Sanyo MPR-414F Instruction Manual

Sanyo MPR-414F - Commercial Solutions Refrigerator Manual

Sanyo MPR-414F manual content summary:

  • Sanyo MPR-414F | Instruction Manual - Page 1
    INSTRUCTION MANUAL Pharmaceutical Refrigerator with Freezer MPR-414F MPR-414FS MPR-414F 32
  • Sanyo MPR-414F | Instruction Manual - Page 2
    this manual is unclear or if there are any inaccuracies. SANYO Electric Biomedical Co., Ltd. All rights reserved. Printed in Japan. CONTENTS PRECAUTIONS FOR SAFE OPERATION CAUTIONS FOR USAGE ENVIRONMENTAL CONDITIONS REFRIGERATOR COMPONENTS INSTALLAITON START-UP OF UNIT OPERATING INSTRUCTIONS ALARMS
  • Sanyo MPR-414F | Instruction Manual - Page 3
    . Symbol shows; ç ˚ this symbol means caution. ç this symbol means an action is prohibited. ç ˔ this symbol means an instruction must be followed. Be sure to keep this manual in a place accessible to users of this unit. < Label on the unit > This mark is labeled on the cover in which the electrical
  • Sanyo MPR-414F | Instruction Manual - Page 4
    the unit is exposed to rain water. Only qualified engineers or service personnel should install the unit. The installation by unqualified personnel may this unit. This may lead to damage to the inner components or electric parts. If this unit is to be used for storing poisons, radioactive material
  • Sanyo MPR-414F | Instruction Manual - Page 5
    doors to prevent accidents such as suffocation. Make sure to prepare a safety check sheet when you request any repair or maintenance for the safety of service personnel. 36 4
  • Sanyo MPR-414F | Instruction Manual - Page 6
    or cord. Do not use the supply plug if its cord is loose. This may cause fire or electric shock. Do not touch any electrical parts such as the power supply plug or any switches with a wet hand. This may cause electric shock. Do not climb onto the unit or do
  • Sanyo MPR-414F | Instruction Manual - Page 7
    . 8. Always open and close the door gently. Rough operation may lead fall down of stocked items, incomplete closing, or damage of door gasket. 9. In the refrigerator compartment, put the items on the shelves and do not contact the items with wall. 10. The cooling pipe is routed in the rear of
  • Sanyo MPR-414F | Instruction Manual - Page 8
    these materials. Especially never use any volatile solvent to clean the plastic or rubber parts. When a neutral dishwashing detergent is used to clean the unit, wipe it up thoroughly with a cloth soaked in clean water. 14. If condensation forms on the front glass or frame surface, wipe it off with
  • Sanyo MPR-414F | Instruction Manual - Page 9
    REFRIGERATOR COMPONENTS 13 15 8 Refrigerator 14 7 63 4 5 1 2 16 17 8 9 10 Freezer 11 4 12 MPR-414F with drawers (optional component) 40 8
  • Sanyo MPR-414F | Instruction Manual - Page 10
    REFRIGERATOR COMPONENTS 1. Glass window: Water can sometimes condense on the glass in areas of high humidity. Wipe off the condensation with a dry soft cloth. (MPR-414FS is used to alarm the abnormality to the remote location. See page 14 for the details. 12. Leveling feet: Use these bolts to adjust
  • Sanyo MPR-414F | Instruction Manual - Page 11
    key (REF./FREEZ.): By pressing this key, chamber temperature display of refrigerator or freezer and each setting is selected. 3. Set key (SET): 5 seconds in the temperature display mode. See "Key lock operation" on page 14. 5. Numerical value shift key ( ): Pressing this key in the setting mode
  • Sanyo MPR-414F | Instruction Manual - Page 12
    away from heat generating sources Avoid installing the unit near heat-emitting appliances such as gas ranges or stoves. Heat can cause inefficient refrigeration. 4. A location with a sturdy and level floor Install the unit on a sturdy floor to avoid vibration and noise. Placing the unit on an
  • Sanyo MPR-414F | Instruction Manual - Page 13
    are dirty, clean them with a diluted neutral dishwashing detergent. (Undiluted detergent can damage the plastic components. For the dilution, refer to the instruction of the detergent.) After the cleaning with the diluted detergent, always wipe it off with a wet cloth. Then wipe off the panels with
  • Sanyo MPR-414F | Instruction Manual - Page 14
    INSTRUCTIONS Temperature setting Table 1 shows the basic operation method. Perform key operation in the sequence indicated in the table. The example in the table is based on the assumption that the refrigerator temperature setting for refrigerator ranges between 2 and 14ˆ. The stored items
  • Sanyo MPR-414F | Instruction Manual - Page 15
    OPERATING INSTRUCTIONS Key lock operation This unit incorporates digit to 1 with the numerical value shift key. 3 Press the set key. The current refrigerator or freezer temperature is displayed. The first digit of the temperature display blinks. Pressing the switch is OFF. Terminal 46 14
  • Sanyo MPR-414F | Instruction Manual - Page 16
    OPERATING INSTRUCTIONS Alarm temperature setting The high temperature and low temperature at which the alarm will be activated are effective for freezer and refrigerator. The following shows the outline of the alarm temperature setting. Display F01 F02 F03 F04 Mode High temp. alarm setting Low
  • Sanyo MPR-414F | Instruction Manual - Page 17
    OPERATING INSTRUCTIONS Table 4 Operation Key operated Indication after operation 1 ----- The current refrigerator temperature is displayed. 2 Press and hold the numerical value shift key for about 5 seconds. The first digit of the temperature display blinks. 3 Set the first
  • Sanyo MPR-414F | Instruction Manual - Page 18
    case of 0oC or lower) Remote alarm is activated after a delay of 15 minutes. (No delay in the case of 0oC or lower) Refrigerator comp. ----- turns OFF. Reset at about 6oC. ----- Remote alarm is activated. Intermittent tone after a delay of 2 ----- minutes. ----- Setting mode is canceled
  • Sanyo MPR-414F | Instruction Manual - Page 19
    SETTING OF ALARM RESUME TIME The alarm buzzer and remote alarm are silenced by pressing BUZZER key on the control panel during alarm condition. The buzzer and remote alarm will be activated again after certain suspension if the alarm condition is continued. The suspension time can be set by
  • Sanyo MPR-414F | Instruction Manual - Page 20
    effect on the chamber temperature. 2. Forced defrost When the ambient humidity is high, or a large amount of damp product is being stored inside the refrigerator, there is a possibility that cycle defrost may not be enough to remove all of the frost on the evaporator. In this case, a forced defrost
  • Sanyo MPR-414F | Instruction Manual - Page 21
    detergent can damage the plastic components. For the dilution, refer to the instruction of the detergent.) After the cleaning with the diluted detergent, always wipe it off with a wet cloth. Then wipe off the cabinet or accessories with a dry cloth. • Never pour water onto or into the unit. Doing
  • Sanyo MPR-414F | Instruction Manual - Page 22
    . For the proper usage of temperature recorder, refer to an instruction manual included with the recorder. The available recorder and its attachment location -4014LH) Recorder fixture(MPR-S30) Or Recorder for both refrigerator & freezer (MTR-G3504) Recorder fixture(MPR-S7) WARING Always disconnect
  • Sanyo MPR-414F | Instruction Manual - Page 23
    the access port on the back of the unit. 4. Take the recorder sensor into the refrigerator chamber through the access port. 5. Fix the recorder sensor on the right wall of the In this case, a dedicated recorder fixture (MPR-S30) is needed. For the installation, refer to "For freezer (MTR-4014LH)"
  • Sanyo MPR-414F | Instruction Manual - Page 24
    of the cabinet even if the recorder for refrigerator (MTR-0620LH) is not installed there. For both refrigerator and freezer (MTR-G3504) Cover panel 1. (MTR-G3504) to the recorder fixture (MPR-S7: optional component) by following the instruction shown in the procedure enclosed with the fixture
  • Sanyo MPR-414F | Instruction Manual - Page 25
    TROUBLESHOOTING If the unit malfunctions, check out the following before calling for service. In the case of no refrigeration or poor refrigeration, transfer the stored items to another refrigerator or freezer before checking out. If nothing operates even when plug in 1. The unit is not connected to
  • Sanyo MPR-414F | Instruction Manual - Page 26
    Pharmaceutical Refrigerator with Freezer Model MPR-414F MPR-414FS External dimensions W800 x D600 x H1805 (mm) Internal dimensions W720 x D495x H1425 (mm) (Refrigerator) W317 x D440x H576 (mm) (Freezer) Effective capacity 340 L (Refrigerator Refrigerator) Weight 126 kg 119 kg Accessories
  • Sanyo MPR-414F | Instruction Manual - Page 27
    .; -5oC to +35oC, No load) Freezer: -20oC to -30oC (Ambient temp.; -5oC to +30oC, No load) Noise level 39 dB (A scale) Maximum pressure 1.46 MPa (Refrigerator), 2.18 MPa (Freezer) Rated voltage AC 110 V AC 115 V AC 220 V AC 220 V AC 230 V AC 240 V Rated frequency 60 Hz 60 Hz 50 Hz
  • Sanyo MPR-414F | Instruction Manual - Page 28
  • Sanyo MPR-414F | Instruction Manual - Page 29
    1
  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29

32
MPR-414F
MPR-414FS
Pharmaceutical Refrigerator
with Freezer
INSTRUCTION MANUAL
MPR-414F