Fluke 6531-40M Product Manual

Fluke 6531-40M Manual

Fluke 6531-40M manual content summary:

  • Fluke 6531-40M | Product Manual - Page 1
    off valve  Variable volume (screw press)  Variable volume piston position indicator Figure 1: E-DWT-H with Hi and Lo Q-RPT Introduction This demo guide is to assist customers and representatives to perform an effective demonstration or evaluation of E-DWT-H. It is written for the more complex
  • Fluke 6531-40M | Product Manual - Page 2
    due to fluid head. For more precise instruments with lower ranges, you may need to zero the DUT before making final test connections. AutoTest will guide you to set each pressure that was defined in Running an E-DWT-H Quick Test. Note. Use the variable volume to control pressures. For more precise
  • Fluke 6531-40M | Product Manual - Page 3
    Once you have set a test pressure, E-DWT-H will automatically test for stability, and indicate to the operator when it is safe to take data. Press [ENT] to collect a data point. Hint: Try collecting a data point while the pressure is still unstable and you have not yet seen the stability light
  • 1
  • 2
  • 3

PN 3469593
April 2014
© 2014 Fluke Corporation. All rights reserved. Specifications are subject to change without notice.
E-DWT-H User Guide
TEST port (DH500F)
Reservoir cap
Priming pump plunger
Reservoir cap vent valve
Reference pressure monitor (RPM4-E-DWT)
Test shut off valve and fine pressure adjustment
‘Lo Q-RPT shut off’ valve position prompt LED
Lo Q-RPT shut off valve
‘Lo Q-RPT active’ valve position prompt LED
‘Lo Q-RPT active’ caution indicator
Reservoir shut off valve
Variable volume (screw press)
Variable volume piston position indicator
Figure 1:
E-DWT-H with Hi and Lo Q-RPT
Introduction
This demo guide is to assist customers and
representatives
to
perform
an
effective
demonstration or evaluation of E-DWT-H.
It is
written for the more complex case of a dual Q-RPT
E-DWT-H model.
Preparation
Plug in the E-DWT-H to power up.
Attach the four variable volume handles.
Open the reservoir cap (vent port) by turning
counter-clockwise.
Open the reservoir shutoff valve.
Back out the variable volume by turning
counter-clockwise.
Back it out all the way,
then turn back in (clockwise)
½
turn.
Set the Lo Q-RPT shutoff valve to the Hi
Q-RPT (clockwise).
Set the RPM4 to the Hi Q-RPT using the
[7/Range] key.
Open the test shutoff valve by turning counter-
clockwise until it is fully backed out.
Turn the
handle back in (clockwise) 4 turns to bring the
valve close to mid-stroke.