Fluke 9132-156 Product Manual

Fluke 9132-156 Manual

Fluke 9132-156 manual content summary:

  • Fluke 9132-156 | Product Manual - Page 1
    9132 Portable IR Calibrator User's Guide Rev. 672501 ENG MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9132-156 | Product Manual - Page 2
    period begins on the date of the shipment. Parts, product repairs, and services are warranted for 90 days. The warranty extends only to the original buyer extend a greater or different warranty on behalf of Hart. Warranty support is available if product is purchased through a Hart authorized sales
  • Fluke 9132-156 | Product Manual - Page 3
    1 Before You Start 1 1.1 Introduction 1 1.2 Symbols Used 1 1.3 Safety Information 3 1.3.1 WARNINGS...3 1.3.2 CAUTIONS...4 1.4 Authorized Service Centers 5 2 Specifications and Environmental Conditions 7 2.1 Specifications 7 2.2 Environmental Conditions 8 3 Quick Start 9 3.1 Unpacking
  • Fluke 9132-156 | Product Manual - Page 4
    Procedure 29 8.2.1 Compute DELTA 29 8.2.2 Compute R0 and ALPHA 30 8.2.3 Accuracy and Repeatability 31 9 Maintenance 33 10 Troubleshooting 35 10.1 Troubleshooting Problems, Possible Causes, and Solutions 35 10.2 Comments 36 10.2.1 EMC Directive 36 MyFlukeStore Shop for Fluke products online
  • Fluke 9132-156 | Product Manual - Page 5
    10.2.2 Low Voltage Directive (Safety 36 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9132-156 | Product Manual - Page 6
    sensors and devices. The user should be familiar with the safety guidelines and operating procedures of the calibrator as described in this user guide. 1.2 Symbols Used Table 1 lists the International Electrical Symbols. Some or all of these symbols may be used on the instrument or in this
  • Fluke 9132-156 | Product Manual - Page 7
    (Alternating Current) AC-DC Battery Complies with European Union directives DC Double Insulated Electric Shock Fuse PE Ground Hot Surface (Burn Hazard) Read the User's Guide (Important Information) Off On  MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9132-156 | Product Manual - Page 8
    Waste Electrical and Electronic Equipment (WEEE) Directive (2002/96/ EC) mark. 1.3 Safety Information Use this instrument only as specified in this manual. Otherwise, the protection provided by the instrument may be impaired. The following definitions apply to the terms "Warning" and "Caution
  • Fluke 9132-156 | Product Manual - Page 9
    listed in the user's guide. ● DO NOT operate this occurs, contact an Authorized Hart Scientific Service Center (see Section ). Always replace the fuse safety guidelines listed in the user's manual. ● Calibration Equipment should only be for optimum durability and trouble free operation, it must
  • Fluke 9132-156 | Product Manual - Page 10
    2 Specifications and Environmental Conditions 2 Specifications and Environmental Conditions 2.1 Specifications Table 2 Specifications Range Accuracy Stability Target Size Resolution Display Heating Times Cooling Times Stabilization Time Target Emissivity Aperture Diameter Computer Interface Power
  • Fluke 9132-156 | Product Manual - Page 11
    for optimum durability and trouble-free operation, it must be handled with care. The instrument should not be operated in an excessively dusty or dirty environment. Maintenance and cleaning recommendations can be found in the Maintenance section of this manual. The instrument operates safely
  • Fluke 9132-156 | Product Manual - Page 12
    . If there is shipping damage, notify the carrier immediately. Verify that the following components are present: ● 9132 Calibrator ● Power Cord ● User's Guide ● Serial Cable ● 9930 Software 3.2 Set Up Place the calibrator on a flat surface with at least 6 inches of free space around the instrument
  • Fluke 9132-156 | Product Manual - Page 13
    9132 Portable IR Calibrator User's Guide When the set-point temperature is changed the controller switches the well heater on or off to raise or lower the temperature. The displayed well
  • Fluke 9132-156 | Product Manual - Page 14
    for 115 VAC (±10%) or 230 VAC (±10%) operation. Heater Voltage Switch - To be used only when changing the input voltage. (See Section 7.2 for instructions on changing the input voltage.) CAUTION: Do not plug the unit into 230 V if the heater switches and fuse holder read 115 V. This action will
  • Fluke 9132-156 | Product Manual - Page 15
    9132 Portable IR Calibrator User's Guide Fan - The fan inside the calibrator has two speeds and runs continuously when the unit is being operated to provide cooling for the instrument. The
  • Fluke 9132-156 | Product Manual - Page 16
    4 Parts and Controls UP - Used to increment the displayed value. EXIT - Used to exit a function and to skip to the next function. Any changes made to the displayed value are ignored. Target Assembly - The target assembly is 57 mm (2.25 in) in diameter and has an emisivity of 0.95. 13 MyFlukeStore
  • Fluke 9132-156 | Product Manual - Page 17
    5 General Operation 5 General Operation 5.1 Changing Display Units The Model 9132 can display temperature in Celsius or Fahrenheit. The temperature units are shipped from the factory set to Celsius. There are two ways to change to Fahrenheit or back to Celsius as described below. 1. Press the "SET"
  • Fluke 9132-156 | Product Manual - Page 18
    6 Controller Operation 6 Controller Operation This chapter discusses in detail how to operate the instrument temperature controller using the front control panel. Using the front panel key-switches and LED display the user may monitor the well temperature, set the temperature set-point in degrees C
  • Fluke 9132-156 | Product Manual - Page 19
    9132 Portable IR Calibrator User's Guide Figure 4 Controller Operation Flowchart 18 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
  • Fluke 9132-156 | Product Manual - Page 20
    6 Controller Operation Press "SET" to accept the new selection and access the set-point value. S Accept selected set-point memory. 6.2.2 Set-point Value The set-point value may be adjusted after selecting the set-point memory and pressing "SET". 4. 200. Set-point 4 value in °C If the set-
  • Fluke 9132-156 | Product Manual - Page 21
    9132 Portable IR Calibrator User's Guide Press "SET" to accept the present setting and continue. S Accept scan setting 6.4.2 Scan Rate The next function in the main menu is the scan rate.
  • Fluke 9132-156 | Product Manual - Page 22
    6 Controller Operation The heater power display is accessed in the secondary menu. Press "SET" and "EXIT" simultaneously and release. The heater power is displayed as a percentage of full power. 100.0 C Well temperature S+E Access heater power in secondary menu SEC Flashes "SEC" for secondary
  • Fluke 9132-156 | Product Manual - Page 23
    9132 Portable IR Calibrator User's Guide 4.1 Proportional band setting To change the proportional band press "UP" or "DOWN". 10.0 New proportional band setting To accept the new setting press "SET". Press "
  • Fluke 9132-156 | Product Manual - Page 24
    6 Controller Operation 6.12.1 Baud Rate The baud rate is the first parameter in the menu. The baud rate setting determines the serial communications transmission rate. bAUd Flashes bAUd and then displays the setting 2400 b Current baud rate The baud rate of the serial communications may be
  • Fluke 9132-156 | Product Manual - Page 25
    the user may restore these values to the factory settings. The user should have a list of these constants and their settings with the instrument manual. CAUTION: DO NOT change the values of the instrument calibration constants from the factory set values. The correct setting of these parameters is
  • Fluke 9132-156 | Product Manual - Page 26
    7 Digital Communications 7 Digital Communications The calibrator is capable of communicating with and being controlled by other equipment through the digital serial interface. With a digital interface the instrument may be connected to a computer or other equipment. This allows the user to set the
  • Fluke 9132-156 | Product Manual - Page 27
    9132 Portable IR Calibrator User's Guide 7.2.1.1 Baud Rate The baud rate is the first parameter in the menu. The display prompts with the baud rate parameter by showing "bAUd". Press "SET"
  • Fluke 9132-156 | Product Manual - Page 28
    7 Digital Communications is sent with the command following a "=". For example, "s=150.0" sets the set-point to 150.0 degrees. In the following list of commands, characters or data within brackets, "[" and "]" are optional for the command. A slash, "/", denotes alternate characters or data. Numeric
  • Fluke 9132-156 | Product Manual - Page 29
    9132 Portable IR Calibrator User's Guide Table 3 Serial Commands Command Description Display Temperature Read current set-point Set current set-point to n Command Format s[etpoint] s[etpoint]=n Command Example s s=200.0 Returned Returned
  • Fluke 9132-156 | Product Manual - Page 30
    8 Calibration Procedure 8 Calibration Procedure Sometimes the user may want to calibrate the instrument to improve the temperature setpoint accuracy. Calibration is done by adjusting the controller probe calibration constants R0, ALPHA, and DELTA so that the temperature of the dry-well as measured
  • Fluke 9132-156 | Product Manual - Page 31
    9132 Portable IR Calibrator User's Guide T1-3 - Measured temperature using a thermometer R1-3 - Value of the set-point resistance from the display. (Press SET and DOWN at the same time.) Where: T1
  • Fluke 9132-156 | Product Manual - Page 32
    8 Calibration Procedure 8.2.3 Accuracy and Repeatability Check the accuracy of the instrument at various points over the calibration range. If instrument does not pass specification at all set-points, repeat the Calibration Procedure. 31 MyFlukeStore Shop for Fluke products online at: www. .com
  • Fluke 9132-156 | Product Manual - Page 33
    a cord with the appropriate gauge wire for the current of the instrument. If there are any questions, call an Authorized Hart Scientific Service Center (see Section ) for more information. ● Before using any cleaning or decontamination method except those recommended by Hart, users should check with
  • Fluke 9132-156 | Product Manual - Page 34
    and solve the problem. If the problem cannot otherwise be solved, contact an Authorized Service Center (see Section 1.4). Be sure to have the model number, serial number, and voltage of your instrument available. Table 4 Troubleshooting Problems Problem Incorrect temperature reading Possible
  • Fluke 9132-156 | Product Manual - Page 35
    10.2 Comments 11.2.1 10.2.1 EMC Directive Hart Scientifics' equipment has been tested to meet the European Electromagnetic Compatibility Directive (EMC Directive, 89/336/EEC). The Declaration of Conformity for your instrument lists the specific standards to which the unit was tested. 11.2.2 10.2.2
  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35

Rev. 672501 ENG
9132
Portable IR Calibrator
User’s Guide
Shop for Fluke products online at:
1.888.610.7664
www.
MyFlukeStore
.com