Fluke 190-502 Manual

Fluke 190-502 Manual

Fluke 190-502 manual content summary:

  • Fluke 190-502 | Manual - Page 1
    ScopeMeter 190 Series II Fluke 190-062, -102, -104, -202, -204, -502 Users Manual May 2011 © 2011 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective companies. MyFlukeStore Shop for Fluke products
  • Fluke 190-502 | Manual - Page 2
    SCOPE 1 2 METER 1 RECORDER 1 SCOPE METER 2 CURSOR ZOOM REPLAY CURSOR ZOOM 3 3 REPLAY 3 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 3
    user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price. Fluke Fluke authorized service center. Fluke
  • Fluke 190-502 | Manual - Page 4
    To locate an authorized service center, visit us on the World Wide Web: http://www.fluke.com or call Fluke using any of the phone numbers listed below: +1-888-993-5853 in U.S.A. and Canada +31-40-2675200 in Europe +1-425-446-5500 from other countries MyFlukeStore Shop for Fluke products online at
  • Fluke 190-502 | Manual - Page 5
    14 Key Illumination...14 Input Connections ...15 Making Input Connections 15 Adjusting the Probe Type Settings 16 Selecting an Input Channel 17 i MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 6
    ScopeMeter 190 Series II Users Manual Displaying an Unknown Signal with Connect-and-View 18 Making Automatic Scope Measurements 19 Freezing the Screen ...20 Using Average, Persistence and 60 Triggering on Edges ...61 ii MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 7
    Contents (continued) Triggering on External Waveforms (models 190-xx2 64 Triggering on Video Signals 65 Triggering on Pulses ...67 5 Using Memory and PC ...95 Charging the Batteries 96 Replacing the Battery Pack 97 iii MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 8
    ScopeMeter 190 Series II Users Manual Calibrating the Voltage Probes 99 Displaying Version and Calibration Information 101 Displaying Battery Information 101 Parts and Accessories 102 Troubleshooting ...107 8 Specifications ...109 Introduction ...109 Oscilloscope ...110 Automatic Scope
  • Fluke 190-502 | Manual - Page 9
    instructions in this manual apply to all ScopeMeter 190 Series II versions (hereafter referred to as the instrument or as the test tool). The versions are listed below. The version 190 on the versions 190-x04. Introduction Version 190-062 190-102 190-104 190-202 190-204 190-502 Description Two 60
  • Fluke 190-502 | Manual - Page 10
    Fluke 190 Series II Users Manual Unpacking the Test Tool Kit The following items are 12 (4x) c (2x) c 6 CAT II 1000V CAT III 1000V CAT IV 600V 13 14 12345 - 12345 - 12345 15 Figure 1. ScopeMeter Test Tool Kit 2 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 11
    sheet + CD ROM with Users Manual (multi-language) and FlukeView ScopeMeter Software for Windows demo package (with restricted functionality) 6 Shipment box (basic version only) Unpacking the Test Tool Kit The Fluke 190-062, 190-102, 190-104, 190-202 and 190-204 include also: # Description
  • Fluke 190-502 | Manual - Page 12
    Fluke 190 Series II Users Manual The Fluke 190-502 includes also: # Description 11 Voltage Probe Set (red) 12 Voltage (one red, one black). Fluke 190-xxx /S versions include also the following items (SCC290 kit): # Description 14 FlukeView ScopeMeter Software for Windows activation key (converts
  • Fluke 190-502 | Manual - Page 13
    , appear throughout the manual. A "Warning" identifies conditions and actions that pose hazard(s) to the user. A "Caution" Fluke's website for recycling information. Conforms to relevant Australian standards Conformité Européenne Alternating Current RoHS China 5 MyFlukeStore Shop for Fluke
  • Fluke 190-502 | Manual - Page 14
    Fluke 190 Series II Users Manual Warning To avoid electrical shock or fire: • Use only the Fluke power supply, Model BC190 (Power and adapters supplied with the product, or indicated by Fluke as suitable for the Fluke 190 Series II ScopeMeter series. • Before use, inspect voltage probes, test leads
  • Fluke 190-502 | Manual - Page 15
    directly transmitted to the product. • Do not use exposed metal BNC or banana plug connectors. Fluke offers cables with plastic, safety designed BNC connectors suitable for the ScopeMeter product, see Chapter 7 'Optional accessories'. • Do not insert metal objects into connectors. • Use the product
  • Fluke 190-502 | Manual - Page 16
    floating' are used in this manual to indicate a measurement in which the product input BNC is connected to a voltage different from earth ground. The isolated input connectors have no exposed metal and are fully insulated to protect against electrical shock. Fluke 190-502: The BNC jacks can
  • Fluke 190-502 | Manual - Page 17
    the battery pack. • The battery pack needs to be charged before use. Use only Fluke approved power adapters to charge the battery pack. Refer to Fluke's safety instructions and Users Manual for proper charging instructions. • Do not leave a battery on prolonged charge when not in use. • The battery
  • Fluke 190-502 | Manual - Page 18
    Fluke 190 Series II Users Manual • Do not short-circuit a battery pack. Do not keep battery packs in a Dispose of properly: do not dispose of the battery as unsorted municipal waste. Go to Fluke's website for recycling information. • Dispose in discharged condition and cover the battery terminals
  • Fluke 190-502 | Manual - Page 19
    basic operations. Powering the Test Tool Follow the procedure (steps 1 through 3) in Figure 2 to power the test tool from a standard ac outlet. See Chapter 6 for instructions on using battery power. Turn the test tool on with the on/off key. The test tool powers up in its last setup configuration
  • Fluke 190-502 | Manual - Page 20
    Fluke 190 Series II Users Manual Resetting the Test Tool If you want to reset the test tool to the factory settings, do the following: 1 Turn the test tool off. 2 USER Press and hold the USER key. 3 Press and release. The test tool turns on, and you should hear a double beep, indicating the
  • Fluke 190-502 | Manual - Page 21
    . 2 F4 Open the Waveform Options menu. This menu is displayed at the bottom of the screen. Actual settings are shown on a yellow background. 3a Figure 4. Basic Navigation 3a Use the blue moment press F4 (CLOSE) 13 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 22
    Fluke 190 Series II Users Manual Hiding Key Labels and Menus You can close a 'Tips' section 'Setting the Display AUTO-Off timer '. Off: in all other situations On: Measurements are stopped, the screen is frozen. (HOLD) Off: Measurements are running. (RUN) On: The range key, the move up/down key,
  • Fluke 190-502 | Manual - Page 23
    (models 190-xx4), or two safety BNC jack inputs and two safety 4-mm banana jack inputs (models 190-xx2). maximally benefit from having independently isolated floating inputs and to avoid problems caused by improper use, read Chapter 6: "Tips". − Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 24
    Fluke 190 Series II Users Manual Adjusting the Probe Type Settings To obtain correct measurement results the test tool probe type factor Current and Temp: select the current probe or temperature probe sensitivity 16 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 25
    pressed keys are illuminated now. The MOVE UP/DOWN key and the RANGE mV/V key applies to all involved input channels. 17 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 26
    Fluke 190 Series II Users Manual Displaying an Unknown Signal with Connect-and-View™ The Connect -and-View feature when the test tool is in MANUAL mode, do the following: 1 MANUAL Perform an Auto Set. AUTO appears at AUTO the top right of the screen, the key illumination is off. The bottom line shows
  • Fluke 190-502 | Manual - Page 27
    Scope Measurements 5 Select the Hz measurement. ENTER Observe that the top left of the screen displays the Hz measurement. (See Figure 8.) To choose also a Peak-Peak measurement for ENTER to the measurements field. 19 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 28
    Fluke 190 Series II Users Manual 5 Open the PEAK menu. ENTER 6 Select the Peak-Peak ENTER measurement. Figure 8 shows an example of the screen with two readings. The character size will be reduced when more then two readings are on. Freezing the Screen You can freeze the screen (all readings
  • Fluke 190-502 | Manual - Page 29
    deviations in a waveform just distort the averaged wave shape, and do not show up on screen clearly. When a signal really changes, for instance when you probe around, it takes quite in video show up on screen instantly. 21 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 30
    Fluke 190 Series II Users Manual Using Persistence, Envelope and Dot-Join to Display Waveforms You can use Persistence to observe dynamic signals to see the upper and lower boundaries of dynamic waveforms (envelope mode). 22 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 31
    . When you select the 2 mV/div range Glitch Detect will automatically be turned Off. In the 2 mV/div range you can set Glitch Detect On manually. 23 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 32
    Fluke 190 Series II Users Manual Suppressing High Frequency Noise Switching the glitch detection off (Glitch: Off) will suppress the high update rate. Normal - optimal trace update rate and zoom range combination 24 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 33
    AC-Coupling After a reset, the test tool is dc-coupled so that ac and dc voltages appear on the screen. Use ac-coupling when you wish to observe a small ac signal that rides on a dc signal. To line below the waveform. 25 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 34
    Fluke 190 Series II Users Manual Variable Input Sensitivity The variable input sensitivity allows 5 Select and accept Variable. ENTER 6 F4 Exit the menu. At the bottom left of the screen the text A Var is displayed. Selecting Variable will turn off cursors and automatic input ranging. 7
  • Fluke 190-502 | Manual - Page 35
    , use the average function or turn off Display Glitches. 3 Jump to Waveform: and Select ENTER Mathematics... to open the Mathematics menu. 27 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 36
    Fluke 190 Series II Users Manual 4 Select Function: +, -, x or XY- ENTER mode. 5 Select the first waveform: ENTER Source 1: A, but also in more leakage. Ensure that the entire waveform amplitude remains on the screen. To use the Spectrum function, do the following: 1 SCOPE Display the SCOPE
  • Fluke 190-502 | Manual - Page 37
    , which can result in aliasing, or too fast, which results in less than one signal period on the screen. 7 F1 8 F2 9 F3 10 F4 Perform a spectrum analysis on trace A, B, C or D. Set off/on (toggle function). 29 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 38
    Fluke 190 Series II Users Manual Figure 11. Spectrum measurement Comparing Waveforms You can display a fixed reference waveform with the Waveform field and ENTER select Reference... to open the WAVEFORM REFERENCE menu. 30 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 39
    from memory and use it as a reference waveform, refer also to Chapter 5 Recalling Screens with Associated Setups. Example of reference waveform with an additional envelope of ±2 pixels: range/div. momentary waveform. 31 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 40
    Fluke 190 Series II Users Manual Pass - Fail Testing You can use a reference waveform as a test template for the actual waveform. If at least one sample of a waveform is outside the test template, the failed or passed scope screen will be stored. Up to 100 screens can be stored. If the memory is
  • Fluke 190-502 | Manual - Page 41
    Meter Making Automatic Meter Measurements (for models 190-xx4) Making Automatic Meter Measurements (for models 190-xx4) The test tool offers a You will see a screen like in Figure 12. 2 F1 Open the Reading .. menu. Figure 12. Meter Screen 33 MyFlukeStore Shop for Fluke products online at:
  • Fluke 190-502 | Manual - Page 42
    Fluke 190 Series II Users Manual Making Relative Meter Measurements A relative measurement displays the present measurement result relative to a defined activity (voltage, temperature) in relation to a known good value. 34 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 43
    the Scope and Meter Making Multimeter Measurements (for models 190-xx2) Making Multimeter Measurements (for models 190-xx2) The screen displays the numeric readings of the measurements on the 14. Meter Connections 35 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 44
    Fluke 190 Series II Users Manual Measuring Resistance Values To measure a resistance, do the following: 1 Connect the red and black Highlight Ohms. 5 ENTER Select Ohms measurement. Figure 15. Resistor Value Readings 36 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 45
    resolution. The next example explains a typical current measurement in Meter mode. Warning Carefully read the instructions about the current probe you are using. To set up the test tool, do the following: 1 Connect a current probe (e.g. Fluke Measurements (for models 190-xx2) Figure
  • Fluke 190-502 | Manual - Page 46
    Fluke 190 Series II Users Manual 6 Observe the sensitivity of the current probe. Highlight the corresponding sensitivity in the menu, e.g. 1 mV/A. 7 ENTER Accept the current measurement. Now, you will see a screen like in Figure 17 Figure 17. Ampere Measurement Readings Selecting Auto/
  • Fluke 190-502 | Manual - Page 47
    the METER key labels. 1 Using the Scope and Meter Making Multimeter Measurements (for models 190-xx2) Now the large reading is the actual input value minus the stored reference value. in relation to a known good value. 39 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 48
    Fluke 190 Series II Users Manual Adjusting the reference value To adjust the reference value, do the following: 5 F3 Display the Adjust step 6 and step 7 until finished. 8 ENTER Enter the new reference value. 40 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 49
    menu, do the following: 1 RECORDER Open the recorder main menu. (See Figure 19). Figure 19. Recorder Main Menu Trendplot Meter is only present in models 190-xx2. 41 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 50
    Fluke 190 Series II Users Manual Plotting Measurements Over Time (TrendPlot™) Use the TrendPlot function time from start appears at the bottom of the screen. The present reading appears on top of the screen. (See Figure 20.) 42 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.
  • Fluke 190-502 | Manual - Page 51
    tool is in automatic mode, automatic vertical scaling is used to fit the TrendPlot graph on the screen. 5 F1 Set RECORDER to STOP to freeze the recorder function. 6 F1 Set RECORDER to RUN Figure 20. TrendPlot Reading 43 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 52
    Fluke 190 Series II Users Manual Displaying Recorded Data When in normal view (NORMAL), only the twelve most recently recorded divisions are displayed on screen. All previous recordings are stored in memory. VIEW ALL shows all data in memory: 7 F3 Display an overview of the full waveform.
  • Fluke 190-502 | Manual - Page 53
    . • The status at the bottom of the screen which includes the time/div setting as well as the total timespan that fits the memory. Note For accurate recordings it is advised to let the instrument first warm up for five minutes. 45 MyFlukeStore Shop for Fluke products online at: www. .com 1.877
  • Fluke 190-502 | Manual - Page 54
    Fluke 190 Series II Users Manual Displaying Recorded Data In Normal view, the samples that roll off the screen are stored in deep of memory. In View All mode, the complete memory contents are displayed on the screen. 4 F3 Press to toggle between VIEW ALL (overview of all recorded samples) and
  • Fluke 190-502 | Manual - Page 55
    xx4) or on Ext. (models 190-xx2) to open the START SINGLE SWEEP ON TRIGGERING or the START SINGLE SWEEP ON EXT. menu. 8 Select one of the Conditions: ENTER and accept the selection. For external triggering (190-xx2) continue at step 9. 47 MyFlukeStore Shop for Fluke products online at: www
  • Fluke 190-502 | Manual - Page 56
    Fluke 190 Series II Users Manual 9 Select the desired trigger slope ENTER (Slope:) and jump to Level samples are continuously saved in deep memory. The last twelve recorded divisions are displayed on the screen. Use View All to display the full memory contents. Note To learn more about the Single
  • Fluke 190-502 | Manual - Page 57
    : • First replay the last screens to find the screen of special interest. • Then zoom Scope Screens When you are in scope mode, the test tool automatically stores the 100 most recent screens. When in time" by stepping through the stored screens to find the screen of your interest. This feature lets
  • Fluke 190-502 | Manual - Page 58
    Fluke 190 Series II Users Manual Replaying Step-by-Step To step through the last scope screens, do the following: 1 REPLAY From scope mode, open the REPLAY menu. Observe that the trace is frozen and that REPLAY appears at the top of the screen (see Figure 23). 2 F1 Step through the previous
  • Fluke 190-502 | Manual - Page 59
    replay. 3 Using Replay, Zoom and Cursors Replaying the 100 Most Recent Scope Screens Turning Off the Replay Function 4 F4 Turn off REPLAY. Capturing 100 Intermittents Automatically "Triggering on Waveforms". 51 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 60
    Fluke 190 Series II Users Manual Zooming in on a Waveform To obtain a more detailed view of a waveform, you can zoom in on a waveform using the ZOOM function. To zoom in on a waveform, do the following: 1 ZOOM Display the ZOOM key labels. ZOOM appears at the top of the screen the screen, you
  • Fluke 190-502 | Manual - Page 61
    . Voltage Measurement with Cursors The screen shows the voltage difference between the two cursors and the voltage at the cursors. (See Figure 25.) Use horizontal cursors to measure the amplitude, high or low value, or overshoot of a waveform. 53 MyFlukeStore Shop for Fluke products online at: www
  • Fluke 190-502 | Manual - Page 62
    Fluke 190 Series II Users Manual Using Vertical Cursors on a Waveform To use the cursors for a Measurement with Cursors 8 Move the right cursor to the desired position on the waveform. The screen shows the time difference between the cursors and the voltage difference between the two markers. (
  • Fluke 190-502 | Manual - Page 63
    a spectrum, do the following: 1 CURSOR From Spectrum measurement display the cursor key label. 2 Move the cursor and observe the readings at the top of the screen. 55 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 64
    Fluke 190 Series II Users Manual Making Rise Time Measurements To measure rise time, do the multiple traces select the required trace A, B, C, D or M (if a math function is active). Select MANUAL or AUTO (this automatically does steps 5 to 7). Move the upper cursor to 100% of the trace height.
  • Fluke 190-502 | Manual - Page 65
    of the signal. (See next section.) • For dedicated applications, use one of the three manual trigger functions: • Edge triggering • Video triggering • Pulse Width triggering • External Triggering (models 190-xx2 only) 57 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 66
    Fluke 190 series II Users Manual Setting Trigger Level and Slope The Connect-and-View™ function enables hands-off triggering to display complex unknown signals. When your test tool is in manual range, do the following: MANUAL AUTO Perform an auto set. AUTO appears at the top right of the screen
  • Fluke 190-502 | Manual - Page 67
    what caused the trigger. In case of a trigger delay, the status at the bottom of the screen will change. For example: 4 Triggering on Waveforms Using Trigger Delay or Pre-trigger When a valid point and waveform display. 59 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 68
    Fluke 190 series II Users Manual Automatic Trigger Options In the trigger menu, instructed to analyze low frequency components for automatic triggering: 4 Select > 1 HZ and return to the ENTER measurement screen. 3 Open the AUTOMATIC TRIGGER menu. ENTER 60 MyFlukeStore Shop for Fluke
  • Fluke 190-502 | Manual - Page 69
    edge triggering to obtain full manual trigger control. To trigger the screen even if there are no triggers. A trace always appears on the screen. mode if you want to update the screen only when valid triggers occur. When screen have adapted to allow further selection of specific edge trigger
  • Fluke 190-502 | Manual - Page 70
    Fluke 190 series II Users Manual Triggering on Noisy Waveforms To reduce jitter on the screen when triggering on noisy waveforms, 3 (page 61) again: 4 Select Single Shot. ENTER The word MANUAL appears at the top of the screen indicating that the test tool is waiting for a trigger. As soon as
  • Fluke 190-502 | Manual - Page 71
    6 Set NCycle to On ENTER Observe that the key labels at the bottom of the screen have been changed to allow further selection of specific N-Cycle trigger settings: 4 Triggering on 31. Figure 31. N-Cycle triggering 63 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 72
    Fluke 190 series II Users Manual Triggering on External Waveforms (models 190-xx2) Use external triggering when you TRIGGER (On Edges) key labels. Observe that the key labels at the bottom of the screen have been adapted to allow selection of two different external trigger levels: 0.12 V and
  • Fluke 190-502 | Manual - Page 73
    Non interlaced a scan rate selection menu will open. Trigger level and slope are now fixed. Observe that the key labels at the bottom of the screen have been changed to allow further selection of specific video trigger settings. 65 MyFlukeStore Shop for
  • Fluke 190-502 | Manual - Page 74
    Fluke 190 series II Users Manual Triggering on Video Frames Use FIELD 1 or FIELD 2 following: 7 F1 Choose FIELD 2. The signal part of the even field is displayed on the screen. Triggering on Video Lines Use ALL LINES to trigger on all line synchronization pulses (horizontal synchronization).
  • Fluke 190-502 | Manual - Page 75
    on narrow pulses only. Observe that the trigger key labels at the bottom of the screen have been adapted to set the pulse conditions: To set the pulse width to 5 5 ms are now displayed on the screen. (See Figure 33). 67 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 76
    Fluke 190 series II Users Manual Tip The test tool stores all triggered screens in the replay memory. For example, if you setup your triggering for glitches, Width on A... to ENTER open the TRIGGER ON PULSE WIDTH menu. 68 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 77
    more than a selectable time in duration. Observe that the trigger menu at the bottom of the screen has been adapted to set the pulse condition: To set the pulse width to 110 ms, continue 34. Triggering on Missing Pulses 69 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 78
    Fluke 190 series II Users Manual 70 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 79
    channels and are covered with dust covers when not in use. USB stick Mini USB Figure 35. Test Tool USB Connections 71 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 80
    Fluke 190 series II Users Manual Saving and Recalling You can: • Save screens and setups to internal memory, and recall them again from memory. The test tool has 15 'screen and setup' memories, 2 'record and setup' memories, and 1 screen image memory. See also Table 1. • Save up to 256 screens and
  • Fluke 190-502 | Manual - Page 81
    OK SAVE are already selected. To modify the name for this particular Screen+Setup or to modify the default name see below 'Editing Names'. 5 ENTER Save the Screen+setup. To resume your measurements press HOLD RUN . 73 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 82
    Fluke 190 series II Users Manual All memories in use If no free memory locations are again. If you want to overwrite the oldest data set, - press F4 . Editing names To name the screen+setup according to your own preferences, continue from step 4 as follows: 5 F1 Open the EDIT NAME menu
  • Fluke 190-502 | Manual - Page 83
    the various test tool modes. To save a TrendPlot press STOP first. Saving Screens in .bmp Format (Print Screen) To save a screen in bitmap (.bmp) format, do the following: 1 Display the SAVE key set, - press F4 . 75 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 84
    Fluke 190 series II Users Manual Deleting Screens with Associated Setups To delete a screen and associated setup, do the following: 1 Display the SAVE key labels. SAVE Recalling Screens with Associated Setups To recall a screen+setup, do the following: 1 Display the SAVE key labels. SAVE 2
  • Fluke 190-502 | Manual - Page 85
    to the filename field. 5 Highlight a file . 6 F2 View the screen, and open the viewer. 7 8 F3 Scroll through all stored screens. Print screen, saves the screen to USB device (if connected) or internal memory. 77 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 86
    Fluke 190 series II Users Manual 9 F4 Exit the View mode. Note: In the VIEW mode the replay screens of a saved 'record+setup' cannot be viewed! Only the screen at the instant of saving can be reviewed in this way. To see all replay screens recall them from memory using the RECALL option.
  • Fluke 190-502 | Manual - Page 87
    Copying-Moving Stored Screens and Setup Files You can copy or move a file from internal memory to a USB device or from a USB device to be copied or moved, Or Select all files 7 ENTER Copy or delete the selected files. 79 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 88
    Fluke 190 series II Users Manual Using FlukeView® With the FlukeView® software you can upload waveform data and screen bitmaps to your PC or ordered using ordering code SW90W. For use with the Fluke Series II ScopeMeter test tools FlukeView® ScopeMeter release V5.0 or higher is required. − The test
  • Fluke 190-502 | Manual - Page 89
    Accessories The following illustrations show the use of the standard accessories such as voltage probes, test leads, and the various clips. 81 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 90
    Fluke 190 Series II Users Manual A Figure 37. HF Voltage Probe Connection Using Ground Spring Warning To avoid electrical shock or fire, do not connect the ground spring to voltages higher than 30 Vrms from earth ground. Figure 38. HF Voltage Probe Connection Using Probe Tip to BNC Adapter (190-
  • Fluke 190-502 | Manual - Page 91
    having four independent instruments. The advantages of having independently floating isolated inputs are: • It allows simultaneous measurement of independently floating signals. 83 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 92
    Fluke 190 Series II Users Manual • Additional safety. Since the commons are not directly connected, the chance of causing short circuit when measuring multiple present on the ground ring near the probe tip, see Figure 41. 84 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 93
    BUS ANALOG GROUND DIGITAL GROUND D A Figure 43. Correct connection of reference leads Figure 42. Parasitic capacitance between analog and digital reference 85 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 94
    Fluke 190 Series II Users Manual ANALOG INPUT DIGITAL CONTROLLER ANALOG GROUND DIGITAL GROUND D angle while placed on a table. The typical position is shown in Figure 45. Kensington® lock support Power adapter input Figure 45. Using the Tilt Stand Note An optional Hanging Hook, ordering code HH290
  • Fluke 190-502 | Manual - Page 95
    test tool. The figure below shows how to attach the strap correctly to the test tool. Figure 46. Fixing the Hangstrap 87 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 96
    Fluke 190 Series II Users Manual Resetting the Test Tool If you want to reset the test tool to the factory settings, without clearing the memories, do the following: 1 Turn the test tool off. 2 USER Press and hold. 3 Press and release. The test tool turns on, and you should hear a double
  • Fluke 190-502 | Manual - Page 97
    following: 1 USER Display the USER key labels. 2 F4 3 4 Enable the arrow keys for manual contrast and backlight adjustment. Adjust the contrast of the screen. Change the 8 'Specifications', section 'Miscellaneous'. 89 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 98
    Fluke 190 Series II Users Manual Changing Date and Time The test tool has a date and time clock. For example, to change the date to 19 April, 2012, do the following: 1 USER Display the USER key labels. 2 F1 Open the USER OPTIONS menu. 3 Open the DATE ADJUST menu. ENTER 4 Choose 2012,
  • Fluke 190-502 | Manual - Page 99
    can set the power shutdown time to 5 minutes as following: 1 USER Display the USER key labels. 2 F1 Open the USER OPTIONS menu. 6 Tips Saving Battery Life Setting the Display AUTO- the selected time is elapsed. 91 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 100
    Fluke 190 Series II Users Manual To turn on the display instructed not to analyze low frequency signal components. However, when you measure frequencies lower than 15 Hz, the test tool must be instructed to analyze low frequency components for automatic triggering: 92 MyFlukeStore Shop for Fluke
  • Fluke 190-502 | Manual - Page 101
    can either be set to dc or left unchanged: 5 Select Unchanged. ENTER When you press the AUTO-MANUAL (auto set) key glitch capture can either be set to On or left unchanged: 6 Select Unchanged the Auto Set Options 93 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 102
    Fluke 190 Series II Users Manual 94 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 103
    this Chapter This chapter covers basic maintenance procedures that can be performed by the user. For complete service, disassembly, repair, and calibration information, see the Service Manual. (www.fluke.com) Warning • Have an approved technician repair the product. • Use only specified replacement
  • Fluke 190-502 | Manual - Page 104
    Fluke 190 Series II Users Manual Charging the Batteries At delivery, the Li-ion batteries may be empty and must be charged for 5 hours (with the test tool turned off) to reach full charge. When battery power is used, the battery indicator at the top of the screen informs you about the condition
  • Fluke 190-502 | Manual - Page 105
    Replacing the Battery Pack Warning Use only the Fluke BP290 (not recommended for 190-xx4) or BP291 for replacement! When no adapter power is supplied, Tool Replacing the Battery Pack Figure 48. Unlocking the Battery Cover 97 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 106
    Fluke 190 Series II Users Manual Figure 49. Removing the Battery Cover Figure 50. Removing the Battery 98 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 107
    Calibrating the Voltage Probes To meet full user specifications, you need to adjust the voltage probes for optimal response. The calibration ENTER Attenuation: 10:1. Probe Cal Ref. Figure 51. Adjusting Voltage Probes 99 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 108
    Fluke 190 Series II Users Manual A message appears asking you whether to start the 10:1 probe housing until a pure square wave is displayed. For instructions to access the trimmer screw in the probe housing see the probe instruction sheet. The test tool automatically calibrates itself to the
  • Fluke 190-502 | Manual - Page 109
    and calibration date: 1 USER Display the USER key labels. 2 F3 Open the VERSION & CALIBRATION screen. 7 Maintaining the by qualified personnel. Contact your local Fluke representative for recalibration. 4 F4 Return to the previous screen. 'Level' indicates the available battery capacity
  • Fluke 190-502 | Manual - Page 110
    Fluke 190 Series II Users Manual Parts and Accessories The following tables list the user-replaceable parts and specific optional accessories for the various test tool models. For more optional accessories, see www.fluke.com. To order replacement parts or additional accessories, contact your
  • Fluke 190-502 | Manual - Page 111
    Red or Blue or Gray or Green), designed for use with the Fluke 190-062,-102,-104,-202, and -204 test tool. The set includes the following items (not ) See Figure 1 on page 2 for item reference. See the VPS410 instruction sheet for voltage/CAT ratings. Replacement Set for Voltage Probe VPS410 The set
  • Fluke 190-502 | Manual - Page 112
    Fluke 190 Series II Users Manual Replacement Parts (continued) Voltage Probe Set, designed for use with the Fluke 190-502 test tool. The set Adapter See Figure 1 on page 2 for item reference. See the VPS510 instruction sheet for voltage/CAT ratings. Replacement Set for Voltage Probe VPS510 The set
  • Fluke 190-502 | Manual - Page 113
    two pieces, black) Li-ion battery (26 Wh, not recommended for models 190-xx4) Li-ion battery (52 Wh) Hangstrap 7 Maintaining the Test Tool into an operational version. • Hard Shell Carrying Case C290 FlukeView® ScopeMeter® Software for Windows® (full version) Hard Shell Carrying Case Ordering
  • Fluke 190-502 | Manual - Page 114
    Fluke 190 Series II Users Manual Optional accessories (continued) External Battery Charger, charges BP291 externally using BC190 High Working Voltage BNC (fully isolated). EBC290 VPS420-R HH290 PM9091 PM9092 PM9093 106 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 115
    You might have a problem with the screen contrast. Press USER , then press F4 . Now you can use the arrow keys to adjust the contrast. 7 Maintaining the Test Tool Troubleshooting • The display 'auto , see Appendix A. MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412 107
  • Fluke 190-502 | Manual - Page 116
    Fluke 190 Series II Users Manual Battery Operated Fluke Accessories Do Not Function • When using battery operated Fluke accessories, always first check the battery condition of the accessory with a Fluke multimeter. 108 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 117
    FLUKE guarantees the properties expressed in numerical values with the stated tolerance. Specified non-tolerance numerical values indicate those that could be nominally expected from the mean of a range of identical ScopeMeter manual contains information and warnings that must be followed by the user
  • Fluke 190-502 | Manual - Page 118
    Fluke 190 Series II Users Manual Oscilloscope Isolated Inputs A,B, C and D (Vertical) Number of Channels Fluke 190-xx2 2 (A, B) Fluke 190-xx4 4 (A,B,C,D) Bandwidth, DC Coupled FLUKE 190-502 500 MHz (-3 dB) FLUKE 190-2xx 200 MHz (-3 dB) FLUKE 190-1xx 100 MHz (-3 dB) FLUKE 190-062 60 MHz (-3 dB
  • Fluke 190-502 | Manual - Page 119
    screen. Use Zoom and Scroll to see waveform details 30k samples > 18k min/max/ average values per measurement Max. Samplerate 190-062: 625 MS/s 190-102/104: 1.25 GS/s 190-202/204: 2.5 GS/s (1 or 2 channels on) 190-204: 1.25 GS/s (3 or 4 channels on) 190-502: 5 GS/s (1 channel on) 190-502
  • Fluke 190-502 | Manual - Page 120
    Fluke 190 Series II Users Manual Trigger and Delay Trigger Modes Automatic, Edge, Video, Pulse Width, N-Cycle, External (190-xx2) Trigger Delay up to +1200 divisions Pre Trigger View one full screen length Delay 12 div to +1200 div Max. Delay 60 s at 5 s/div Automatic Connect-and-View Trigger
  • Fluke 190-502 | Manual - Page 121
    >T) or 500 ns (=T, ≠T), a maximum of 10 s, and a resolution of 0.01 div. with a minimum of 50 ns 8 Specifications Oscilloscope 20 mV Automatic Capturing Scope Screens Capacity 100 Scope Screens For viewing screens, see Replay function. 113 MyFlukeStore Shop for Fluke products online at: www.
  • Fluke 190-502 | Manual - Page 122
    Fluke 190 Series II Users Manual 127. At least 1.5 waveform period must be visible on the screen. General Inputs A, B, C and D DC Common Mode Rejection VDC) Maximum Voltage with 10:1 probe 1000 V direct (1:1 300 V Maximum Resolution with 10:1 probe 1 mV direct (1:1 100 µV Full Scale Reading 999
  • Fluke 190-502 | Manual - Page 123
    least 1.5 waveform period is on the screen AC+DC Voltage (True RMS) Maximum Voltage with 10:1 probe 1000 V direct (1:1 300 V Maximum Resolution with 10:1 probe 1 mV direct (1:1 10 ns/div and 10 periods on the screen). 115 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 124
    Fluke 190 Series II Users Manual Duty Cycle (DUTY) Range 4.0 % to 98.0 % Resolution 0.1 % (when period > 2 div) Full Scale Reading 999 counts (3-digit display) Accuracy (logic or pulse 0.5 % +2 counts) Pulse Width (PULSE) Resolution (with GLITCH off 1/100 division Full Scale Reading 999
  • Fluke 190-502 | Manual - Page 125
    Phase (A and B, C and D) Range 180 to +180 degrees Resolution 1 degree Accuracy 0.1 Hz to 1 MHz 2 degrees 1 MHz to 10 MHz 3 degrees Fluke 190-xx4 Meter Measurements for Fluke 190-xx4 Four of the Automatic Scope Measurements as defined above may be displayed at the same time, using larger screen
  • Fluke 190-502 | Manual - Page 126
    pF) Max. Input Voltage Fluke 190-202, -102, -062 1000 V CAT III 600 V CAT IV (For detailed specifications, see "Safety") Max. Input Voltage Fluke 190-502 1000 V CAT II 600 V CAT III (For detailed specifications, see "Safety") Meter Functions Ranging Auto, Manual Modes Normal, Relative General
  • Fluke 190-502 | Manual - Page 127
    frequencies the frequency roll off of the Meter input starts affecting accuracy. Normal Mode DC Rejection 50 dB 8 Specifications Meter Measurements for Fluke 190-xx2 AC+DC Voltage (True RMS) Ranges......500.0 mV, 5.000 V, 50.00 V, 500.0 V, 1100 V Full Scale Reading 5000 counts Accuracy DC to
  • Fluke 190-502 | Manual - Page 128
    Fluke 190 Series II Users Manual Recorder TrendPlot (Meter or Scope) Chart recorder that plots a graph of min and max values of Meter or Continuous Roll Start/Stop on Trigger Time Reference time from start, time of day 120 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 129
    individual samples Replay Displays a maximum of 100 captured quad input Scope screens. Replay modes Step by Step, Replay as Animation 8 Specifications Zoom, (power over time between cursors using powertrace AxB or CxD) 121 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 130
    Fluke 190 Series II Users Manual Miscellaneous Display View Area 126.8 x 88.4 mm (4.99 x 3.48 inches) Resolution 320 x 240 pixels Backlight LED (Temperature compensated) Brightness Power Adapter: 200 cd/m2 Battery Power: 90 cd/ m2 Display Auto-OFF time (battery saving) .........30
  • Fluke 190-502 | Manual - Page 131
    Each memory can contain one screen image External Memory USB stick, 2GB max Mechanical Size 265 x 190 x 70 mm (10.5 x 7.5 x 2.8 in) Weight FLUKE 190-xx4........... 2.2 kg (4.8 lbs) including battery FLUKE 190-5xx........... 2.2 kg (4.8 lbs) including battery FLUKE 190-xx2........... 2.1 kg (4.6 lbs
  • Fluke 190-502 | Manual - Page 132
    Fluke 190 Series II Users Manual Environmental Environmental MIL-PRF-28800F, Class 2 Temperature Operating: CATII 1000V 3 km (10 000 feet) CATIV 600V, CATIII 1000V 2 km (6 600 feet) Operating, Fluke 190-502: CATIII 300V, CATII 600V 3 km (10 000 feet) CATIV 300V, CATIII 600V 2 km (6 600
  • Fluke 190-502 | Manual - Page 133
    600 V CAT IV 300 V CAT III 1000 V CAT III 600 V CAT IV FLUKE 190-502 300 V CAT III 1000 V CAT II 600 V CAT III 300 V CAT III 1000 V CAT II 600 V CAT III 8 Specifications Safety Max. Floating Voltage FLUKE 190-xx4, -102, -202, -062 (test tool or test tool + VPS410) From any terminal to earth ground
  • Fluke 190-502 | Manual - Page 134
    Fluke 190 Series II Users Manual MAX. INPUT VOLTAGE (Vrms) 500 200 100 50 20 10 5 2 1 0.01 0.02 0.05 0.1 0.2 0.5 1 2 5 10 20 50 100 FREQUENCY (MHz) Figure 52. Max. Input Voltage vs. Frequency VOLTAGE (Vrms) 1000 500 CAT III (-502: CAT II) CAT IV (-502: CAT III) 200 100 50 30 20 10 5 2 1 0.01 0.
  • Fluke 190-502 | Manual - Page 135
    1 % 20 kHz to 1 MHz 2 % 1 MHz to 25 MHz 3 % For higher frequencies the probe's roll off starts affecting the accuracy. For further probe specifications see the instruction sheet supplied with the VPS510 probe set. 127 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 136
    Fluke 190 Series II Users Manual Electromagnetic Immunity The Fluke 190 Series II test tools, including standard accessories, conform with the EEC directive 2004/108 E = 3V/m 500 mV to 1000 V, 500 Ohm to 30 MOhm ranges 128 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 137
    No visible disturbance Frequency range 2 GHz to 2.7 GHz 8 Specifications Electromagnetic Immunity Table 6 E = 1V/m 500 mV to 1000 V , 500 Ohm to 30 MOhm ranges 129 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 138
    Fluke 190 Series II Users Manual 130 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 139
    Li-ion Battery, 105 -C- C290 Hard Case, 105 Calibrating the Test Tool, 101 Calibrating Voltage Probes, 99, 122 calibration date, 101 number, 101 Capturing 100 Screens, 51, 113 Case, 105 131 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 140
    ScopeMeter 190 Series II Users Manual Charge Time, 122 Charging, Key, 3, 4 demo version, 80 installation, 80 Software, 3, 4 Freezing the Screen, 20 Frequency (Hz), 115 Frequency Response, 110, 118 -G- Glitch Capture, 23 132 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 141
    Adapter, 102 Power Down Timer, 91 Power Factor, 116 Power Factor (Power...), 19 power shutdown, 90 Powering the Test Tool, 11 Pre Trigger, 59 Print Screen, 75 133 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 142
    ScopeMeter 190 Series II Users Manual Probe Accessory Extension Set, 105 Probe Calibration, 99, 122 Probe Replacement Set, 103, 104 Probe Tip to BNC Adapter, 4 Probe Type, 16, 33 Pulse Trigger, 67 Pulse Width, 116 Pulse Width Trigger, 113 -R- Readings, 19 190-104, 33, 35 190-204, 33 Recalibrating,
  • Fluke 190-502 | Manual - Page 143
    On Edges, 61 On Pulses, 67 On Video, 65 On Waveforms, 57 Troubleshooting, 107 TV Triggering, 65 -U- Unpacking, 2 USB drivers, 80 USB Interface Lines, 66 Video Trigger, 65, 113 Viewing Stored Screens, 77 Voltage Probe Set, 3 Voltage probe set 100 Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 144
    ScopeMeter 190 Series II Users Manual 136 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 145
    Appendices Appendix Title Page A Installing USB Drivers A-1 B Battery Pack MSDS B-1 C Instrument Security Procedures C-1 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 146
    MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 147
    . This is required for installation on Win 7. Note: The Fluke 190 Series II instrument requires two drivers to be loaded in sequence. − 1st it requires installation of the Fluke 190 ScopeMeter USB driver − 2nd it requires installation of the Fluke USB Serial port Both of these drivers need to be
  • Fluke 190-502 | Manual - Page 148
    Fluke 190 Series II Users Manual Installing the USB Drivers To install the USB drivers, do the following: 1 Connect the Fluke 190 Series II is not required to power off. When there is no driver loaded for the Fluke 190 Series II instrument, Windows will show that there is New Hardware detected, and
  • Fluke 190-502 | Manual - Page 149
    with the ScopeMeter, which contains the drivers. 3 Follow the instructions on screen. When the driver has finished installation click 'Finish' to complete the first step of the driver installation. A Appendices Installing USB Drivers A-3 MyFlukeStore Shop for Fluke
  • Fluke 190-502 | Manual - Page 150
    Fluke 190 Series II Users Manual 4 After completing the first step the New Hardware Wizard internet, you need to load the CD-ROM, supplied with the ScopeMeter, which contains the drivers. 5 Follow the instructions on screen. When the driver finished installation click 'Finish' to complete the final
  • Fluke 190-502 | Manual - Page 151
    the device manager for your Windows version) From the device manager click on the + sign to Expand the 'Universal Serial Bus controllers'. The 'Fluke 190 ScopeMeter' should be listed here. From the device manager click on the + sign to Expand the 'Ports (COM & LPT)' Universal Serial Bus controllers
  • Fluke 190-502 | Manual - Page 152
    Fluke 190 Series II Users Manual Notes 1) Sometimes application software may require a different port number. (for example in the range Com 1..4). In this situation the COM port number can be changed manually. To manually assign a different COM port number right click on 'Fluke USB Serial Port COM
  • Fluke 190-502 | Manual - Page 153
    Li-ion Battery Pack Contact Fluke for a Battery Material Safety Data Sheet (MSDS) or Compliance Information. B Appendices Battery Pack MSDS Appendix B Battery Pack MSDS B-1 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 154
    Fluke 190 Series II Users Manual B-2 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 155
    Memory The 190 Series II ScopeMeter test tools has screens and instrument setups Security Summary The operating code (instrument firmware) stored in D5000, D5002 can be read using dedicated remote interface commands (only available for use by Product Development). C-1 MyFlukeStore Shop for Fluke
  • Fluke 190-502 | Manual - Page 156
    Fluke 190 Series II Users Manual The instrument firmware can be loaded using a dedicated Fluke software program (only available for Fluke saved screens and instrument setups: 1 Push the SAVE button. SAVE 2 F4 Press F4 - FILE OPTIONS... If you see a screen like figure C-1, no screens and
  • Fluke 190-502 | Manual - Page 157
    sources, open the battery door present at the rear panel and lift the battery out of the instrument. C Appendices Instrument Security Procedures C-3 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • Fluke 190-502 | Manual - Page 158
    Fluke 190 Series II Users Manual C-4 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412
  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122
  • 123
  • 124
  • 125
  • 126
  • 127
  • 128
  • 129
  • 130
  • 131
  • 132
  • 133
  • 134
  • 135
  • 136
  • 137
  • 138
  • 139
  • 140
  • 141
  • 142
  • 143
  • 144
  • 145
  • 146
  • 147
  • 148
  • 149
  • 150
  • 151
  • 152
  • 153
  • 154
  • 155
  • 156
  • 157
  • 158

May 2011
© 2011 Fluke Corporation. All rights reserved. Specifications are subject to change without notice.
All product names are trademarks of their respective companies.
ScopeMeter 190 Series II
Fluke 190-062, -102, -104, -202, -204, -502
Users Manual
Shop for Fluke products online at:
1.877.766.5412
www.
MyFlukeStore
.com