Fluke 717-500G Fluke 717 and 718 Pressure Calibrator Datasheet

Fluke 717-500G Manual

Fluke 717-500G manual content summary:

  • Fluke 717-500G | Fluke 717 and 718 Pressure Calibrator Datasheet - Page 1
    992 g (35 oz) Power: 9V battery ANSI/NEDA 1604A or IEC 6LR619V alkaline; two batteries in 718 Battery Life: 4 to 20 hours typical, depending on functions used Warranty: Three years (one year on pressure pump in Fluke 718) Display: LCD, 5 digit pressure and current simultaneous Accuracy: 0.05 % Note
  • Fluke 717-500G | Fluke 717 and 718 Pressure Calibrator Datasheet - Page 2
    o-rings seat properly. The Calibrator is now ready for use. Functional Pressure Specifications Model Range 718 1G -1 PSI to +1 PSI, -68.9 mbar to 68.9 mbar (-6.89 kPa to 6.89 kPa) 718 30G -12 PSI to 30 PSI, (-850 mbar to 2 bar, -85 to 206.84 kPa) 718 100G -12 PSI to 100 PSI, (-850 mbar to
  • Fluke 717-500G | Fluke 717 and 718 Pressure Calibrator Datasheet - Page 3
    Per pressure module specs Accuracy: To 0.025 % of full span, per pressure module specs Functions: Zero, Min, Max, Hold, Damp Note: media compatibility per pressure module specs Fluke 718 series via built in pump Range: Loop Power: -12 PSI or -850 mbar to full scale, Supported Pressure Units
  • 1
  • 2
  • 3

Fluke Pressure
Calibrators
Technical Data
Fluke 718 Pressure Calibrator
The complete solution to pressure calibration
The Fluke 718 pressure calibrator family (4 models)
provide a total pressure calibration solution for
transmitters, gauges and switches. This compact
pressure calibration solution is about a third of the
size of comparable instruments and weighs just one
kilogram (two pounds). The unique, easy to clean
out access ports built into the pump protect the
pump from fluids and provides the ability to service
the pump without disassembly.
Pressure Source and Milliamp Measurement to
calibrate and maintain almost any pressure device
Integrated pump is easily cleaned when
accidentally exposed to fluids:
– reduces cost of ownership and repairs
– service the pump in the field
Error % calculation for quicker pass/fail decision
making in the field
Best in class mA accuracy of 0.015 % for better
measurement confidence and workload coverage
Min/Max/Hold to capture changing measurements
Switch test feature for quicker and more accurate
pressure switch testing
New 1 and 300 PSI pressure ranges mean fewer
extra tools required
1, 30, 100 & 300 PSI ranges now available
General Specifications
Maximum voltage:
30 volts: Non-operating
Temperature:
-40 °C to 60 °C
Operating temperature:
-10 °C to 55° C
Relative humidity:
95 % (10 °C to 30 °C); 75 % (30 °C to 40 °C);
45 % (40 °C to 50 °C); 35 % (50 °C to 55 °C)
Operating altitude:
3,000 m max
Shock:
1 m drop test
Vibration:
Random, 2 g, 5 Hz – 500 Hz
Safety:
CSA C22.2 No. 1010.1: 1992
EMC:
EN50082-1: 1992 and EN55022: 1994 Class B
Size/weight (717 with holster):
201 mm x 98 mm x 52 mm
(7.93 in x 3.86 in x 2.06 in) 600 g (21 oz)
Size/weight (718 with holster):
216 mm x 94 mm x 66 mm
(8.50 in x 3.72 in x 2.60 in) 992 g (35 oz)
Power:
9V battery ANSI/NEDA 1604A or IEC 6LR619V alkaline; two
batteries in 718
Battery Life:
4 to 20 hours typical, depending on functions used
Warranty:
Three years (one year on pressure pump in Fluke 718)
Display:
LCD, 5 digit pressure and current simultaneous
Accuracy:
0.05 %
Note:
For use with non-corrosive gasses