Invacare MA85 Owners Manual

Invacare MA85 Manual

Invacare MA85 manual content summary:

  • Invacare MA85 | Owners Manual - Page 1
    microAIR® MA80 Series and MA85 (with Alternating Pressure) Series MA80, MA80B42, MA80B48, MA80RSR, MA85, MA85B42, MA85B48, MA85RSR en Low Air Loss Mattress User Manual This manual MUST be given to the user of the product. BEFORE using this product, this manual MUST be read and saved for future
  • Invacare MA85 | Owners Manual - Page 2
    or modification in whole or in part is prohibited without prior written permission from Invacare. Trademarks are identified by ™ and ®. All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted. Making Life's Experiences Possible is a registered
  • Invacare MA85 | Owners Manual - Page 3
    Using the Front Panel 20 5.1.1 Power Button 21 5.1.2 Pulse Button (MA80 Power Unit 21 5.1.3 Alternating Pressure Button (MA85 Power Unit) . Service Life 31 6.2 Cleaning the System 31 6.3 Changing the Filter 33 6.4 Disposal 34 6.5 Storing the System 34 7 Troubleshooting 35 7.1 Troubleshooting
  • Invacare MA85 | Owners Manual - Page 4
    microAIR® MA80 Series and MA85 (with Alternating Pressure) Series 1 General 1.1 Symbols Signal words are used in this manual and apply to hazards of efficient, trouble-free use. THE INFORMATION CONTAINED IN THIS DOCUMENT IS SUBJECT TO CHANGE WITHOUT NOTICE. 1.2 Intended Use The microAIR® MA80 and MA
  • Invacare MA85 | Owners Manual - Page 5
    this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as user manuals, service manuals or instruction sheets supplied with this product or optional equipment. DANGER! Risk of Death, Injury
  • Invacare MA85 | Owners Manual - Page 6
    microAIR® MA80 Series and MA85 (with Alternating Pressure) Series DANGER! Risk Of Death, Injury Or due to misuse or modification: - DO NOT use this product in any way other than described in this manual. - DO NOT modify the product or any components of the product. WARNING! Risk Of Death Or Injury
  • Invacare MA85 | Owners Manual - Page 7
    power cords and outlets for damage. Stop using the product immediately if the cord or outlet is damaged. Contact qualified technician or Invacare for service. WARNING! Risk Of Injury Or Damage To avoid electrical damage, product damage, shock and/or personal injury: - Keep all bed components and
  • Invacare MA85 | Owners Manual - Page 8
    microAIR® MA80 Series and MA85 (with Alternating Pressure) Series WARNING! Risk Of Injury Or Damage To avoid personal injury or damage: - Check for leakage from electrical components such as batteries and circuit boards. - Remove product from service and contact dealer or Invacare for repair if
  • Invacare MA85 | Owners Manual - Page 9
    risk of death or injury from entrapment or falling: - Invacare suggests rails be in the raised or guard position whenever a oxygen administering equipment. - Do not allow oxygen tent to extend below mattress support level. Installation WARNING! Risk Of Injury Or Damage To avoid serious injury or
  • Invacare MA85 | Owners Manual - Page 10
    microAIR® MA80 Series and MA85 (with Alternating Pressure) Service DANGER! Risk of Death, Injury, or Damage To avoid death, injury or damage due to improper maintenance or inspection. Always maintain and inspect equipment per the instructions in this manual. Contact a qualified technician or Invacare
  • Invacare MA85 | Owners Manual - Page 11
    3 Overview 3.1 Label Locations Labels are subject to change without notice. This label is on the back of the unit and also contains the serial number. Overview WARNING! Risk of Injury, Damage or Death Missing or damaged labels may contribute to injury, damage or death. - Ensure labels are present
  • Invacare MA85 | Owners Manual - Page 12
    microAIR® MA80 Series and MA85 (with Alternating Pressure) Series 3.2 Technical Data Electrical Specifications Input Voltage Ac: Input Frequency: Current: Maximum Power Consumption: Circuit Protection: Mode Of Operation: MA80/MA85 110 V 60 Hz 2A 150 ± 10 W Dual fused, 250 V, 5 A fast blow fuses
  • Invacare MA85 | Owners Manual - Page 13
    Timer: Support Surface Inflation Time: Patient Comfort Control Pressures: Soft Pressure: Firm Pressure: Cycle Time (minutes): Patient Contact: MA80 MA85 350 Control Unit Dimensions (Lxwxh): Weight: Power Cord: Connection: Packaging: Air Filter: MA80 MA85 15" x 6.5" x 11" 15" x 6.5" x 11" 15
  • Invacare MA85 | Owners Manual - Page 14
    microAIR® MA80 Series and MA85 (with Alternating Pressure) Series 4 Setup 4.1 Unpacking 1. Remove any loose packing from the carton. 2. Carefully remove all components from the carton. 3. Inspect all components for damage. Contact dealer or Invacare if components are damaged. DO NOT use product.
  • Invacare MA85 | Owners Manual - Page 15
    4.2 Mattress Replacement System Installation CAUTION! Risk Of Damage To avoid damage to the mattress: - Do not strap the mattress to the bed frame at the head and foot ends. - Secure all mattress straps. - Secure the straps to the bed deck at the head and foot ends and to the frame at the center of
  • Invacare MA85 | Owners Manual - Page 16
    microAIR® MA80 Series and MA85 (with Alternating Pressure) Series 4.3 Side Rails Installation Refer to the Bed Rail Entrapment Risk Notification Guide at www.invacare.com for additional safety information. Refer to the instructions provided with the side rails for the installation procedure. 4.4
  • Invacare MA85 | Owners Manual - Page 17
    3. Perform one of the following: • MA80 Power Unit-Perform the following: C D Setup • MA85 Power Unit-Perform the following: a. Squeeze and hold the tabs C on the hose connector D. b. Push the hose connectors
  • Invacare MA85 | Owners Manual - Page 18
    microAIR® MA80 Series and MA85 (with Alternating Pressure) Series 4.6 Connecting the Power Cord WARNING! Risk Of Injury Or Damage To avoid serious injury or product damage from
  • Invacare MA85 | Owners Manual - Page 19
    needs to be installed and used according to the EMC information in this manual. Setup This equipment has been tested and found to comply with EMC limits line powered device to a different electrical power circuit. Contact Invacare for any additional EMC information or EMC tables for the device environment
  • Invacare MA85 | Owners Manual - Page 20
    microAIR® MA80 Series and MA85 (with Alternating Pressure) Series 5 Usage 5.1 Using the Front Panel DETAIL "A" - MA80 FRONT PANEL DETAIL "B" - MA85 FRONT PANEL 20 1148138-I
  • Invacare MA85 | Owners Manual - Page 21
    a power outage, the control unit will go through its system initialization routine for a few seconds and then resume the desired function. 5.1.2 Pulse Button (MA80 Power Unit) 1. Press the Pulse button B to turn on Pulse mode and select one of four Pulse Times - 1, 3, 5, or 10 minutes. The pulse
  • Invacare MA85 | Owners Manual - Page 22
    microAIR® MA80 Series and MA85 (with Alternating Pressure) Series 5.1.4 Adjusting mmHg). The Comfort Control LED displays the patient comfort pressure levels from 0 to 9 and provides a guide to the caregiver to set approximate comfort pressure level depending on the patient weight. If the patient's
  • Invacare MA85 | Owners Manual - Page 23
    5.1.7 Fowler Button Manual Mode 1. Press the Fowler button F to activate the patient fowler mode. The Fowler LED illuminates when in this mode. When this mode is activated, the
  • Invacare MA85 | Owners Manual - Page 24
    microAIR® MA80 Series and MA85 (with Alternating Pressure) Series Fowler Transmitter Battery Replacement (if equipped) care, patient turning or patient cleaning. 2. Press the Max Inflate button again to manually disengage Max Inflate mode. This will deactivate the Max Inflate LED. 24 1148138-I
  • Invacare MA85 | Owners Manual - Page 25
    internal memory and retains the previous settings during the power outage. During a power outage, the MA80 series and MA85 series mattresses do not retain air. The mattress will deflate in the event of outlet while unit is on. See troubleshooting section of this manual if alarm fails to sound. 25
  • Invacare MA85 | Owners Manual - Page 26
    microAIR® MA80 Series and MA85 (with Alternating Pressure) Series 5.1.11 Low Pressure LED In the event that the mattress hose disconnects, an alarm sounds and Low Pressure LED K flashes Amber. Once the low pressure problem is fixed, the control unit resumes operation in the previously set mode.
  • Invacare MA85 | Owners Manual - Page 27
    to adjust the comfort settings. 11. Use a regular pillow to help support and stabilize the patient's head. Usage 5.4 Transferring Patient From/To a Gurney or lower the bed to match the gurney height. Refer to the user manual provided with the bed. 5. When the mattress has reached maximum firmness,
  • Invacare MA85 | Owners Manual - Page 28
    microAIR® MA80 Series and MA85 (with Alternating Pressure) Series 5.5 Transferring the Patient From/ pressure. 4. Raise or lower the bed to match the wheelchair height. Refer to the user manual provided with the bed. 5. When the mattress has reached maximum firmness, perform one of the following
  • Invacare MA85 | Owners Manual - Page 29
    5.6 Preparing for CPR Procedure MA80 A B Usage 1. Press and hold the tabs A on the hose connector B. 2. Pull the hose connector B from the control unit connectors. C 3. Disconnect the red CPR connector C located on the hose. 1148138-I 29
  • Invacare MA85 | Owners Manual - Page 30
    microAIR® MA80 Series and MA85 (with Alternating Pressure) Series MA85 1. Lift the connector cover A. 2. Pull the hose connector C from the control unit connectors. If the hose connector
  • Invacare MA85 | Owners Manual - Page 31
    6.1 Service Life WARNING! Risk of Injury or Damage Use of the product beyond this time period may cause product damage and injury. - This product has an expected lifetime of one (1) years when used in accordance with safety instructions, maintenance intervals and correct use stated in this manual
  • Invacare MA85 | Owners Manual - Page 32
    microAIR® MA80 Series and MA85 (with Alternating Pressure) Series 7. Perform one of the ii. Wash with warm water (below 120°F). iii. Add detergent and disinfectant according to the manufacturer's instructions. iv. Remove excess water. v. Set the dryer to the lowest setting (below 120°F). vi. Dry
  • Invacare MA85 | Owners Manual - Page 33
    9. Perform the following steps to clean the control unit and hose fittings: a. Wipe all controls, chassis and hose fittings with a quaternary disinfectant solution. b. Using a nylon brush, gently clean all crevices as they can harbor microorganisms. c. Air dry all treated surfaces. 10. Perform the
  • Invacare MA85 | Owners Manual - Page 34
    microAIR® MA80 Series and MA85 (with Alternating Pressure) Series 6.4 Disposal Follow local governing ordinances and recycling plans regarding disposal of the product or components normally used in
  • Invacare MA85 | Owners Manual - Page 35
    7 Troubleshooting 7.1 Troubleshooting Problem Mattress not inflating Not alternating properly No power Fowler Transmitter not working 1148138-I Troubleshooting Cause Solution Mattress hose disconnected Connect hose synchronized to power unit Refer to Folwer transmitter setup in this manual 35
  • Invacare MA85 | Owners Manual - Page 36
    microAIR® MA80 Series and MA85 (with Alternating Pressure) Series Problem Alarm-Power Failure Cause Power cord disconnected No power in the power source Alarm-Hose Disconnected Power outage Blown fuse Mattress hose disconnected Solution Connect
  • Invacare MA85 | Owners Manual - Page 37
    service, please contact the dealer from whom you purchased your Invacare product. In the event you do not receive satisfactory warranty service, please write directly to Invacare SHALL NOT APPLY TO PROBLEMS ARISING FROM NORMAL WEAR OR FAILURE TO ADHERE TO THE ENCLOSED INSTRUCTIONS. IN ADDITION, THE
  • Invacare MA85 | Owners Manual - Page 38
    microAIR® MA80 Series and MA85 (with Alternating Pressure) Series PRODUCTS DAMAGED BY REASON OF REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC CONSENT OF INVACARE; PRODUCTS DAMAGED BY CIRCUMSTANCES BEYOND INVACARE'S CONTROL; PRODUCTS REPAIRED BY ANYONE OTHER THAN AN INVACARE DEALER, SUCH
  • Invacare MA85 | Owners Manual - Page 39
    Notes
  • Invacare MA85 | Owners Manual - Page 40
    Manufacturer KAP Medical 1395 Pico Street Corona, CA 92881 Distributed by: Invacare Corporation USA One Invacare Way Elyria, Ohio USA 44035 800-333-6900 1148138-I 2019-04-04 *1148138I* www.invacare.com Canada 570 Matheson Blvd E Unit 8 Mississauga Ontario L4Z 4G4 Canada 800-668-5324 Making Life's
  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40

microAIR® MA80 Series and
MA85 (with Alternating Pressure)
Series
MA80, MA80B42, MA80B48, MA80RSR, MA85,
MA85B42, MA85B48, MA85RSR
en
Low Air Loss Mattress
User Manual
This manual MUST be given to the user of the product.
BEFORE using this product, this manual MUST be read and saved for
future reference.