HP Healthcare Edition HC271 Product Notice AMS

HP Healthcare Edition HC271 Manual

HP Healthcare Edition HC271 manual content summary:

  • HP Healthcare Edition HC271 | Product Notice AMS - Page 1
    Consult instructions for use. Alternating current Indications for use/ intended use • This display is intended for healthcare professionals mode HC271: available HC271p: not available IP Protection IP32 on the front glass Contacting HP Go to www.hp.com/support to locate your user guide, drivers
  • HP Healthcare Edition HC271 | Product Notice AMS - Page 2
    the front glass Contacting HP Go to www.hp.com/support to locate your user guide, drivers, and software. For safety, regulatory, and environmental information, refer to the Product Notices provided with your product. R Regulatory Notice HC271/HC271p Clinical Review Monitors HP Inc.,1501 Page Mill
  • 1
  • 2

Regulatory Notice
HC271/HC271p Clinical Review Monitors
HP Inc.,1501 Page Mill Road,
Palo Alto, CA 94304, United States
WARNING!
Electricity.
CAUTION:
Indicates a hazardous situation that, if
not avoided, could result in damage to equipment
or loss of information.
Consult instructions for use.
Alternating current
Indications for use/
intended use
• This display is intended for healthcare
professionals viewing patient information,
records, medical examination images, or results
in a hospital or clinical medical environment.
• This display is not intended for use in
interpreting or analyzing patient information,
records, medical examination images, or
results. The display is intended only to provide
information to assist healthcare professionals in
making an independent diagnosis or treatment
decision regarding an individual patient.
• Do not operate this display with any surgical,
life-support, or radiological equipment
(mammography equipment included).
• Do not use this display with radiology,
pathology, or mammography systems for
patient diagnosis purposes.
Connecting to external
devices
All the equipment intended to connect
this display shall be certified according to
international standards IEC60601-1, IEC60950
or other IEC/ISO Standards applicable to the
equipment.
All configurations of equipment must comply
with the system standard IEC60601-1.
Connecting any additional equipment to the
signal input port or signal output port of this
display is considered configuring a medical
system. Therefore, the system must comply
with the requirements of the system standard
IEC60601-1.
Disconnecting devices
A power cord is included with the display. If
another cord is used, use only a power source
and connection appropriate for this display. For
information on the correct power cord set to
use with the display, refer to the
Product Notices
provided in your documentation kit.
WARNING!
To reduce the risk of electric shock or
damage to the equipment:
• Plug the power cord into an AC outlet that is easily
accessible at all times.
• Disconnect power from the computer by unplugging
the power cord from the AC outlet.
• If provided with a 3-pin attachment plug on the power
cord, plug the cord into a grounded (earthed) 3-pin
outlet. Do not disable the power cord grounding
pin, for example, by attaching a 2-pin adapter. The
grounding pin is an important safety feature.
• Installations must be set up by qualified personnel
only.
• It is the responsibility of the operating institution
to choose qualified personnel for installation and
operation of the device.
Cleaning the display
1. Turn off the display.
2. Disconnect any external devices.
3. Wipe the exterior with a soft, water-
dampened cloth to remove any visible soiling.
4. Next, use any of the following germicidal
wipe formulations to safely disinfect all
exposed surfaces of your monitor. Refer
to the directions for use provided by the
manufacturer of the wipes.
CAUTION:
The cloth should be moist, but not wet.
Water dripping into the ventilation or other points
of entry can cause damage to the monitor. Allow the
unit to air-dry before use.
Regulatory Notice
HC271/HC271p Clinical Review Monitors
HP Inc.,1501 Page Mill Road,
Palo Alto, CA 94304, United States
WARNING!
Electricity.
CAUTION:
Indicates a hazardous situation that, if
not avoided, could result in damage to equipment
or loss of information.
Consult instructions for use.
Alternating current
Indications for use/
intended use
• This display is intended for healthcare
professionals viewing patient information,
records, medical examination images, or results
in a hospital or clinical medical environment.
• This display is not intended for use in
interpreting or analyzing patient information,
records, medical examination images, or
results. The display is intended only to provide
information to assist healthcare professionals in
making an independent diagnosis or treatment
decision regarding an individual patient.
• Do not operate this display with any surgical,
life-support, or radiological equipment
(mammography equipment included).
• Do not use this display with radiology,
pathology, or mammography systems for
patient diagnosis purposes.
Connecting to external
devices
All the equipment intended to connect
this display shall be certified according to
international standards IEC60601-1, IEC60950
or other IEC/ISO Standards applicable to the
equipment.
All configurations of equipment must comply
with the system standard IEC60601-1.
Connecting any additional equipment to the
signal input port or signal output port of this
display is considered configuring a medical
system. Therefore, the system must comply
with the requirements of the system standard
IEC60601-1.
Disconnecting devices
A power cord is included with the display. If
another cord is used, use only a power source
and connection appropriate for this display. For
information on the correct power cord set to
use with the display, refer to the
Product Notices
provided in your documentation kit.
WARNING!
To reduce the risk of electric shock or
damage to the equipment:
• Plug the power cord into an AC outlet that is easily
accessible at all times.
• Disconnect power from the computer by unplugging
the power cord from the AC outlet.
• If provided with a 3-pin attachment plug on the power
cord, plug the cord into a grounded (earthed) 3-pin
outlet. Do not disable the power cord grounding
pin, for example, by attaching a 2-pin adapter. The
grounding pin is an important safety feature.
• Installations must be set up by qualified personnel
only.
• It is the responsibility of the operating institution
to choose qualified personnel for installation and
operation of the device.
Cleaning the display
1. Turn off the display.
2. Disconnect any external devices.
3. Wipe the exterior with a soft, water-
dampened cloth to remove any visible soiling.
4. Next, use any of the following germicidal
wipe formulations to safely disinfect all
exposed surfaces of your monitor. Refer
to the directions for use provided by the
manufacturer of the wipes.
CAUTION:
The cloth should be moist, but not wet.
Water dripping into the ventilation or other points
of entry can cause damage to the monitor. Allow the
unit to air-dry before use.
PRINTER: Replace this box with Printed- In (PI) Statement(s) as per spec.
NOTE: This box is simply a placeholder. PI Statement(s) do not have to fit
inside the box but should be placed in this area.
*L39819-DN2*
L39819-DN2
Technical specifications
All specifications represent the typical
specifications provided by HP’s component
manufacturers; actual performance may vary either
higher or lower.
For the latest specifications or additional
specifications on this product, go to http://www.
hp.com/go/quickspecs/ and search for your
specific display model to find the model-specific
QuickSpecs.
Display
68.6 cm (27.0 in) wide screen
Type
AHVA
Viewable image size
68.6 cm (27.0 in) diagonal
Weight (Net)
HC271: 5.0 kg (11.0 lbs)
HC271p: 5.15 kg (11.4 lbs)
Weight with stand*
HC271: 7.16 kg (15.8 lbs)
HC271p: 7.31 kg (16.1 lbs)
*
(Optional) HP Stand Kit, product number 4BX37AA,
Weight: 2160 g, Dimensions: 210 x 196.9 x 434.8 mm
(W x H x D)
Dimensions
Height
Display head only: 39.25 cm (15.45 in)
With stand: 55.79 cm (21.96 in)
Depth
Display head only: 5.11 cm (2.01 in)
With stand: 19.7 cm (7.76 in)
Width
Display head only:
64.41 cm (25.36 in)
With stand: 64.41 cm (25.36 in)
Maximum graphic resolution
2560 x 1440 (60 Hz)
Optimum graphic resolution
2560 x 1440 (60 Hz)
Dot pitch
0.233 (H) x 0.233 (V) mm
Pixels per inch
109 PPI
Horizontal frequency
30 kHz to 90 kHz
Vertical refresh rate
50 Hz to 60 Hz
Environmental requirements
Temperature
Operating 5°C to 35°C (41°F to 95°F)
Storage/transportation
-20°C to 60°C (-3.99°F to 140°F)
Humidity
Operating: 20%-80%
Storage/transportation: 5% to 95% RH at 38.7°C
Altitude
Operating 0 m to 2,000 m (0 to 6,562 ft)
Storage/transportation: 0 m to 12,192 m (0 to
40,000 ft)
Atmospheric pressure range
Operating: 80 kPa to 101 kPa
Storage/transportation: 24.0 kPa to 101 kPa
Relative humidity
20% to 80%
Power source
100 V ac to 240 V ac 50/60 Hz
Measured power consumption
Full power: 55 W
Typical settings: 45 W
Sleep: <0.5 W
Off: <0.3 W
Input terminal
One DisplayPort port, one HDMI port, one VGA
port, one USB Type-B port (upstream), and two
USB 2.0 Type-A ports (downstream)
DICOM mode
HC271: available
HC271p: not available
IP Protection
IP32 on the front glass
Contacting HP
Go to www.hp.com/support to locate your
user guide, drivers, and software. For safety,
regulatory, and environmental information,
refer to the
Product Notices
provided with your
product.
© Copyright 2018 HP Development Company, L.P.
The information contained herein is subject to change without
notice. The only warranties for HP products and services are set
forth in the express warranty statements accompanying such
products and services. Nothing herein should be construed as
constituting an additional warranty. HP shall not be liable for
technical or editorial errors or omissions contained herein.
Second Edition: August 2018
First Edition: July 2018
Technical specifications
All specifications represent the typical
specifications provided by HP’s component
manufacturers; actual performance may vary either
higher or lower.
For the latest specifications or additional
specifications on this product, go to http://www.
hp.com/go/quickspecs/ and search for your
specific display model to find the model-specific
QuickSpecs.
Display
68.6 cm (27.0 in) wide screen
Type
AHVA
Viewable image size
68.6 cm (27.0 in) diagonal
Weight (Net)
HC271: 5.0 kg (11.0 lbs)
HC271p: 5.15 kg (11.4 lbs)
Weight with stand*
HC271: 7.16 kg (15.8 lbs)
HC271p: 7.31 kg (16.1 lbs)
*
(Optional) HP Stand Kit, product number 4BX37AA,
Weight: 2160 g, Dimensions: 210 x 196.9 x 434.8 mm
(W x H x D)
Dimensions
Height
Display head only: 39.25 cm (15.45 in)
With stand: 55.79 cm (21.96 in)
Depth
Display head only: 5.11 cm (2.01 in)
With stand: 19.7 cm (7.76 in)
Width
Display head only:
64.41 cm (25.36 in)
With stand: 64.41 cm (25.36 in)
Maximum graphic resolution
2560 x 1440 (60 Hz)
Optimum graphic resolution
2560 x 1440 (60 Hz)
Dot pitch
0.233 (H) x 0.233 (V) mm
Pixels per inch
109 PPI
Horizontal frequency
30 kHz to 90 kHz
Vertical refresh rate
50 Hz to 60 Hz
Environmental requirements
Temperature
Operating 5°C to 35°C (41°F to 95°F)
Storage/transportation
-20°C to 60°C (-3.99°F to 140°F)
Humidity
Operating: 20%-80%
Storage/transportation: 5% to 95% RH at 38.7°C
Altitude
Operating 0 m to 2,000 m (0 to 6,562 ft)
Storage/transportation: 0 m to 12,192 m (0 to
40,000 ft)
Atmospheric pressure range
Operating: 80 kPa to 101 kPa
Storage/transportation: 24.0 kPa to 101 kPa
Relative humidity
20% to 80%
Power source
100 V ac to 240 V ac 50/60 Hz
Measured power consumption
Full power: 55 W
Typical settings: 45 W
Sleep: <0.5 W
Off: <0.3 W
Input terminal
One DisplayPort port, one HDMI port, one VGA
port, one USB Type-B port (upstream), and two
USB 2.0 Type-A ports (downstream)
DICOM mode
HC271: available
HC271p: not available
IP Protection
IP32 on the front glass
Contacting HP
Go to www.hp.com/support to locate your
user guide, drivers, and software. For safety,
regulatory, and environmental information,
refer to the
Product Notices
provided with your
product.
EMC Notice
Electromagnetic immunity
Immunity test
IEC 60601 Test levels
Compliance level
Electromagnetic environment–Guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/burst
IEC 61000-4-4
±2 kV for power supply lines ±1
kV for input/ output lines
±2 kV for power supply lines ±1
kV for input/ output lines
The main power quality should be that of a typical commercial or hospital
environment
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV line(s) to line(s)
±2 kV line(s) to earth
The main power quality should be that of a typical commercial or hospital
environment.
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC 61000-4-11
< 5%
U
T
1 (> 95% dip in
U
T
) for
0.5 cycle, 40%
U
T
(60% dip in
U
T
)
for 5 cycles, 70%
U
T
(30% dip in
U
T
) for 25 cycles, < 5%
U
T
(>95%
dip in
U
T
) for 5s
< 5%
U
T
(> 95% dip in
U
T
) for 0.5
cycle, 40%
U
T
(60% dip in
U
T
) for
5 cycles, 70%
U
T
(30% dip in
U
T
)
for 25 cycles, < 5%
U
T
(>95% dip
in
U
T
) for 5s
The main power quality should be that of a typical commercial or hospital
environment. If the user of the HC271/HC271p requires continued operation
during interruptions in the main power, it is recommended that the HC271/
HC271p be powered from an uninterruptible power supply or a battery.
Power frequency
(50 Hz) magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields should be at least characteristic of a typical
location in a typical commercial or hospital environment.
Conducted RF induced by RF fields
IEC 61000-4-6
3 Vrms 150 kHz - 80 MHz 6Vrms
ISM and Amateur Radio Bands
3 Vrms 150 kHz - 80 MHz, 6Vrms
ISM and Amateur Radio Bands
Portable and mobile RF communications equipment should be used no
closer to any part of the OEV262H, including cables, than the recommended
separation distance calculated from the equation appliance to the frequency
of the transmitter. Recommended separation distance: d = 1.2 √P, d = 1.2
√P 80 MHz to 800 MHz, d = 2.3 √P 800 MHz to 2.7 GHz, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recommended separation distance in
meters (m). Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey
a
, should be less than the compliance level in each
frequency range
b
. Interference may occur in the vicinity of equipment marked
with following symbol:
Radiated RF EM fields and
Proximity fields from RF wireless
communications equipment
IEC 61000-4-3
10 V/m, 80 MHz - 2,7 GHz,
80% AM 1kHz
10 V/m
80 MHz – 2,7 GHz
385MHz (18Hz Pulse Modulation)
27 V/m
450MHz (FM+/- 5KHz deviation
1kHz sine or 18Hz Pulse
Modulation)
28 V/m
710MHz (217Hz PM)
9 V/m
745MHz (217Hz PM)
9 V/m
780MHz (217Hz PM)
9 V/m
810MHz (18Hz PM)
28 V/m
870MHz (18Hz PM)
28 V/m
930MHz (18Hz PM)
28 V/m
1720MHz (217Hz PM)
28 V/m
1845MHz (217Hz PM)
28 V/m
1970MHz (217Hz PM)
28 V/m
2450MHz (217Hz PM)
28 V/m
5240MHz (217Hz PM)
9 V/m
5500MHz (217Hz PM)
9 V/m
5785MHz (217Hz PM)
9 V/m
At 80 MHz and 800 MHz, the higher frequency
range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength
in the location in which the HC271/HC271p is used exceeds the applicable RF compliance level above, the HC271/HC271p should be observed to verify normal operation. If abnormal performance
is observed, additional measures may be necessary, such as reorienting or relocating the HC271/HC271p.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
EMC Notice
Electromagnetic immunity
Immunity test
IEC 60601 Test levels
Compliance level
Electromagnetic environment–Guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/burst
IEC 61000-4-4
±2 kV for power supply lines ±1
kV for input/ output lines
±2 kV for power supply lines ±1
kV for input/ output lines
The main power quality should be that of a typical commercial or hospital
environment
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV line(s) to line(s)
±2 kV line(s) to earth
The main power quality should be that of a typical commercial or hospital
environment.
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC 61000-4-11
< 5%
U
T
1 (> 95% dip in
U
T
) for
0.5 cycle, 40%
U
T
(60% dip in
U
T
)
for 5 cycles, 70%
U
T
(30% dip in
U
T
) for 25 cycles, < 5%
U
T
(>95%
dip in
U
T
) for 5s
< 5%
U
T
(> 95% dip in
U
T
) for 0.5
cycle, 40%
U
T
(60% dip in
U
T
) for
5 cycles, 70%
U
T
(30% dip in
U
T
)
for 25 cycles, < 5%
U
T
(>95% dip
in
U
T
) for 5s
The main power quality should be that of a typical commercial or hospital
environment. If the user of the HC271/HC271p requires continued operation
during interruptions in the main power, it is recommended that the HC271/
HC271p be powered from an uninterruptible power supply or a battery.
Power frequency
(50 Hz) magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields should be at least characteristic of a typical
location in a typical commercial or hospital environment.
Conducted RF induced by RF fields
IEC 61000-4-6
3 Vrms 150 kHz - 80 MHz 6Vrms
ISM and Amateur Radio Bands
3 Vrms 150 kHz - 80 MHz, 6Vrms
ISM and Amateur Radio Bands
Portable and mobile RF communications equipment should be used no
closer to any part of the OEV262H, including cables, than the recommended
separation distance calculated from the equation appliance to the frequency
of the transmitter. Recommended separation distance: d = 1.2 √P, d = 1.2
√P 80 MHz to 800 MHz, d = 2.3 √P 800 MHz to 2.7 GHz, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recommended separation distance in
meters (m). Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey
a
, should be less than the compliance level in each
frequency range
b
. Interference may occur in the vicinity of equipment marked
with following symbol:
Radiated RF EM fields and
Proximity fields from RF wireless
communications equipment
IEC 61000-4-3
10 V/m, 80 MHz - 2,7 GHz,
80% AM 1kHz
10 V/m
80 MHz – 2,7 GHz
385MHz (18Hz Pulse Modulation)
27 V/m
450MHz (FM+/- 5KHz deviation
1kHz sine or 18Hz Pulse
Modulation)
28 V/m
710MHz (217Hz PM)
9 V/m
745MHz (217Hz PM)
9 V/m
780MHz (217Hz PM)
9 V/m
810MHz (18Hz PM)
28 V/m
870MHz (18Hz PM)
28 V/m
930MHz (18Hz PM)
28 V/m
1720MHz (217Hz PM)
28 V/m
1845MHz (217Hz PM)
28 V/m
1970MHz (217Hz PM)
28 V/m
2450MHz (217Hz PM)
28 V/m
5240MHz (217Hz PM)
9 V/m
5500MHz (217Hz PM)
9 V/m
5785MHz (217Hz PM)
9 V/m
At 80 MHz and 800 MHz, the higher frequency
range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength
in the location in which the HC271/HC271p is used exceeds the applicable RF compliance level above, the HC271/HC271p should be observed to verify normal operation. If abnormal performance
is observed, additional measures may be necessary, such as reorienting or relocating the HC271/HC271p.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Electromagnetic emissions
The HC271/HC271p is intended for use in the
electromagnetic environment specified below. The
customer or the user of the HC271/HC271p should assure
that it is used in such an environment.
Emissions test
Compliance
Electromagnetic
environment–
Guidance
RF emissions
CISPR 32
Group 1
The HC271/HC271p
uses RF energy only
for its internal function.
Therefore, its RF
emissions are very low
and are not likely to
cause any interference
in nearby electronic
equipment.
RF emissions
CISPR 32
Class B
The HC271/HC271p
is suitable for use in
all establishments,
including domestic
establishments
and those directly
connected to the public
low-voltage power
supply network that
supplies buildings used
for domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Class D
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Disposal of waste equipment
by users
This symbol means do not dispose of your
product with your other household waste.
Instead, you should protect human health
and the environment by handing over your waste
equipment to a designated collection point for the
recycling of waste electrical and electronic
equipment. For more information, please contact
your household waste disposal service or go to
FCC Statement
This device complies with Part 15 and Part 18
of the FCC Rules. Operation is subject to the
following two conditions: (1) this device may not
cause harmful interference, and (2) this device
must accept any interference received, including
interference that may cause undesired operation.
Electromagnetic emissions
The HC271/HC271p is intended for use in the
electromagnetic environment specified below. The
customer or the user of the HC271/HC271p should assure
that it is used in such an environment.
Emissions test
Compliance
Electromagnetic
environment–
Guidance
RF emissions
CISPR 32
Group 1
The HC271/HC271p
uses RF energy only
for its internal function.
Therefore, its RF
emissions are very low
and are not likely to
cause any interference
in nearby electronic
equipment.
RF emissions
CISPR 32
Class B
The HC271/HC271p
is suitable for use in
all establishments,
including domestic
establishments
and those directly
connected to the public
low-voltage power
supply network that
supplies buildings used
for domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Class D
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Disposal of waste equipment
by users
This symbol means do not dispose of your
product with your other household waste.
Instead, you should protect human health
and the environment by handing over your waste
equipment to a designated collection point for the
recycling of waste electrical and electronic
equipment. For more information, please contact
your household waste disposal service or go to
FCC Statement
This device complies with Part 15 and Part 18
of the FCC Rules. Operation is subject to the
following two conditions: (1) this device may not
cause harmful interference, and (2) this device
must accept any interference received, including
interference that may cause undesired operation.
S