thermometer
FVF-SC1
FlukeView Forms
Software including
interface cable
51 II
52 II
53 II
54 II
Thermocouple types
J,K,T,E
J,K,T,E
J,K,T,E,N,R,S
J,K,T,E,N,R,S
Number of inputs
Single
Dual
Single
Dual
Time stamp
Relative Time
Relative Time
Time of Day
Time of Day
Splash/Dust resistant
Dual display with backlight
Min/Max/Avg recording
(T
1
-T
2
) True differential
Data logging up to 500 points
IR data port for interface
to PC
Compatible with optional
FlukeView Software
Temperature range:
J-type
Thermocouples
-210°C to 1200°C
(-346°F to 2192°F)
K-type
Thermocouples
-200°C to 1372°C
(-328°F to 2501°F)
T-type
Thermocouples
-250°C to 400°C
(-418°F to 752°F)
E-type
Thermocouples
-150°C to 1000°C
(-238°F to 1832°F)
N-type**
Thermocouples
-200°C to 1300°C
(-328°F to 2372°F)
R** and S-type** Thermocouples
0°C to 1767°C
(32°F to 3212°F)
Temperature accuracy
Above –100°C (-148°F):
J, K, T, E, and N-type**
± [ 0.05% + 0.3°C (0.5°F) ]
R** and S-type**
± [ 0.05% + 0.4°C (0.7°F) ]
Below –100°C (-148°F):
J, K, E, and N-types
± [ 0.20% + 0.3°C (0.5°F) ]
T-type
± [ 0.50% + 0.3°C (0.5°F) ]
**Only the Fluke Models 53 and 54 Series II thermometers are capable of measuring N, R, and S-type thermocouples.
The Fluke 50 Series II contact
thermometers offer fast response and
laboratory accuracy (0.05% + 0.3°C) in
a rugged handheld test tool.
Large backlit dual display shows any
combination of T
1
, T
2
(52 and 54 only),
T
1
-T
2
(52 and 54 only) plus MIN, MAX,
or AVG
Relative time clock on MIN, MAX, and
AVG provides a time reference for major
events
Electronic Offset function allows
compensation of thermocouple errors to
maximize overall accuracy
Readout in °C, °F, or Kelvin (K)
Sleep mode increases battery life
Battery door allows easy battery
replacement without breaking the
calibration seal
Fluke 51 II
Fluke 52 II
Fluke 53 II
Fluke 54 II