Fluke 190-504/AM Product Manual

Fluke 190-504/AM Manual

Fluke 190-504/AM manual content summary:

  • Fluke 190-504/AM | Product Manual - Page 1
    ScopeMeter® Test Tool 190 Series II Fluke 190-062, -102, -104, -202, -204, -502, -504 Users Manual May 2011, Rev.2, 1/14 © 2011-2014 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
  • Fluke 190-504/AM | Product Manual - Page 2
    hpp00.eps
  • Fluke 190-504/AM | Product Manual - Page 3
    Fluke does not warrant that software will be error free or operate without interruption. Fluke Fluke. Warranty support is available if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price. Fluke Fluke authorized service center. Fluke
  • Fluke 190-504/AM | Product Manual - Page 4
    SERVICE CENTERS To locate an authorized service center, visit us on the World Wide Web: http://www.fluke.com or call Fluke using any of the phone numbers listed below: +1-888-993-5853 in U.S.A. and Canada +31-40-2675200 in Europe +1-425-446-5500 from other countries
  • Fluke 190-504/AM | Product Manual - Page 5
    Table of Contents Chapter Title Page Chapter 1...1-11 1 Using the Scope and Meter 1-11 Powering the Test Tool 1-11 Resetting the Test Tool 1-12 Navigating a Menu ...1-13 Hiding Key Labels and Menus 1-14 Key Illumination ...1-15 Input Connections ...1-15 Making Input Connections 1-
  • Fluke 190-504/AM | Product Manual - Page 6
    ScopeMeter® Test Tool 190 Series II Users Manual Using Average for Smoothing Waveforms Smart average ...Using Persistence, Envelope and Dot-Join to Display Waveforms Displaying Glitches Suppressing High Frequency Noise Acquiring Waveforms...Setting
  • Fluke 190-504/AM | Product Manual - Page 7
    Contents (continued) Displaying Recorded Data 2-44 Changing the Recorder Options 2-44 Turning Off the TrendPlot Display 2-44 Recording Scope Waveforms In Deep Memory (Scope Record 2-45 Starting a Scope Record Function 2-45 Displaying Recorded Data 2-46 Using Scope Record in Single Sweep Mode
  • Fluke 190-504/AM | Product Manual - Page 8
    ScopeMeter® Test Tool 190 Series II Users Manual Triggering on Noisy Waveforms Making a Single Acquisition N-Cycle Triggering ...Triggering on External Waveforms (models 190-xx2 Triggering on Video Signals Triggering on Video Frames Triggering on Video Lines Triggering on Pulses ...Detecting
  • Fluke 190-504/AM | Product Manual - Page 9
    Parts and Accessories...7-102 Replacement Parts ...7-102 Optional Accessories 7-105 Troubleshooting ...7-108 The Test Tool Shuts Down After a Short Time Recognize The Test Tool 7-108 Battery Operated Fluke Accessories Do Not Function 7-109 8 Specifications ...8-111 Introduction ...8-111 v
  • Fluke 190-504/AM | Product Manual - Page 10
    ScopeMeter® Test Tool 190 Series II Users Manual Oscilloscope ...Isolated Inputs A,B, C and D (Vertical Horizontal ...Trigger and Delay ...Automatic Connect-and-View Trigger Edge Trigger ...Isolated External Trigger (190 Measurements for Fluke 190-xx4 Meter Measurements for Fluke 190-xx2 8-112
  • Fluke 190-504/AM | Product Manual - Page 11
    Contents (continued) Meter Input (Banana Jacks 8-120 Meter Functions...8-120 General...8-120 Ohms (Ω)...8-121 Continuity (CONT) ...8-121 Diode ...8-121 Temperature (TEMP 8-121 DC Voltage (VDC) ...8-121 AC Voltage (VAC)...8-121 AC+DC Voltage (True RMS 8-122 Amperes (AMP) ...8-122 Recorder ...8-122
  • Fluke 190-504/AM | Product Manual - Page 12
    ScopeMeter® Test Tool 190 Series II Users Manual 10:1 Probe VPS410 ...8-129 Accuracy ...8-129 Electromagnetic Immunity 8-130 viii
  • Fluke 190-504/AM | Product Manual - Page 13
    instructions in this manual apply to all ScopeMeter® Test Tool 190 Series II versions (hereafter referred to as the instrument or as the test tool). The versions are listed below. The version 190 190-x04. Introduction Version 190-062 190-102 190-104 190-202 190-204 190-502 190-504 Description
  • Fluke 190-504/AM | Product Manual - Page 14
    ScopeMeter® Test Tool 190 Series II Users Manual Unpacking the Test Tool Kit The items in Figure 1 are included in your test tool kit: 1 2 4 3 a c b d e f 7 8 9 10 Note When new, the rechargeable Li-ion battery
  • Fluke 190-504/AM | Product Manual - Page 15
    Chapter 6 for mounting instructions) 3 Power Adapter (country dependent, may vary from what is shown in Figure 1) 4 USB interface cable for PC connection (USB-A to mini-USB-B) 5 Safety Information sheet + CD ROM with Users Manual (multi-language) and FlukeView® ScopeMeter® Software for Windows demo
  • Fluke 190-504/AM | Product Manual - Page 16
    ® Test Tool 190 Series II Users Manual Fluke 190-xxx /S versions include also the following items (SCC290 kit): # Description 12 FlukeView® ScopeMeter® Software for Windows activation key (converts FlukeView® DEMO status into full operational status). 13 Hard Shell Carrying Case Safety Information
  • Fluke 190-504/AM | Product Manual - Page 17
    used on the product and in this manual.  Risk of Danger. Important information. See Manual.  Double Insulated  Conforms to relevant Do not dispose of this product as unsorted municipal waste. Go to Fluke's website for recycling information. Measurement Category III is applicable to test
  • Fluke 190-504/AM | Product Manual - Page 18
    ScopeMeter® Test Tool 190 Series II Users Manual Warning To avoid electrical shock or fire: • Use only the Fluke power supply, Model BC190 (Power Adapter). • Before use, check that the selected/indicated range on the BC190 matches the local line power voltage and frequency. •
  • Fluke 190-504/AM | Product Manual - Page 19
    directly transmitted to the product. • Do not use exposed metal BNC or banana plug connectors. Fluke offers cables with plastic, safety designed BNC connectors suitable for the ScopeMeter® Test Tool product, see Chapter 7 'Optional accessories'. • Do not insert metal objects into connectors. • Use
  • Fluke 190-504/AM | Product Manual - Page 20
    ScopeMeter® Test Tool 190 Series II Users Manual IV refers to the overhead or underground utility service of an installation. Measurement Category III refers to battery pack The battery pack Fluke model BP290 (26 Wh)/BP291 (52 Wh) has been tested in accordance with the UN Manual of Tests and Criteria
  • Fluke 190-504/AM | Product Manual - Page 21
    the battery pack. • The battery pack needs to be charged before use. Use only Fluke approved power adapters to charge the battery pack. Refer to Fluke's safety instructions and Users Manual for proper charging instructions. • Do not leave a battery on prolonged charge when not in use. Safe Use of
  • Fluke 190-504/AM | Product Manual - Page 22
    ScopeMeter® Test Tool 190 Series II Users Manual • Alteration of battery pack: there shall be no attempt to open, properly: do not dispose of the battery as unsorted municipal waste. Go to Fluke's website for recycling information. • Dispose in discharged condition and cover the battery terminals
  • Fluke 190-504/AM | Product Manual - Page 23
    operations. Powering the Test Tool Follow the procedure (steps 1 through 3) in Figure 2 to power the test tool from a standard ac outlet. See Chapter 6 for instructions on using battery power. Turn the test tool on with the on/off key. The test tool powers up in its last setup configuration. The
  • Fluke 190-504/AM | Product Manual - Page 24
    ScopeMeter® Test Tool 190 Series II Users Manual Resetting the Test Tool If you want to reset the test tool to the factory settings, do the following: 1 Turn the test tool off. 2 USER
  • Fluke 190-504/AM | Product Manual - Page 25
    Navigating a Menu The following example shows how to use the test tool's menus to select a function. Sequently follow steps 1 through 3 to open the scope menu and to choose an item. 1 SCOPE Press the SCOPE key to display the labels that define the present use for the four blue function keys at
  • Fluke 190-504/AM | Product Manual - Page 26
    ScopeMeter® Test Tool 190 Series II Users Manual Note To exit the menu at any moment press F4 and the F1...F4 key labels, apply to the illuminated channel key(s). Off: - MANUAL AUTO TRIGGER On: Manual operating mode. Off: Automatic operating mode, optimizes the waveform position, range, time base
  • Fluke 190-504/AM | Product Manual - Page 27
    BNC jack signal inputs (models 190-xx4), or two safety BNC jack inputs and two safety 4-mm banana jack inputs (models 190-xx2). See Figure 5. − To maximally benefit from having independently isolated floating inputs and to avoid problems caused by improper use, read Chapter 6: "Tips". − For an
  • Fluke 190-504/AM | Product Manual - Page 28
    ScopeMeter® Test Tool 190 Series II Users Manual Figure 6. Scope Connections 16 Adjusting the Probe Type Settings To obtain correct measurement results the test tool probe type settings must correspond to the connected
  • Fluke 190-504/AM | Product Manual - Page 29
    Selecting an Input Channel To select an input channel, do the following: A Press the required channel key (A...D): B - the channel is turned on C - labels for the F1...F4 keys are shown. Press the channel key again D to turn the labels off/on (toggle). mV RANGE MOVE V - the channel key
  • Fluke 190-504/AM | Product Manual - Page 30
    ScopeMeter® Test Tool 190 Series II Users Manual Displaying an Unknown Signal with Connect-and-View™ The several signals. To enable the Connect-and-View feature when the test tool is in MANUAL mode, do the following: 1 MANUAL Perform an Auto Set. AUTO appears at AUTO the top right of the screen, the
  • Fluke 190-504/AM | Product Manual - Page 31
    Making Automatic Scope Measurements The test tool offers a wide range of automatic scope measurements. In addition to the waveforms you can display four numeric readings: READING 1 ... 4. These readings are selectable independently, and the measurements can be done on the input A, input B, input C
  • Fluke 190-504/AM | Product Manual - Page 32
    ScopeMeter® Test Tool 190 Series II Users Manual 5 Open the PEAK menu. ENTER 6 Select the Peak-Peak ENTER measurement. Figure 8 shows an example of the screen with two readings. The character size will
  • Fluke 190-504/AM | Product Manual - Page 33
    Using Average, Persistence and Glitch Capture Using Average for Smoothing Waveforms To smooth the waveform, do the following: 1 SCOPE Display the SCOPE key labels. 2 F4 Open the WAVEFORM OPTIONS menu. 3 Jump to Average: 4 Select On... to open the AVERAGE ENTER menu. 1 Using the Scope and
  • Fluke 190-504/AM | Product Manual - Page 34
    ScopeMeter® Test Tool 190 Series II Users Manual Figure 9. Smoothing a Waveform 22 Using Persistence, Envelope and Dot-Join to Display Waveforms You can use Persistence to observe dynamic signals. See Figure 10. 1 SCOPE
  • Fluke 190-504/AM | Product Manual - Page 35
    . When you select the 2 mV/div range Glitch Detect will automatically be turned Off. In the 2 mV/div range you can set Glitch Detect On manually. 23
  • Fluke 190-504/AM | Product Manual - Page 36
    ScopeMeter® Test Tool 190 Series II Users Manual Suppressing High Frequency Noise Switching the glitch detection off (Glitch: Off) will suppress the high frequency noise on a waveform. Averaging will suppress the noise even
  • Fluke 190-504/AM | Product Manual - Page 37
    Normal - optimal waveform update rate and zoom range combination 4 F4 Exit the menu. See also Table 2 in Chapter 8. Selecting AC-Coupling After a reset, the test tool is dc-coupled so that ac and dc voltages appear on the screen. Use ac-coupling when you wish to observe a small ac signal that
  • Fluke 190-504/AM | Product Manual - Page 38
    ScopeMeter® Test Tool 190 Series II Users Manual Variable Input Sensitivity The variable input sensitivity variable input sensitivity on for example input A, do the following: 1 Apply the input signal. 2 MANUAL AUTO Perform an Auto Set (AUTO must appear at the top of the screen). An Auto Set
  • Fluke 190-504/AM | Product Manual - Page 39
    Working with Noisy Waveforms To suppress high frequency noise on waveforms, you can limit the working bandwidth to 10 kHz or 20 MHz. This function smoothes the displayed waveform. For the same reason, it improves triggering on the waveform. To choose the 10 kHz bandwidth on for example input A, do
  • Fluke 190-504/AM | Product Manual - Page 40
    ScopeMeter® Test Tool 190 Series II Users Manual 4 Select Function: +, -, x or XY- ENTER mode. 5 Select the first waveform: ENTER Source 1: A, B, C or D 6 Select the second waveform: ENTER Source 2: A, B, C or D The mathematical function key labels
  • Fluke 190-504/AM | Product Manual - Page 41
    3 Jump to Waveform: and select ENTER Mathematics... to open the Mathematics menu. 4 Select Function: Spectrum. ENTER 5 Select the source waveform for the ENTER spectrum: Source : A, B, C or D 6 Select Window: Auto (automatic ENTER windowing), Hanning, Hamming, or None (no windowing).
  • Fluke 190-504/AM | Product Manual - Page 42
    ScopeMeter® Test Tool 190 Series II Users Manual Figure 11. Spectrum Measurement Comparing Waveforms You can display a fixed reference waveform with the actual waveform for comparison. To create a reference waveform and to display
  • Fluke 190-504/AM | Product Manual - Page 43
    4 Select On to display the reference ENTER waveform. This can be: - the last used reference waveform (if not available no reference waveform will be shown). - the envelope waveform if the persistence function Envelope is on. Select Recall... to recall a saved waveform (or waveform envelope)
  • Fluke 190-504/AM | Product Manual - Page 44
    ScopeMeter® Test Tool 190 Series II Users Manual Pass - Fail Testing You can use a reference waveform as a test template for the actual waveform. If at least one sample of a waveform is outside the
  • Fluke 190-504/AM | Product Manual - Page 45
    1 Using the Scope and Meter Making Automatic Meter Measurements (for models 190-xx4) Making Automatic Meter Measurements (for models 190-xx4) The test tool offers a wide range of automatic meter measurements. You can display four large numeric readings: READING 1 ... 4. These readings are
  • Fluke 190-504/AM | Product Manual - Page 46
    ScopeMeter® Test Tool 190 Series II Users Manual Making Relative Meter Measurements A relative measurement displays the present measurement result relative to a defined reference value. The following example shows how to perform a relative voltage
  • Fluke 190-504/AM | Product Manual - Page 47
    finished. 9 ENTER Enter the new reference value. 1 Using the Scope and Meter Making Multimeter Measurements (for models 190-xx2) Making Multimeter Measurements (for models 190-xx2) The screen displays the numeric readings of the measurements on the meter input. Making Meter Connections Use
  • Fluke 190-504/AM | Product Manual - Page 48
    ScopeMeter® Test Tool 190 Series II Users Manual Measuring Resistance Values To measure a resistance, do the following: 1 Connect the red and black test leads from the 4-mm banana jack inputs to the resistor. 2
  • Fluke 190-504/AM | Product Manual - Page 49
    the instructions about the current probe you are using. To set up the test tool, do the following: 1 Connect a current probe (e.g. Fluke key labels. 1 Using the Scope and Meter Making Multimeter Measurements (for models 190-xx2) Figure 16. Measurement Setup 3 F1 Open the MEASUREMENT menu. 4
  • Fluke 190-504/AM | Product Manual - Page 50
    ScopeMeter® Test Tool 190 Series II Users Manual 6 Observe the sensitivity of the current probe. Highlight the corresponding sensitivity in the menu, e.g. 1 mV/A. 7 Accept the current measurement. ENTER Now, you will see a screen
  • Fluke 190-504/AM | Product Manual - Page 51
    measurement. First obtain a reference value: 1 METER Display the METER key labels. 1 Using the Scope and Meter Making Multimeter Measurements (for models 190-xx2) Now the large reading is the actual input value minus the stored reference value. The bargraph indicates the actual input value. The
  • Fluke 190-504/AM | Product Manual - Page 52
    ScopeMeter® Test Tool 190 Series II Users Manual Adjusting the reference value To adjust the reference value, do the following: 5 F3 Display the Adjust Reference menu. 6 Select the digit you want to adjust. 7
  • Fluke 190-504/AM | Product Manual - Page 53
    menu, do the following: 1 Open the recorder main menu. RECORDER (See Figure 19). Figure 19. Recorder Main Menu Trendplot Meter is only present in models 190-xx2. 41
  • Fluke 190-504/AM | Product Manual - Page 54
    ScopeMeter® Test Tool 190 Series II Users Manual Plotting Measurements Over Time (TrendPlot™) Use the TrendPlot function to plot a graph of Scope or Meter measurements (readings) as function of time. Note Because the
  • Fluke 190-504/AM | Product Manual - Page 55
    . 6 F1 Set RECORDER to RUN to restart. Note Scope TrendPlot is not possible on cursor related measurements. As an alternative you may use the PC software FlukeView® ScopeMeter®. Figure 20. TrendPlot Reading 43
  • Fluke 190-504/AM | Product Manual - Page 56
    ScopeMeter® Test Tool 190 Series II Users Manual Displaying Recorded Data When in normal view (NORMAL), only the twelve most recently recorded divisions are displayed on screen. All previous recordings are stored in
  • Fluke 190-504/AM | Product Manual - Page 57
    2 Using The Recorder Functions Recording Scope Waveforms In Deep Memory (Scope Record) Recording Scope Waveforms In Deep Memory (Scope Record) The SCOPE RECORD function is a roll mode that logs a long waveform of each active input. This function can be used to monitor waveforms like motion control
  • Fluke 190-504/AM | Product Manual - Page 58
    ScopeMeter® Test Tool 190 Series II Users Manual Displaying Recorded Data In Normal view, the samples that roll off the screen are stored in deep memory. When the memory is full, recording continues
  • Fluke 190-504/AM | Product Manual - Page 59
    Stop recording to unlock the OPTIONS... softkey. 6 F2 Open the RECORDER OPTIONS menu. 7 Jump to the Mode: field, select on Trigger... (models 190-xx4) or ENTER on Ext. (models 190-xx2) to open the START SINGLE SWEEP ON TRIGGERING or the START SINGLE SWEEP ON EXT. menu. 8 Select one of the
  • Fluke 190-504/AM | Product Manual - Page 60
    ScopeMeter® Test Tool 190 Series II Users Manual 9 Select the desired trigger slope (Slope:) and jump to Level:. ENTER 10 Select the 0.12V or 1.2V trigger level and accept all recorder ENTER options.
  • Fluke 190-504/AM | Product Manual - Page 61
    Chapter 3 Using Replay, Zoom and Cursors About this Chapter This chapter covers the capabilities of the analysis functions Cursor, Zoom, and Replay. These functions can be used with one or more of the primary functions Scope, TrendPlot or Scope Record. It is possible to combine two or three
  • Fluke 190-504/AM | Product Manual - Page 62
    ScopeMeter® Test Tool 190 Series II Users Manual Replaying Step-by-Step To step through the last scope screens, do the following: 1 REPLAY From scope mode, open the REPLAY menu. Observe that the
  • Fluke 190-504/AM | Product Manual - Page 63
    3 Using Replay, Zoom and Cursors Replaying the 100 Most Recent Scope Screens Replaying Continuously You can also replay the stored screens continuously, like playing a video tape. To replay continuously, do the following: 1 REPLAY From Scope mode, open the REPLAY menu. Observe that the
  • Fluke 190-504/AM | Product Manual - Page 64
    ScopeMeter® Test Tool 190 Series II Users Manual Zooming in on a Waveform To obtain a more detailed view of a waveform, you can zoom in on a waveform using the ZOOM function. To zoom in on a
  • Fluke 190-504/AM | Product Manual - Page 65
    Making Cursor Measurements Cursors allow you to make precise digital measurements on waveforms. This can be done on live waveforms, recorded waveforms, and on saved waveforms. Using Horizontal Cursors on a Waveform To use the cursors for a voltage measurement, do the following: 1 CURSOR From
  • Fluke 190-504/AM | Product Manual - Page 66
    ScopeMeter® Test Tool 190 Series II Users Manual Using Vertical Cursors on a Waveform To use the cursors for a time measurement (T, 1/T), for a mVs-mAs-mWs measurement, or for an RMS measurement of the waveform
  • Fluke 190-504/AM | Product Manual - Page 67
    Notes − For mVs select probe type 'Voltage'. − For mAs select probe type 'Current'. − For mWs select mathematical function x, and probe type 'Voltage' for one channel and 'Current' for the other channel. Using Cursors on a Mathematical Result (+ - x) Waveform Cursor measurements on, for examle, a
  • Fluke 190-504/AM | Product Manual - Page 68
    ScopeMeter® Test Tool 190 Series II Users Manual Making Rise Time Measurements To measure rise time, do the the required waveform A, B, C, D or M (if a math function is active). 4 F3 Select MANUAL or AUTO (this automatically does steps 5 to 7). 5 Move the upper cursor to 100% of the waveform
  • Fluke 190-504/AM | Product Manual - Page 69
    control the trigger level, slope, and trigger delay for a better view of the signal. (See next section.) • For dedicated applications, use one of the four manual trigger functions: • Edge triggering • Video triggering • Pulse Width triggering • External Triggering (models
  • Fluke 190-504/AM | Product Manual - Page 70
    ScopeMeter® Test Tool 190 Series II Users Manual Setting Trigger Level and Slope The Connect-and-View™ function enables hands-off triggering to display complex unknown signals. When your test tool is in manual range, do the following: MANUAL AUTO Perform an auto set. AUTO appears at the top
  • Fluke 190-504/AM | Product Manual - Page 71
    Using Trigger Delay or Pre-trigger You can begin to display the waveform some time before or after the trigger point has been detected. Initially, you have a half screen (6 divisions) of pre-trigger view (negative delay). To set the trigger delay, do the following: 5 MOVE Hold down to adjust
  • Fluke 190-504/AM | Product Manual - Page 72
    ScopeMeter® Test Tool 190 Series II Users Manual Automatic Trigger Options In the trigger menu, settings for automatic However, when you measure frequencies lower than 15 Hz, the test tool must be instructed to analyze low frequency components for automatic triggering: 4 Select > 1 HZ and return
  • Fluke 190-504/AM | Product Manual - Page 73
    Triggering on Edges If the signal is unstable or has a very low frequency, use edge triggering to obtain full manual trigger control. To trigger on rising edges of the input A waveform, do the following: 1 TRIGGER Display the TRIGGER key labels. 2 F4 Open the TRIGGER OPTIONS
  • Fluke 190-504/AM | Product Manual - Page 74
    ScopeMeter® Test Tool 190 Series II Users Manual Triggering on Noisy Waveforms To reduce jitter on A waveform, continue from step 3 (page 61) again: 4 Select Single Shot. ENTER 62 The word MANUAL appears at the top of the screen indicating that the test tool is waiting for a trigger. As soon
  • Fluke 190-504/AM | Product Manual - Page 75
    N-Cycle Triggering N-Cycle triggering enables you to create a stable picture of for example n-cycle burst waveforms. Each next trigger is generated after the waveform has crossed the trigger level N times in the direction that complies with the selected trigger slope. To select N-Cycle triggering,
  • Fluke 190-504/AM | Product Manual - Page 76
    ScopeMeter® Test Tool 190 Series II Users Manual Triggering on External Waveforms (models 190-xx2) Use external triggering when you want to display waveforms on inputs A and B while triggering on a third signal. You can choose external triggering with automatic
  • Fluke 190-504/AM | Product Manual - Page 77
    Triggering on Video Signals To trigger on a video signal, first select the standard of the video signal you are going to measure: 1 Apply a video signal to the red input A. 2 TRIGGER Display the TRIGGER key labels. 4 Triggering on Waveforms Triggering on Video Signals 3 F4 Open the Trigger
  • Fluke 190-504/AM | Product Manual - Page 78
    ScopeMeter® Test Tool 190 Series II Users Manual Triggering on Video Frames Use FIELD 1 or FIELD 2 to trigger either on the first half of the frame (odd) or on the second half of
  • Fluke 190-504/AM | Product Manual - Page 79
    Triggering on Pulses Use pulse width triggering to isolate and display specific pulses that you can qualify by time, such as glitches, missing pulses, bursts or signal dropouts. Detecting Narrow Pulses To set the test tool to trigger on narrow positive pulses shorter than 5 ms, do the following: 1
  • Fluke 190-504/AM | Product Manual - Page 80
    ScopeMeter® Test Tool 190 Series II Users Manual Tip The test tool stores all triggered screens in the replay memory. For example, if you setup your triggering for glitches, you can capture 100
  • Fluke 190-504/AM | Product Manual - Page 81
    4 Select the positive pulse icon to ENTER trigger on a positive pulse, then jump to Condition: . 5 Select >t, then jump to Update: . ENTER 6 Select On Trigger and exit the ENTER menu. The test tool is now prepared to trigger on pulses that are more than a selectable time in duration.
  • Fluke 190-504/AM | Product Manual - Page 82
    ScopeMeter® Test Tool 190 Series II Users Manual 70
  • Fluke 190-504/AM | Product Manual - Page 83
    information on computer communication at the end of this chapter. Using the USB Ports The test tool is provided with two USB ports: • a USB-host port to connect an external flash memory drive ('USB-stick') for data storage. • a mini-USB-B port which allows you to connect the test tool to a PC for
  • Fluke 190-504/AM | Product Manual - Page 84
    ScopeMeter® Test Tool 190 Series II Users Manual Saving and Recalling You can: • Save screens and setups to internal memory, and recall them again from memory. The test tool has 30 'screen and
  • Fluke 190-504/AM | Product Manual - Page 85
    1 screen setup + replay screens/record data setup + trendplot data screen image (imagexxx.bmp - a screen image can be copied to a USB stick connected to the test tool. The USB stick connected to a PC allows you to insert the image for instance into a text document. The copying function is available
  • Fluke 190-504/AM | Product Manual - Page 86
    ® Test Tool 190 Series II Users Manual 3 F1 Select the target memory INT (internal memory) or USB (USB device). Observe the new SAVE menu if you select USB. You can save data in .csv format to a USB stick. The saved .csv file can be used to analyze the data in e.g. FlukeView® ScopeMeter® or in
  • Fluke 190-504/AM | Product Manual - Page 87
    All memories in use If no free memory locations are available a message pops up that proposes to you to overwrite the oldest data set. Do one of the following: If you don't want to overwrite the oldest data set, - press F3 , the delete one or more memory locations, and save again. If you want
  • Fluke 190-504/AM | Product Manual - Page 88
    ScopeMeter® Test Tool 190 Series II Users Manual Saving Screens in .bmp Format (Print Screen) To save a screen in bitmap (.bmp) format, do the following: 1 Display the SAVE key labels. SAVE 2 F3 Save the screen to: - Internal memory (INT) if no USB device is connected. - a USB a USB device.
  • Fluke 190-504/AM | Product Manual - Page 89
    do the following: 1 Display the SAVE key labels. SAVE 2 F2 3 F1 4 Open the RECALL menu. Select the source, internal memory (INT) or a USB device. Highlight DATA . 5 ENTER Accept your choice and jump to the filename field 6 Select the file to be recalled. 7 ENTER Recall the selected
  • Fluke 190-504/AM | Product Manual - Page 90
    ScopeMeter® Test Tool 190 Series II Users Manual Viewing Stored Screens To scroll through the memories while looking at the stored screens, do the following: 1 Display the SAVE key labels. SAVE 2 F2 Open the RECALL menu. 3 F1 Select the source, internal memory (INT) or a USB device. 4
  • Fluke 190-504/AM | Product Manual - Page 91
    7 ENTER Copy or delete the selected files. Using FlukeView® With the FlukeView® software you can upload waveform data and screen bitmaps to your PC or notebook computer for further processing. USB drivers for the test tool and a FlukeView® Demo version with restricted functionality are available
  • Fluke 190-504/AM | Product Manual - Page 92
    USB drivers, see Appendix A. • Install the FlukeView® Demo version. For information about installing and using the FlukeView®. ScopeMeter® software see the FlukeView® Users Manual code SW90W. For use with the Fluke Series II ScopeMeter®test tools FlukeView®ScopeMeter®release V5.1 or higher is
  • Fluke 190-504/AM | Product Manual - Page 93
    Chapter 6 Tips About this Chapter This chapter gives you information and tips on how you can make the best use of the test tool. Using the Standard Accessories The following illustrations show the use of the standard accessories such as voltage probes, test leads, and the various clips. 81
  • Fluke 190-504/AM | Product Manual - Page 94
    ScopeMeter® Test Tool 190 Series II Users Manual A Figure 37. HF Voltage Probe Connection Using Ground Spring Warning To avoid electrical shock or fire, do not connect the ground spring to voltages higher
  • Fluke 190-504/AM | Product Manual - Page 95
    should be aware of parasitic capacitance. (See Figure 39, Figure 41, Figure 42 and Figure 43.) Note The input channels are electrically isolated from the USB port and from the power adapter input. 83
  • Fluke 190-504/AM | Product Manual - Page 96
    ScopeMeter® Test Tool 190 Series II Users Manual AB C D Figure 39. Parasitic capacitance between probes, instrument and environment Note: Parasitic capacitances such as shown in Figure 39, 41 and 43 can cause ringing
  • Fluke 190-504/AM | Product Manual - Page 97
    MOTION/MOTOR CONTROLLER DC BUS + DIGITAL GROUND D A M - DC BUS Figure 41. Parasitic capacitance between analog and digital reference 6 Tips Using the Independently Floating Isolated Inputs ANALOG INPUT DIGITAL CONTROLLER ANALOG GROUND DIGITAL GROUND D A Figure 42. Correct connection of
  • Fluke 190-504/AM | Product Manual - Page 98
    ScopeMeter® Test Tool 190 Series II Users Manual ANALOG INPUT DIGITAL CONTROLLER ANALOG GROUND DIGITAL while placed on a table. The typical position is shown in Figure 44. Kensington® lock support Power adapter input Figure 44. Using the Tilt Stand Note An optional Hanging Hook, ordering code
  • Fluke 190-504/AM | Product Manual - Page 99
    Kensington®-lock The test tool is provided with a security slot compatible with a Kensington® lock, see Figure 44. The Kensington Security Slot along with a locking cable provides physical security against thefts of opportunity. Locking cables can be obtained from, for example, laptop computer
  • Fluke 190-504/AM | Product Manual - Page 100
    ScopeMeter® Test Tool 190 Series II Users Manual Resetting the Test Tool If you want to reset the test tool to the factory settings, without clearing the memories, do the following: 1 Turn the
  • Fluke 190-504/AM | Product Manual - Page 101
    To adjust the contrast and backlight brightness, do the following: 1 USER Display the USER key labels. 2 F4 3 4 Enable the arrow keys for manual contrast and backlight adjustment. Adjust the contrast of the screen. Change the backlight. Note The new contrast and brightness are stored until
  • Fluke 190-504/AM | Product Manual - Page 102
    ScopeMeter® Test Tool 190 Series II Users Manual Changing Date and Time The test tool has a date and time clock. For example, to change the date to 19 April, 2013, do the following: 1
  • Fluke 190-504/AM | Product Manual - Page 103
    Saving Battery Life When operated on the battery, the test tool conserves power by shutting itself down. If you have not pressed a key for at least 30 minutes, the test tool turns itself off automatically. Automatic power shutdown will not occur if TrendPlot or Scope Record is on, but the backlight
  • Fluke 190-504/AM | Product Manual - Page 104
    ScopeMeter® Test Tool 190 Series II Users Manual Setting the Display AUTO-off Timer Initially the display AUTO-off timer is disabled (no automatic display turn off). You can set the display AUTO-
  • Fluke 190-504/AM | Product Manual - Page 105
    you measure frequencies lower than 15 Hz, the test tool must be instructed to analyze low frequency components for automatic triggering: 6 Tips Changing the Hz and up, then jump to ENTER Input Coupling: When you press the AUTO-MANUAL (auto set) key, the input coupling can either be set to dc or
  • Fluke 190-504/AM | Product Manual - Page 106
    ScopeMeter® Test Tool 190 Series II Users Manual 94
  • Fluke 190-504/AM | Product Manual - Page 107
    Chapter This chapter covers basic maintenance procedures that can be performed by the user. For complete service, disassembly, repair, and calibration information, see the Service Manual. (www.fluke.com) Warning • Have an approved technician repair the product. • Use only specified replacement parts
  • Fluke 190-504/AM | Product Manual - Page 108
    ScopeMeter® Test Tool 190 Series II Users Manual Charging the Batteries At delivery, the Li-ion batteries may be 96 Figure 46. Charging the Batteries Alternatively, you may choose to exchange the battery (Fluke accessory BP290 or BP291) with a fully charged one, and use the external battery
  • Fluke 190-504/AM | Product Manual - Page 109
    the Battery Pack Warning Use only the Fluke BP290 (not recommended for 190-xx4) or BP291 for replacement! When : − Store the data in the test tool's non volatile flash memory, on a computer or a USB stick. − Connect the power adapter. To replace the battery pack, proceed as follows: 1. Remove all
  • Fluke 190-504/AM | Product Manual - Page 110
    ScopeMeter® Test Tool 190 Series II Users Manual Figure 48. Removing the Battery Cover Figure 49. Removing the Battery 98
  • Fluke 190-504/AM | Product Manual - Page 111
    Calibrating the Voltage Probes To fully meet user specifications, you need to adjust the voltage probes for optimal response. The calibration consists of a high frequency adjustment and a dc calibration for 10:1 probes and 100:1 probes. The probe calibration matches the probe to the input channel.
  • Fluke 190-504/AM | Product Manual - Page 112
    ScopeMeter® Test Tool 190 Series II Users Manual A message appears asking you whether to start the 10:1 until a pure square wave is displayed. For instructions to access the trimmer screw in the probe housing see the probe instruction sheet. The test tool automatically calibrates itself to
  • Fluke 190-504/AM | Product Manual - Page 113
    number with latest calibration date, and installed (software) options. The test tool specifications (see Chapter 8) are based on a 1 year calibration cycle. Recalibration must be carried out by qualified personnel. Contact your local Fluke representative for recalibration. 4 F4 Return to the
  • Fluke 190-504/AM | Product Manual - Page 114
    ScopeMeter® Test Tool 190 Series II Users Manual Parts and Accessories The following tables list the user-replaceable parts and specific optional accessories for the various test tool models. For more optional accessories, see www.fluke.com. To order replacement parts or additional accessories,
  • Fluke 190-504/AM | Product Manual - Page 115
    Clip (black) • Ground Spring for Probe Tip (black) • Insulation Sleeve (black) See Figure 1 on page 2 for item reference. See the VPS410-II instruction sheet for voltage/CAT ratings. 7 Maintaining the Test Tool Parts and Accessories VPS410-II-R (red) VPS410-II-B (blue) VPS410-II-G (gray) VPS410-II
  • Fluke 190-504/AM | Product Manual - Page 116
    ScopeMeter® Test Tool 190 Series II Users Manual Replacement Parts (continued) Replacement Set for Sleeve for Probe Tip (black) See Figure 1 on page 2 for item reference. See the VPS410 instruction sheet for voltage/CAT ratings. BNC Feedthrough 50 Ohm (1 Watt) terminator (set of two pieces, black
  • Fluke 190-504/AM | Product Manual - Page 117
    Optional Accessories Item Voltage Probe Set, designed for use with the Fluke 190-50x test tool. The set includes the following items (not available separately): • 10:1 Voltage Probe, 500 MHz (red or blue or gray or green) • Hook
  • Fluke 190-504/AM | Product Manual - Page 118
    ScopeMeter® Test Tool 190 Series II Users Manual Optional accessories (continued) Probe Accessory Extension Set - VPS410, VPS410- into an operational version. • Hard Shell Carrying Case C290 FlukeView® ScopeMeter® Software for Windows® (full version) Hard Shell Carrying Case External Battery Charger
  • Fluke 190-504/AM | Product Manual - Page 119
    probe, 2.5 m. 1:1 30 MHz voltage probe, 1.2 m. Dual Banana Plug male to female BNC Dual Banana Jack female to male BNC Automotive Troubleshooting Kit Service Kit Meter and Drive Applications 7 Maintaining the Test Tool Parts and Accessories HH290 PM9091 PM9092 PM9093 TRM50 VPS212-R (red), VPS212
  • Fluke 190-504/AM | Product Manual - Page 120
    ScopeMeter® Test Tool 190 Series II Users Manual Troubleshooting might have a problem with the screen to a software hangup, do USB port for communication with a computer! • Make sure that no SAVE/RECAL/COPY/MOVE action from or to a USB-stick is being performed. • Make sure that the USB drivers
  • Fluke 190-504/AM | Product Manual - Page 121
    Battery Operated Fluke Accessories Do Not Function • When using battery operated Fluke accessories, always first check the battery condition of the accessory with a Fluke multimeter or follow the procedure given for that particular accessory. 7 Maintaining the Test Tool Troubleshooting 109
  • Fluke 190-504/AM | Product Manual - Page 122
    ScopeMeter® Test Tool 190 Series II Users Manual 110
  • Fluke 190-504/AM | Product Manual - Page 123
    mean of a range of identical ScopeMeter® test tools. The Test Tool meets the specified accuracy 30 minutes and two complete acquisitions after power on. Specifications are based on a 1-year calibration cycle. Environmental Data The environmental data mentioned in this manual are based on the results
  • Fluke 190-504/AM | Product Manual - Page 124
    ScopeMeter® Test Tool 190 Series II Users Manual Oscilloscope Isolated Inputs A,B, C and D (Vertical) Number of Channels Fluke 190-xx2 2 (A, B) Fluke 190-xx4 4 (A,B,C,D) Bandwidth, DC Coupled FLUKE 190-50x 500 MHz (-3 dB) FLUKE 190-2xx 200 MHz (-3 dB) FLUKE 190-1xx 100 MHz (-3 dB) FLUKE 190-062
  • Fluke 190-504/AM | Product Manual - Page 125
    1.25 GS/s 10 µs to 120 s/div 125 MS/s FLUKE 190-062: 10 ns to 4 µs /div (all channels up 190-062: 625 MS/s 190-102/104: 1.25 GS/s 190-202/204: 2.5 GS/s (1 or 2 channels on) 190-204: 1.25 GS/s (3 or 4 channels on) 190-50x: 5 GS/s (1 channel on) 190-50x: 2.5 GS/s (2 channels on) 190-504
  • Fluke 190-504/AM | Product Manual - Page 126
    ScopeMeter® Test Tool 190 Series II Users Manual Trigger and Delay Trigger Modes Automatic, Edge, Video, Pulse Width, N-Cycle, External (190-xx2) Trigger Delay up to +1200 divisions Pre Trigger View one full screen length Delay 12 div to +1200 div Max. Delay 48 s at 4 s/div
  • Fluke 190-504/AM | Product Manual - Page 127
    Video Trigger Standards ....PAL, PAL+, NTSC, SECAM, Non-interlaced Modes Lines, Line Select, Field 1 or Field 2 Source A Polarity Positive, Negative Sensitivity 0.7 division sync level Pulse Width Trigger Screen Update On Trigger, Single Shot Trigger Conditions T, >T, =T (±10 %), ≠T(±10 %)
  • Fluke 190-504/AM | Product Manual - Page 128
    ScopeMeter® Test Tool 190 Series II Users Manual Automatic Scope Measurements The accuracy of all readings is Maximum Resolution with 10:1 probe 1 mV direct (1:1 100 µV Full Scale Reading 999 counts Accuracy, FLUKE 190-xx2 DC coupled: DC to 60 Hz 1.5 % +10 counts) AC coupled, low frequencies: 50
  • Fluke 190-504/AM | Product Manual - Page 129
    ) 20 kHz to 1 MHz 5 % + 20 counts) 1 MHz to 25 MHz 10 % + 20 counts) For higher frequencies the instrument's frequency roll off starts affecting accuracy. Accuracy, FLUKE 190-xx4 DC to 60 Hz 3 % + 10 counts) 60 Hz to 20 kHz 4 % + 15 counts) 20 kHz to 1 MHz 6 % + 20 counts) 1 MHz to 25 MHz 10
  • Fluke 190-504/AM | Product Manual - Page 130
    ScopeMeter® Test Tool 190 Series II Users Manual Amperes (AMP) With Optional Current Probe or Current Shunt Ranges same as VDC, VAC, VAC+DC Probe Sensitivity 100 µV/A, 1 mV/A, 10 mV/A, 100 mV/A, 400
  • Fluke 190-504/AM | Product Manual - Page 131
    V/Hz Purpose to show the measured Vpwm value (see Vpwm) divided by the fundamental frequency on Variable AC Motor Speed drives. Accuracy Vrms + %Hz Note AC motors are designed for use with a rotating magnetic field of constant strength. This strength depends on the applied vottage (Vpwm) divided
  • Fluke 190-504/AM | Product Manual - Page 132
    ScopeMeter® Test Tool 190 Series II Users Manual Meter Measurements for Fluke 190-xx4 Four of the Automatic Scope scope waveform information. For specifications see Automatic scope Measurements above. Meter Measurements for Fluke 190-xx2 The accuracy of all measurements is within ± (% of reading +
  • Fluke 190-504/AM | Product Manual - Page 133
    ms Diode Maximum Voltage Reading 2.8 V Open Circuit Voltage 4 V Accuracy 2 % +5 counts) Measurement Current 0.5 mA, ±20 % 8 Specifications Meter Measurements for Fluke 190-xx2 Temperature (TEMP) With Optional Temperature Probe Ranges (°C or °F 40.0 to +100.0 ° -100.0 to +250.0 ° -100.0 to +500
  • Fluke 190-504/AM | Product Manual - Page 134
    ScopeMeter® Test Tool 190 Series II Users Manual AC+DC Voltage (True RMS) Ranges ..... 500.0 mV, 5.000 V, 50.00 V, 500.0 V, 1100 V Full Scale Reading 5000 counts Accuracy DC to 60 Hz 1 % +10 counts)
  • Fluke 190-504/AM | Product Manual - Page 135
    Acquisition Modes Single Sweep Continuous Roll Start/Stop on Trigger Time Reference time from start, time of day Zoom, Replay and Cursors Zoom Zoom ranges from full record overview to detailed view of individual samples Replay Displays a maximum of 100 captured quad input Scope screens. Replay
  • Fluke 190-504/AM | Product Manual - Page 136
    ScopeMeter® Test Tool 190 Series II Users Manual Miscellaneous Display View Area 126.8 x 88.4 mm (4.99 x 3.48 / m2 Display Auto-OFF time (battery saving) ......... 30 seconds, 5 minutes or disabled Power FLUKE 190-xx4, -50x: Rechargeable Li-ion Battery (model BP291): Operating Time up to 7 hours (
  • Fluke 190-504/AM | Product Manual - Page 137
    one screen image External Memory USB stick, 2GB max Mechanical Size 265 x 190 x 70 mm (10.5 x 7.5 x 2.8 in) Weight FLUKE 190-xx4........... 2.2 kg (4.8 lbs) including battery FLUKE 190-5xx........... 2.2 kg (4.8 lbs) including battery FLUKE 190-xx2........... 2.1 kg (4.6 lbs) including battery
  • Fluke 190-504/AM | Product Manual - Page 138
    ScopeMeter® Test Tool 190 Series II Users Manual Environmental Environmental MIL-PRF-28800F, Class 2 Temperature Operating: battery installed 0 to 40 °C (32 to 104 °F) no battery installed 0 to 50 °C (32 to 122 °F) Storage 20
  • Fluke 190-504/AM | Product Manual - Page 139
    V CAT IV Via VPS410 1000 V CAT III 600 V CAT IV METER/EXT banana input 1000 V CAT III 600 V CAT IV 8 Specifications Safety Max. Floating Voltage FLUKE 190-xxx (test tool or test tool + VPS410) From any terminal to earth ground ......... 1000 V CAT III 600 V CAT IV Between any terminal 1000 V CAT
  • Fluke 190-504/AM | Product Manual - Page 140
    ScopeMeter® Test Tool 190 Series II Users Manual MAX. INPUT VOLTAGE (Vrms) 500 200 100 50 20 10 5 2 1 0.01 0.02 0.05 0.1 0.2 0.5 1 2 5 10 20 50 100 FREQUENCY (MHz) Figure 51. Max. Input Voltage vs.
  • Fluke 190-504/AM | Product Manual - Page 141
    1 % 20 kHz to 1 MHz 2 % 1 MHz to 25 MHz 3 % For higher frequencies the probe's roll off starts affecting the accuracy. For further probe specifications see the instruction sheet supplied with the VPS410 probe set. 8 Specifications 10:1 Probe VPS410 129
  • Fluke 190-504/AM | Product Manual - Page 142
    ScopeMeter® Test Tool 190 Series II Users Manual Electromagnetic Immunity The Fluke 190 Series II test tools, including standard accessories are compliant to EN61326-1, with the addition of the following table Scope Mode (10 ms/div: Waveform disturbance
  • Fluke 190-504/AM | Product Manual - Page 143
    Appendices Appendix Title Page A Installing USB Drivers A-1 B Battery Pack MSDS B-1
  • Fluke 190-504/AM | Product Manual - Page 144
    ScopeMeter® Test Tool 190 Series II Users Manual
  • Fluke 190-504/AM | Product Manual - Page 145
    for installation on Win 7. Note The Fluke 190 Series II instrument requires two drivers to be loaded in sequence. − 1st it requires installation of the Fluke 190 ScopeMeter® USB driver − 2nd it requires installation of the Fluke USB Serial port Both of these drivers need to be installed to be able
  • Fluke 190-504/AM | Product Manual - Page 146
    ScopeMeter® Test Tool 190 Series II Users Manual Installing the USB Drivers To install the USB drivers, do the following: 1 Connect the Fluke 190 Series II instrument to the PC. The USB cable can be plugged in and out (hot-swap) when both the computer and the instrument are on. It is not required to
  • Fluke 190-504/AM | Product Manual - Page 147
    to the internet, you need to load the CD-ROM, supplied with the ScopeMeter®, which contains the drivers. 3 Follow the instructions on screen. When the driver has finished installation click 'Finish' to complete the first step of the driver installation. A Appendices Installing USB Drivers A-3
  • Fluke 190-504/AM | Product Manual - Page 148
    ScopeMeter® Test Tool 190 Series II Users Manual 4 After completing the first step the New Hardware Wizard will start again to install the USB Serial Port Driver. Click 'Next' to install the software automatically. Windows will download the drivers automatically from the Windows Driver Distribution
  • Fluke 190-504/AM | Product Manual - Page 149
    to Expand the 'Universal Serial Bus controllers'. The 'Fluke 190 ScopeMeter®' should be listed here. From the device manager Fluke USB Serial Port COM(5)' should be listed here. Note that the COM port number may differ and is automatically assigned by Windows. A Appendices Installing USB Drivers
  • Fluke 190-504/AM | Product Manual - Page 150
    ScopeMeter® Test Tool 190 Series II Users Manual Notes 1) Sometimes application software may require a different port number. (for example in the range Com 1..4). In this situation the COM port number can be changed manually. To manually assign a different COM port number right click on 'Fluke USB
  • Fluke 190-504/AM | Product Manual - Page 151
    Li-ion Battery Pack Contact Fluke for a Battery Material Safety Data Sheet (MSDS) or Compliance Information. B Appendices Battery Pack MSDS Appendix B Battery Pack MSDS B-1
  • Fluke 190-504/AM | Product Manual - Page 152
    ScopeMeter® Test Tool 190 Series II Users Manual B-2
  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122
  • 123
  • 124
  • 125
  • 126
  • 127
  • 128
  • 129
  • 130
  • 131
  • 132
  • 133
  • 134
  • 135
  • 136
  • 137
  • 138
  • 139
  • 140
  • 141
  • 142
  • 143
  • 144
  • 145
  • 146
  • 147
  • 148
  • 149
  • 150
  • 151
  • 152

May 2011, Rev.2, 1/14
© 2011-2014 Fluke Corporation. All rights reserved. Specifications are subject to change without notice.
All product names are trademarks of their respective companies.
ScopeMeter
®
Test Tool 190 Series II
Fluke 190-062, -102, -104, -202, -204, -502, -504
Users Manual