Fluke 9100S-D-156 Product Manual

Fluke 9100S-D-156 Manual

Fluke 9100S-D-156 manual content summary:

  • Fluke 9100S-D-156 | Product Manual - Page 1
    9100S Dry-Well Calibrator Users Guide 9100S___ugeng0000 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 2
    Limited Warranty & Limitation of Liability Each product from Fluke Corporation, ("Fluke") is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is one year for Dry-Well Calibrators. The warranty period begins on the date of the shipment.
  • Fluke 9100S-D-156 | Product Manual - Page 3
    Safety Information 3 1.3.1 Warnings...3 1.3.2 Cautions...4 1.4 Authorized Service Centers 5 2 Specifications and Environmental Conditions 9 2.1 Specifications 5.2 Changing Display Units 17 6 Controller Operation 19 6.1 Well Temperature 19 6.2 Temperature Set-point 19 6.2.1 Programmable Set-
  • Fluke 9100S-D-156 | Product Manual - Page 4
    9100S Dry-Well Calibrator  6.3.1 Scan Control 22 6.3.2 Scan Rate...22 6.4 Set-point Resistance 22 6.5 Secondary Menu 23 6.6 Heater Power 23 6.7 Proportional Band 23 6.8 Controller Configuration 24 6.9 Operating Parameters 25 6.
  • Fluke 9100S-D-156 | Product Manual - Page 5
      11 Troubleshooting 43 11.1 Troubleshooting Problems, Possible Causes, and Solutions 43 11.2 CE Comments 44 11.2.1 EMC Directive 44 11.2.2 Low Voltage Directive (Safety 44 iii MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 6
    9100S Dry-Well Calibrator  Tables Table 1 International Electrical Symbols 1 Table 2 Controller communications commands 32 iv MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 7
    Figure 1 Back Panel 13 Figure 2 Side view of the 9100S showing the prop stand 14 Figure 3 Front panel 15 Figure 4 Fixed block options 16 Figure 5 Controller operation flow chart 21 Figure 6 Serial cable wiring 31 v MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 8
    MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 9
    , to any desired temperature within the instrument's range. Multiple fault protection devices insure user and instrument safety and protection. This dry-well calibrator was designed for portability, low cost, and ease of operation. Through proper use, the instrument will provide continuous accurate
  • Fluke 9100S-D-156 | Product Manual - Page 10
    9100S Dry-Well Calibrator Symbols Used Symbol Description AC-DC Battery Complies with European Union directives DC Double Insulated Electric Shock Fuse PE Ground Hot Surface (Burn Hazard) Read the User's Guide . 2 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 11
    probes. ●● If the instrument is used in a manner not in accordance with the equipment design, the operation of the dry-well may be impaired or safety hazards may arise. ●● This instrument is intended for indoor use only. 3 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 12
    9100S Dry-Well Calibrator Safety Information BURN HAZARDS ●● DO NOT turn the instrument upside down with the inserts in place; the 122°F (5°C to 50°C). Allow sufficient air circulation by leaving at least 6 inches (15 4 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 13
    the calibration parameters. ●● DO NOT operate this instrument in an excessively wet, oily, dusty, or dirty environment. Always keep the well and inserts clean and clear of foreign material. ●● The dry-well is a precision instrument. Although it has been designed for optimum durability and trouble
  • Fluke 9100S-D-156 | Product Manual - Page 14
    9100S Dry-Well Calibrator Authorized Service Centers 6 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 15
    You Start Authorized Service Centers E-mail: [email protected] When contacting these Service Centers for support, please have the following information available: ●● Model Number ●● Serial Number ●● Voltage ●● Complete description of the problem 7 MyFlukeStore Shop for Fluke products online at
  • Fluke 9100S-D-156 | Product Manual - Page 16
    MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 17
    at equal depths within wells 35 to 375°C: 9.5 SERVICEABLE PARTS 2.2 Environmental Conditions Although the instrument has been designed for optimum durability and trouble calibration environment should be minimized ●● altitudes less than 2000 meters ●● indoor use only 9 MyFlukeStore Shop for Fluke
  • Fluke 9100S-D-156 | Product Manual - Page 18
    MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 19
    9100S Dry-well ●● Power Cord ●● User's Guide with Report of Calibration ●● RS-232 Cable ●● 9930 Interface-it Software and User's Guide 3.2 Set-up Place the calibrator the calibrator. Turn on the power to the calibrator by the temperature set-point on the calibrator using the front panel keys. The
  • Fluke 9100S-D-156 | Product Manual - Page 20
    temperature. The well may require 5 to 10 minutes to reach the set-point depending on the span. Another 5 to 10 minutes is required to stabilize within ±0.1°C of the set-point. Ultimate stability may take 15 to 20 minutes more of stabilization time. MyFlukeStore Shop for Fluke products online at
  • Fluke 9100S-D-156 | Product Manual - Page 21
    4 Parts and Controls The user should become familiar with the dry-well calibrator and its parts: (See Figures 1, 2, and 3). 4.1 Rear the corners of the instrument are provided for airflow. The area around the calibrator must be kept clear to allow adequate ventilation. The air is directed from
  • Fluke 9100S-D-156 | Product Manual - Page 22
    9100S Dry-Well Calibrator Side View RS-232 - The RS-232 serial port provides a means for connecting the instrument to a computer or a printer using the included serial cable. WARNING: Always leave enough clearance in front of the calibrator 2 Side view of the 9100S showing the prop stand CAUTION:
  • Fluke 9100S-D-156 | Product Manual - Page 23
    panel is the well opening where a and actual temperatures but also various calibrator functions, settings, and constants. calibrator temperature set-point, access and set other operating parameters, and access and set calibration the sensor that is to be calibrated is inserted. A high-temperature
  • Fluke 9100S-D-156 | Product Manual - Page 24
    9100S Dry-Well Calibrator Constant Temperature Block Assembly the temperature of the block. The entire 1/4" (6.4 mm) 1/8" (3.2 mm) Figure 4 Fixed block options Block "D" 3 mm 4 mm 6 mm 3/16" (4.8 mm) 4 mm 6 mm 3 mm 16 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 25
    set the temperature set-point on the calibrator using the front panel keys. The procedure well heater on or off to raise or lower the temperature. The displayed well temperature gradually changes until it reaches the set-point temperature. The well Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 26
    MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 27
    dry-well temperature controller using the front control panel. Using the front panel key-switches and LED display the user may monitor the well conveniently set the calibrator to a previously Well temperature in degrees Celsius Access set-point memory 19 MyFlukeStore Shop for Fluke products online
  • Fluke 9100S-D-156 | Product Manual - Page 28
    so that when the set-point is changed the dry-well heats or cools at a specified rate (degrees per minute) until it reaches the new set-point. With the scan disabled the dry-well heats or cools at the maximum possible rate. MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 29
    Adjust R0 SET/EXIT Select ALPHA DO NOT CHANGE THESE VALUES Adjust ALPHA SET/EXIT Select DELTA Adjust DELTA SET/EXIT 21 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 30
    9100S Dry-Well Calibrator Set-point Resistance 6.3.1 Scan Control The scan is controlled with the scan on/off function that appears in and then the whole number of the current set-point resistance setting is displayed 22 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 31
    6.6 Heater Power The temperature controller controls the temperature of the well by pulsing the heater on and off. The total power knowing the amount of heating, the user can tell if the calibrator is heating up to the set-point, cooling down, or Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 32
    9100S Dry-Well Calibrator Controller Configuration to lower the temperature back down. In this way the temperature is maintained at a fairly constant level. The temperature stability of the well and response time depend on the width of the proportional band. If the band is too wide, the well
  • Fluke 9100S-D-156 | Product Manual - Page 33
    Controller Operation Serial Interface Parameters parameters, serial interface parameters, and calibration parameters. The menus are selected using the up and down arrows and then . The baud rate parameter is indicated by, 25 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 34
    9100S Dry-Well Calibrator Serial Interface Parameters bAUd Flashes bAUd for one second and then displays the setting 2400b Current baud rate mode setting The mode may be changed using the up or down arrows (u d). 26 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 35
    to the factory settings. The user should have a list of these constants and their settings with the instrument manual. CAUTION: DO NOT change the values of the instrument calibration constants from the factory set values. The correct setting of these parameters is important to the safety, proper
  • Fluke 9100S-D-156 | Product Manual - Page 36
    9100S Dry-Well Calibrator Calibration Parameters new value. Pressing "EXIT" causes the parameter to be skipped ignoring any changes that may have been and then the ALPHA setting is displayed 38530 Current ALPHA setting 28 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 37
    Controller Operation Calibration Parameters To change the ALPHA setting, press the up or down arrows. 38600 New ALPHA setting To accept "EXIT" to continue without storing the new value. S Accept the new DELTA setting 29 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 38
    MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 39
    a linefeed (ASCII decimal 10)), and the duplex to HALF, disabling echo. GND 5 9 GND 4 8 RxD TxD 3 7 TxD RxD 2 6 1 Figure 6 Serial cable wiring 31 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 40
    9100S Dry-Well Calibrator Interface Commands The serial port can be used to transmit measurements to a computer or printer or to change ON srat: 99.9 {C or F}/min srat:12.4 C/min ON or OFF 0.1 to 99.9°C 0.2 to 179.8°F 32 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 41
    0.006 Set DELTA calibration parameter de[lta]=n de=1.37 de: 9.99999 0-3.0 Miscellaneous Other Commands Read firmware version number *ver[sion] *ver ver.9999x,9.99 ver.9100S,1.01 Read structure of RETURNED column. 33 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 42
    MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 43
    well. Once the probe is inserted into the well well, it may be compared to the calibrator Dry-well Characteristics There is a temperature gradient vertically in the test well dry-well. However, actual heat losses vary with design of the thermometer probes inserted into the calibrator well calibration
  • Fluke 9100S-D-156 | Product Manual - Page 44
    MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 45
    most useful operating range of the dry-well. The farther apart the calibration temperatures, the greater the calibrated temperature range. However, the calibration error is also greater over the the following formula: 37 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 46
    9100S Dry-Well Calibrator Calibration Procedure 9.2.1 Compute DELTA A = T3 − T2 B = T2 − T1 C =  T3 100   1 − T3 100   −  T2 100   1 − T2 100   D =  T2 100  above (Section ) 38 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 47
    5 on page 21). 9.2.3 Accuracy and Repeatability Check the accuracy of the dry-well at various points over the calibration range. If drywell does not pass specification at all set-points, repeat the Calibration Procedure. 39 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 48
    MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 49
    Fluke, users should check with an Authorized Service Center (see Section 1.4, Authorized Service Centers on page 5) to be sure that the proposed method will not damage the equipment. ●● If the instrument is used in a manner not in accordance with the equipment design, the operation of the dry-well
  • Fluke 9100S-D-156 | Product Manual - Page 50
    MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9100S-D-156 | Product Manual - Page 51
    Troubleshooting Troubleshooting Problems, Possible Causes, and Solutions 11 Troubleshooting This section contains information on troubleshooting, CE Comments, and a wiring diagram. 11.1 Troubleshooting Problems of Calibration. Reprogram the parameters into the instrument (see Section , Calibration
  • Fluke 9100S-D-156 | Product Manual - Page 52
    11.2 CE Comments 11.2.1 EMC Directive Fluke's equipment has been tested to meet the European Voltage Directive (Safety) In order to comply with the European Low Voltage Directive (73/23/EEC), Fluke equipment has been designed to meet the IEC 1010-1 (EN 61010-1) and the IEC 1010-2010
  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52

9100S
Dry-Well Calibrator
Users Guide
9100S___ugeng0000
Shop for Fluke products online at:
1.888.610.7664
www.
MyFlukeStore
.com