Invacare TDXSEAT Owners Manual

Invacare TDXSEAT Manual

Invacare TDXSEAT manual content summary:

  • Invacare TDXSEAT | Owners Manual - Page 1
    ™ Powered Seating Tilt Recline Elevate Tilt/Recline Tilt/Elevate Tilt/Recline/Elevate DEALER: This manual MUST be given to the user of the product. USER: BEFORE using this product, read this manual and save for future reference. For more information regarding Invacare products, parts, and services
  • Invacare TDXSEAT | Owners Manual - Page 2
    , CAUTIONS AND INSTRUCTIONS, CONTACT INVACARE TECHNICAL SUPPORT BEFORE ATTEMPTING TO SERVICE OR OPERATE THIS EQUIPMENT - OTHERWISE, INJURY OR DAMAGE MAY RESULT. REFERENCE DOCUMENTS DOCUMENT MK6i™Electronics Field Service Guide MK6i Electronics Service Manual Formula TRE Service Manual PART NUMBER
  • Invacare TDXSEAT | Owners Manual - Page 3
    WHEELCHAIRS 17 Pinch Points...17 SECTION 4-SAFETY INSPECTION/TROUBLESHOOTING 18 Safety Inspection Checklists...18 Troubleshooting...20 SECTION 5-WHEELCHAIR OPERATION 23 A Note About Drive Lock-Out ...23 Operating Powered /Replacing Telescoping Front Rigging Support 38 Installing Adjustable Angle
  • Invacare TDXSEAT | Owners Manual - Page 4
    From/Into the Wheelchair 66 When to invacare.com Please have your model number and purchase date available to complete your registration. Any registration information you submit will only be used by Invacare Corporation and protected as required by applicable laws and regulations. Formula™ Powered
  • Invacare TDXSEAT | Owners Manual - Page 5
    manual WHEELCHAIR USER As a manufacturer of wheelchairs, Invacare endeavors to supply a wide variety of wheelchairs an unoccupied wheelchair during wheelchair, in a moving vehicle of any type. It is Invacare's position that users of wheelchairs Invacare cannot and does not recommend any wheelchair
  • Invacare TDXSEAT | Owners Manual - Page 6
    Invacare accessories. Accessories designed by other manufacturers have not been tested by Invacare and are not recommended for use with Invacare more frequently. Refer to wheelchair base owner's manual for additional safety and operation information. ƽ POWERED SEATING SYSTEM WARNINGS This seating
  • Invacare TDXSEAT | Owners Manual - Page 7
    LABEL LOCATIONS LABEL LOCATIONS NOTE: Refer to Owner's Manual shipped with wheelchair base for additional labels located on wheelchair base. All Wheelchairs Weight Capacity Label located here WA R N I N G Refer to Service Manual to reposition the seat - otherwise serious personal injury and
  • Invacare TDXSEAT | Owners Manual - Page 8
    LABEL LOCATIONS Wheelchairs with Elevate Wheelchairs with Ventilator Trays Formula™ Powered Seating 8 Part No 1122170
  • Invacare TDXSEAT | Owners Manual - Page 9
    TYPICAL PRODUCT PARAMETERS TYPICAL PRODUCT PARAMETERS Formula Powered Seating Systems SEAT WIDTH RANGE: SEAT DEPTH items that the user may require [back pack, etc.]). Example: If weight limitation of the wheelchair is 300 lbs and additional items equal 25 lbs, subtract 25 lbs from 300 lbs. This
  • Invacare TDXSEAT | Owners Manual - Page 10
    Full Length LEGRESTS: Swingaway Footrests, Power or Manual Center Mount Footrests, Manual or Power Elevating Legrests HEADRESTS: Curved, Contoured, 275 lbs. ƽ WARNING If the seating system is mounted onto a power wheelchair that has a weight limitation greater than that of the seating system,
  • Invacare TDXSEAT | Owners Manual - Page 11
    understanding these instructions and any additional instructional material such as Owner's Manuals, Service Manuals or Instruction Sheets supplied , repairing or servicing the wheelchair, ALWAYS turn the wheelchair power Off, otherwise, injury or damage may occur. Invacare products are specifically
  • Invacare TDXSEAT | Owners Manual - Page 12
    Off immediately and reenter set-up specifications. Repeat this section until the wheelchair performs to specifications. DO NOT leave the power button On when entering or exiting your wheelchair. DO NOT operate on roads, streets or highways. DO NOT climb, go up or down ramps or traverse slopes
  • Invacare TDXSEAT | Owners Manual - Page 13
    the seat. DO NOT attempt to adjust the drive lock-out. Have the wheelchair serviced by a qualified technician. Use only the controls listed in Operating Powered Seating Systems on page 24 to activate the powered seating functions*. DO NOT USE any other actuator controls. Such devices may result
  • Invacare TDXSEAT | Owners Manual - Page 14
    as a seat for any kind of weight training. If occupant uses said wheelchair as a weight training apparatus, Invacare shall not be liable for bodily injury and the warranty is void. Weight Limitation Refer to Typical Product Parameters in this manual and the Typical Product Parameters in the owner
  • Invacare TDXSEAT | Owners Manual - Page 15
    radios, and cellular phones. The interference (from radio wave sources) can cause the powered wheelchair to release its brakes, move by itself, or move in unintended directions. It can also far as we know, are not likely to cause EMI problems to your powered wheelchair. Part No 1122170 15 Formula
  • Invacare TDXSEAT | Owners Manual - Page 16
    or turn ON personal communication devices, such as cellular phones, while the powered wheelchair is turned ON; 2) Be aware of nearby transmitters, such as radio of this wheelchair as manufactured by Invacare may adversely affect the EMI immunity levels. Formula™ Powered Seating 16 Part No 1122170
  • Invacare TDXSEAT | Owners Manual - Page 17
    NOTE: For more information about safety/handling of the wheelchair, refer to the owner's manual that shipped with the wheelchair base. Pinch Points ƽ WARNING A pinch point exists between head tube cap and Point Pinch Points FIGURE 3.1 Pinch Points Part No 1122170 17 Formula™ Powered Seating
  • Invacare TDXSEAT | Owners Manual - Page 18
    just like any other vehicle. Routine maintenance will extend the life and efficiency of your wheelchair. NOTE: Refer to wheelchair base owner's manual for additional safety inspection and troubleshooting information. Initial adjustments should be made to suit your personal body structure needs and
  • Invacare TDXSEAT | Owners Manual - Page 19
    SECTION 4-SAFETY INSPECTION/TROUBLESHOOTING ❑ Make sure seating systems with the recline function have the retaining pin in place at the top of each to release and adjustment levers engage properly. ❑ Adjustable height arms operate and lock securely. Part No 1122170 19 Formula™ Powered Seating
  • Invacare TDXSEAT | Owners Manual - Page 20
    °). completely lowered). Refer to Operating Powered Seating Systems on page 24. Contact Invacare/Dealer for service if Motor lock levers are this does not solve the problem. disengaged. Engage motor lock levers. See wheelchair base owner's manual. Low batteries. Charge batteries. Faulty
  • Invacare TDXSEAT | Owners Manual - Page 21
    TROUBLESHOOTING SYMPTOM PROBABLE CAUSE SOLUTIONS ƽ SPM Battery Fault displays and the Batteries need to be charged wheelchair power, reconnect the joystick the wheelchair does not drive. disconnected (Error code 32). of input device and turn power Contact Dealer/Invacare for Service. Have
  • Invacare TDXSEAT | Owners Manual - Page 22
    SOLUTIONS Contact Dealer/Invacare for Service. Contact Dealer/Invacare for Service. Controller programmed improperly. Contact Dealer/Invacare to have controller reprogrammed. Wheelchair does not respond to commands. Poor battery terminal connection. Have terminals cleaned. Power indicator off
  • Invacare TDXSEAT | Owners Manual - Page 23
    or elevate/lower the seat. DO NOT attempt to adjust the drive lock-out. Have the wheelchair serviced by a qualified technician. The wheelchair user MUST have a clear line of sight to drive safely. On initial chair delivery and after adjusting the back angle, drive lock-out switch, tilt
  • Invacare TDXSEAT | Owners Manual - Page 24
    wheelchair operates at maximum speed while in an elevated position, DO NOT operate the wheelchair. Have the wheelchair serviced . If such devices are used, Invacare shall not be liable and the Power Legs) MPJ+ POWERED SEATING CONTROL 4 POLE MOTOR SWITCH OPTION THROUGH THE JOYSTICK (TTJ) Powered
  • Invacare TDXSEAT | Owners Manual - Page 25
    1. Make sure the wheelchair is on a level surface. 2. Press the powered seating switch once to activate the actuator. 3. Press the powered seating switch again ( tilt function. Refer to the electronics manual for complete switch option operating instructions. 3. Release four way switch to neutral
  • Invacare TDXSEAT | Owners Manual - Page 26
    For this procedure, refer to FIGURE 5.2 on page 27. 1. Make sure the wheelchair is on a level surface. 2. Select a drive that has the Actuator Control four of the following icons display. The icons indicate the available powered seating operation choices. The location of the icons on the screen
  • Invacare TDXSEAT | Owners Manual - Page 27
    DETAIL "A" SELECTING THE ACTUATOR CONTROL SWITCH MODE Mode Icons DETAIL "B" ACTUATOR CONTROL SWITCH MODE EXAMPLE SCREEN Operation Icons SECTION 5-WHEELCHAIR OPERATION Actuator Control Switch Mode Icon Tilt Down Icon (Move the joystick right to use this function) DETAIL "C" - EXAMPLE SCREEN
  • Invacare TDXSEAT | Owners Manual - Page 28
    SECTION 5-WHEELCHAIR OPERATION MPJ+ Joystick Switches and Indicators NOTE: For this procedure, refer to FIGURE 5.3. Drive the joystick housing. 1. Rotate the knob clockwise (forward) to increase the speed of the wheelchair to the programmed max speed. Formula™ Powered Seating 28 Part No 1122170
  • Invacare TDXSEAT | Owners Manual - Page 29
    is located in front of the joystick and provides information on the status of the wheelchair through a backlit display. The LCD display is readable in both bright sunlight and complete The Information Center displays current data on the wheelchair. Part No 1122170 29 Formula™ Powered Seating
  • Invacare TDXSEAT | Owners Manual - Page 30
    SECTION 5-WHEELCHAIR OPERATION Refer to LCD Display table on page 30 for descriptions Battery Level and will change depending on the available battery power. This indicator is shown on every screen. This area displays status or instructions. The status indicator will show a "Warning" (exclamation
  • Invacare TDXSEAT | Owners Manual - Page 31
    Indicator Battery Gauge Display Icon appears when a joystick command will drive the Status/Info ECU wheelchair Message Area FIGURE 5.5 LCD Display Screens - Main Screen Status Indicator Driving Screen NOTE: that are connected to the chair. Part No 1122170 31 Formula™ Powered Seating
  • Invacare TDXSEAT | Owners Manual - Page 32
    SECTION 5-WHEELCHAIR OPERATION ICON DESCRIPTION 4-Switch Attendant Control RIM Control ECU (1 to 4) Proportional Attendant Control Compact Joystick Sip and Puff Control ASL Digital Control ASL Analog Control Shark Power Module Actuator FIGURE 5.7 LCD Display Screens - Connected Devices Screen
  • Invacare TDXSEAT | Owners Manual - Page 33
    Mode* • Information Center Display Selection (does not require Reset activation at power up) If any of the above modes are selected, the control will with the basic or professional memory card for saving or reading wheelchair parameters. Preparing the Joystick for Use NOTE: For this procedure,
  • Invacare TDXSEAT | Owners Manual - Page 34
    to release the joystick mounting tube from the mounting bracket. 2. Remove the joystick from the wheelchair. Washers Mounting Bracket (Threaded Plate) Arm Tube 3. Remove the three hex screws that the joystick mounting tube into the mounting bracket. Formula™ Powered Seating 34 Part No 1122170
  • Invacare TDXSEAT | Owners Manual - Page 35
    RIGGINGS ƽ WARNING After ANY adjustments, repair or service and BEFORE use, make sure that all attaching NOT stand on the flip-up footboard. When getting in or out of the wheelchair, make sure that the flip-up footboard is in the upward position. PINCH POINT No 1122170 35 Formula™ Powered Seating
  • Invacare TDXSEAT | Owners Manual - Page 36
    Repeat STEPS 1-3 for the opposite side of the wheelchair footrest, if necessary. NOTE: PH904A style front rigging shown. PHAL4A front rigging adjust the same way. Footrest Support Locknut Lower Footrest Lug Bolt FIGURE 6.3 Footrest Height Adjustment Formula™ Powered Seating 36 Part No 1122170
  • Invacare TDXSEAT | Owners Manual - Page 37
    lower footrest to the footrest support and tighten securely. 6. Repeat STEPS 1-5 for the opposite wheelchair footrest, if necessary. 7. Reinstall the footrest(s) onto the wheelchair. Refer to Installing/ Removing screw and coved spacer. Securely tighten. Part No 1122170 37 Formula™ Powered Seating
  • Invacare TDXSEAT | Owners Manual - Page 38
    raise front rigging to the desired height. • Lowering - Support front rigging with one hand away from the release lever. ensure stability - otherwise personal injury and/or damage to the wheelchair and surrounding property may result. NOTE: For this procedure, Powered Seating 38 Part No 1122170
  • Invacare TDXSEAT | Owners Manual - Page 39
    SECTION 6-FRONT RIGGINGS Side Rail Spacers Cap Screws Side Rail Cap Screws Spacers Threaded Blocks Telescoping Front Tube FIGURE 6.7 Adjusting/Replacing Telescoping Front Rigging Support Part No 1122170 39 Formula™ Powered Seating
  • Invacare TDXSEAT | Owners Manual - Page 40
    support tube at the desired height. 2. Position mounting screw, washers, spacer, and locknut on the footrest support Footrest Support support until the footplate hinge remains in the up position. 5. Check the up and down motion of the footplate hinge to make sure the user of the wheelchair Support
  • Invacare TDXSEAT | Owners Manual - Page 41
    Footrest Support Footplate 2. Slowly turn nylon adjustment screw in or out until footplate is perpendicular to the footrest assembly or the desired inversion or eversion is obtained (FIGURE 6.11). FIGURE 6.11 Perpendicular and/or Inversion/Eversion Adjustment Part No 1122170 41 Formula™ Powered
  • Invacare TDXSEAT | Owners Manual - Page 42
    6.12 Composite/Articulating Footplate Heel Loop Replacement NOTE: When securing heel loop to the footrest assembly, tighten mounting screw until the spacer is secure. Formula™ Powered Seating 42 Part No 1122170
  • Invacare TDXSEAT | Owners Manual - Page 43
    front of the wheelchair) and swing the legrest to the outside of the wheelchair. 2. Lift UP on the legrest and remove from the wheelchair. 3. Repeat STEPS 1- 2 for opposite side of wheelchair. Bolt/Nut FIGURE 6.13 Installing/Removing Elevating Legrests Part No 1122170 43 Formula™ Powered Seating
  • Invacare TDXSEAT | Owners Manual - Page 44
    back on the release lever until the leg is at the desired height. • Lowering - Support leg with one hand and push release lever downward with other hand. Adjust Calfpad Secure Calfpad 3. Turn the calfpad towards the inside of the wheelchair. Formula™ Powered Seating 44 Part No 1122170
  • Invacare TDXSEAT | Owners Manual - Page 45
    seat frame. Installing Power Center Mount Footrests 1. Position the link drive tube between the mounting holes of the center support (Detail "C"). rigging pivot pin secure the center mount footrest to the mounting bracket. Manual Center Mount Footrests 1. Engage the release lever with one hand, hold
  • Invacare TDXSEAT | Owners Manual - Page 46
    of the seat frame. Center Mount Footrest DETAIL "A" Power Center Mount Footrest Top of Support DETAIL "B" Link Plate Short Quick-Release Pin Center Pin Center Support Center Mount Footrest Mounting Holes (Not Shown) DETAIL "C" Long Quick-Release Pin Link Drive Tube Manual Center Mount
  • Invacare TDXSEAT | Owners Manual - Page 47
    Height of the Center Mount Footrest ƽ WARNING While the wheelchair is moving, minimum ground clearance for the front rigging is three inches. If the wheelchair is not moving, the front rigging MUST maintain a the Height of the Center Mount Footrest Part No 1122170 47 Formula™ Powered Seating
  • Invacare TDXSEAT | Owners Manual - Page 48
    the Angle of the Manual Center Mount Footrest ƽ WARNING While the wheelchair is moving, minimum ground clearance for the front rigging is three inches. If the wheelchair is not moving, the FIGURE 6.18 Adjusting the Footplate Angle - After 2/14/07 Formula™ Powered Seating 48 Part No 1122170
  • Invacare TDXSEAT | Owners Manual - Page 49
    2/15/07 Adjusting the Tension of the Flip Up Footplate NOTE: For this procedure, refer to FIGURE 6.20. NOTE: This procedure applies to both power and manual center mount footrests. NOTE: The tension can be adjusted to increase or decrease the rotation effort of the flip up footplates. 1. Loosen the
  • Invacare TDXSEAT | Owners Manual - Page 50
    SECTION 6-FRONT RIGGINGS Calf Pad Adjustment for Power Center Mount Footrests NOTE: For this procedure, refer to FIGURE 6.21 on page Pad Mounting Bracket Calf Pad Depth Adjustment Screws FIGURE 6.21 Calf Pad Adjustment for Power Center Mount Footrests Formula™ Powered Seating 50 Part No 1122170
  • Invacare TDXSEAT | Owners Manual - Page 51
    6.22). NOTE: Make sure the legrest sits flush on the seat frame. 3. Rotate the power legs toward the INSIDE of the wheelchair until it locks in place. NOTE: The footplate will be on the INSIDE of the wheelchair when locked in place. 4. Repeat STEPS 1-3 for the opposite legrest. Part No 1122170 51
  • Invacare TDXSEAT | Owners Manual - Page 52
    jumper cable. 2. Push legrest release handle and swing legrest to the outside of the wheelchair. 3. Lift up on powered legrest and remove from wheelchair. 4. Repeat STEPS 1-3 for opposite power legrest. Mounting Hole DETAIL "A" Mounting Pin Calf Pad DETAIL "B" Footplate Jumper Cable Legrest
  • Invacare TDXSEAT | Owners Manual - Page 53
    service and before use, make sure that all attaching hardware is tightened securely - otherwise injury or damage may result. Installing/Removing Flip Back Armrests ƽ WARNING Make sure the flip back armrest release and height adjustment levers are in the locked position before using the wheelchair
  • Invacare TDXSEAT | Owners Manual - Page 54
    back armrest release and height adjustment levers are in the locked position before using the wheelchair. NOTE: For this procedure, refer to FIGURE 7.2. Positioning Flip Back Armrests for User height adjustment lever into the down (vertical) position. Formula™ Powered Seating 54 Part No 1122170
  • Invacare TDXSEAT | Owners Manual - Page 55
    refer to FIGURE 7.3. Removing Armrests 1. Lift the armrest release lever at the front of the wheelchair to the unlocked (horizontal) position. 2. Lift up on the armrest and remove from the front . FIGURE 7.3 Removing/Installing Reclining Armrests Part No 1122170 55 Formula™ Powered Seating
  • Invacare TDXSEAT | Owners Manual - Page 56
    Height NOTE: For this procedure, refer to FIGURE 7.4 on page 57. 1. Make sure the seating system is in the full upright position. Refer to Operating Powered Seating Systems on page 58. 2. Remove the socket bolt and locknut that secure the front of the upper armrest to the lower armrest (Detail
  • Invacare TDXSEAT | Owners Manual - Page 57
    "C" - FORMULA TRE SEATING SYSTEM WITH CONTOURA BACK Back Cane Cap Screws Locknut Arm Adapter Plate FIGURE 7.4 Adjusting Reclining Armrest Height Part No 1122170 57 Formula™ Powered Seating
  • Invacare TDXSEAT | Owners Manual - Page 58
    SECTION 8-SEAT SECTION 8-SEAT ƽ WARNING After ANY adjustments, repair or service and before use, make sure that all attaching hardware is tightened securely - otherwise 100° Upper Hex Screw Washer Lower Hex Screw FIGURE 8.1 Adjusting the Back Angle Formula™ Powered Seating 58 Part No 1122170
  • Invacare TDXSEAT | Owners Manual - Page 59
    Quick-Release Pin Tab Mounting Screw Quick-Release Pin Tab Washers Seat Positioning Strap Halves FIGURE 8.2 Replacing Seat Positioning Strap Part No 1122170 59 Formula™ Powered Seating
  • Invacare TDXSEAT | Owners Manual - Page 60
    or service and before use, make sure all attaching hardware is tightened securely - otherwise injury or damage may result. CAUTION If using a ventilator, verify headrest support does not Headrest Stop FIGURE 9.1 Removing/Installing/Adjusting Headrest Formula™ Powered Seating 60 Part No 1122170
  • Invacare TDXSEAT | Owners Manual - Page 61
    headrest to the desired depth by sliding the headrest towards the front of the wheelchair or towards the rear of the wheelchair. 3. If necessary, reposition the headrest to the desired position (headrest will Screw Headrest Part No 1122170 FIGURE 9.2 Replacing Headrest 61 Formula™ Powered Seating
  • Invacare TDXSEAT | Owners Manual - Page 62
    power to the wheelchair is OFF before performing these procedures. The use of rubber gloves is recommended when working with batteries. Invacare strongly recommends that battery installation and battery replacement always be done by a qualified technician. After ANY adjustments, repair or service
  • Invacare TDXSEAT | Owners Manual - Page 63
    perform this procedure. • Wire Cutter 1. Remove the batteries from the wheelchair. Refer to Removing/Installing the Batteries From/Into the Wheelchair on page 66. 2. Cut the tie-wraps that secure the battery POSITIVE (+) battery post of the battery. Part No 1122170 63 Formula™ Powered Seating
  • Invacare TDXSEAT | Owners Manual - Page 64
    each battery terminal cover over top of each battery terminal. 12. Secure battery terminal covers in place with one tie-wrap. 13. Install batteries into wheelchair. Refer to Removing/Installing the Batteries From/Into the Wheelchair on page 66. Formula™ Powered Seating 64 Part No 1122170
  • Invacare TDXSEAT | Owners Manual - Page 65
    Screw NOTE: Only battery cable ends/brackets shown for clarity. Front Battery NEGATIVE (-) Battery Terminal/Post FIGURE 10.1 Replacing Batteries Part No 1122170 65 Formula™ Powered Seating
  • Invacare TDXSEAT | Owners Manual - Page 66
    -ELEVATE BATTERIES Cleaning Battery Terminals ƽ WARNING Most batteries are not sold with instructions. However, warnings are frequently noted on the cell caps. Read them carefully of battery door. 6. Remove the battery door from the front of wheelchair. Formula™ Powered Seating 66 Part No 1122170
  • Invacare TDXSEAT | Owners Manual - Page 67
    support bracket to the battery box. B. Remove the battery support bracket. 8. Disconnect the controller from the batteries at the rear of the wheelchair (STEP 6) DETAIL "B" - WHEELCHAIRS WITH TRBKTS Thumb Screws (STEP Tray Locknuts (STEP 7) Medium Battery Support Bracket (STEP 7) FIGURE 10.2
  • Invacare TDXSEAT | Owners Manual - Page 68
    it from the battery box. NOTE: Battery cables not shown for clarity. REAR OF WHEELCHAIR Disconnect Batteries from Controller Here Rear Battery Battery Box YELLOW Battery Connector Tab BLACK Battery the Battery Door and Rear Shroud on page 69. Formula™ Powered Seating 68 Part No 1122170
  • Invacare TDXSEAT | Owners Manual - Page 69
    box. NOTE: For this procedure, refer to FIGURE 10.2 on page 67. 1. Wheelchairs with TRBKTS Only - Install the battery support bracket onto the battery box (Detail "B" of FIGURE 10.2 on page 67) using the otherwise serious personal injury may result. Part No 1122170 69 Formula™ Powered Seating
  • Invacare TDXSEAT | Owners Manual - Page 70
    or clamps. ALWAYS use the recharging plug located on the back of the joystick. DO NOT attempt to recharge the batteries and operate the power wheelchair at the same time. During use and charging, unsealed batteries will vent hydrogen gas which is explosive in the right concentration with air
  • Invacare TDXSEAT | Owners Manual - Page 71
    immediately. Contact an Invacare dealer. NOTE: If performing the charging procedures independently, READ and CAREFULLY follow the individual instructions for each charger (supplied or purchased). NOTE: If charging instructions are not supplied, consult a qualified service technician for proper
  • Invacare TDXSEAT | Owners Manual - Page 72
    SECTION 10-ELEVATE BATTERIES MPJ+ Joystick Charger/Programming Port (On Front of Joystick) FIGURE 10.6 Charging Batteries Formula™ Powered Seating 72 Part No 1122170
  • Invacare TDXSEAT | Owners Manual - Page 73
    may occur. Before adjusting, repairing or servicing the seating system, ALWAYS turn the wheelchair power OFF, otherwise, injury or damage may may have additional clearance requirements. ALWAYS refer to the ventilator's instructions prior to installation for any additional requirements. NOTE: For
  • Invacare TDXSEAT | Owners Manual - Page 74
    the ventilator battery harness connector to the ventilator battery harness extension (Detail "A"). 8. Wheelchairs with recline systems only - Perform the following steps: A. Test the recline function Tray FIGURE 1 Preparing the Ventilator Tray for Use Formula™ Powered Seating 74 Part No 1122170
  • Invacare TDXSEAT | Owners Manual - Page 75
    'S CONTROL, AND SUCH EVALUATION WILL BE SOLELY DETERMINED BY INVACARE. THE WARRANTY SHALL NOT APPLY TO PROBLEMS ARISING FROM NORMAL WEAR AND TEAR OR FAILURE TO ADHERE TO THE PRODUCT INSTRUCTIONS. A CHANGE IN OPERATING NOISE, PARTICULARLY RELATIVE TO MOTORS AND GEARBOXES DOES NOT CONSTITUTE A FAILURE
  • Invacare TDXSEAT | Owners Manual - Page 76
    BY INVACARE. THE WARRANTY SHALL NOT APPLY TO PROBLEMS ARISING FROM NORMAL WEAR AND TEAR OR FAILURE TO ADHERE TO THE PRODUCT INSTRUCTIONS. /PROVINCIAL LAWS AND REQUIREMENTS. Invacare Corporation USA One Invacare Way Elyria, Ohio USA 44036-2125 800-333-6900 www.invacare.com Canada 570 Matheson Blvd
  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76

Owner’s Operator and Maintenance Manual
DEALER:
This manual MUST be given to
the user of the product.
USER:
BEFORE using this product, read this
manual and save for future reference.
For more information regarding
Invacare products, parts, and services,
please visit www.invacare.com
Formula™
Powered Seating
Tilt
Recline
Elevate
Tilt/Recline
Tilt/Elevate
Tilt/Recline/Elevate