Fluke 975 FE 975 Users Manual

Fluke 975 Manual

Fluke 975 manual content summary:

  • Fluke 975 | FE 975 Users Manual - Page 1
    PN 2507213 August 2006 © 2006 Fluke Corporation. All rights reserved. Printed in U.S.A. All product names are trademarks of their respective companies. ® 975 AirMeterTM Test Tool Users Manual
  • Fluke 975 | FE 975 Users Manual - Page 2
    not warrant that software will be error free or operate without interruption. Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available
  • Fluke 975 | FE 975 Users Manual - Page 3
    Table of Contents Title Page Introduction...1 Features ...1 Contacting Fluke ...2 Safety Information ...2 Symbols ...3 Shipping Contents ...4 Using the Meter ...6 Using the Meter ...6 Softkeys and Pushbuttons 6 Meter Power ...7 Meter Power ...7 Startup and Self Test 8 Automatic Backlight 9
  • Fluke 975 | FE 975 Users Manual - Page 4
    975 Users Manual CO Alarm ...10 The Setup Menu ...10 Taking Measurements 11 Temperature, Relative Humidity, CO, CO , Dew Point, and Wet Bulb 2 Measurements ...11 % Outside Air ...11 % Outside Air (Temperature 12 % Outside Air (CO 12 2 Air Velocity and Volume Flow Rate 13 Standard Velocity and
  • Fluke 975 | FE 975 Users Manual - Page 5
    Contents (continued) Changing the AA Batteries 25 General Specifications 27 Calculated Parameters 28 Environmental Specifications 29 Physical Specifications 29 Agency Approvals, Certifications and Standards Compliance 30 Replaceable Parts and Accessories 30 iii
  • Fluke 975 | FE 975 Users Manual - Page 6
    975 Users Manual iv
  • Fluke 975 | FE 975 Users Manual - Page 7
    List of Tables Table Title Page 1. Symbols...3 2. 975 Shipping Contents ...5 3. Softkeys and Pushbuttons...6 4. Startup Failure Messages...9 5. Replaceable Parts and Accessories 32 v
  • Fluke 975 | FE 975 Users Manual - Page 8
    975 Users Manual vi
  • Fluke 975 | FE 975 Users Manual - Page 9
    List of Figures Figure Title Page 1. 975 Shipping Contents ...4 2. Softkeys and Pushbuttons...6 3. AC Adapter Input and USB Port 8 4. Calibration Set Up ...23 5. AA Battery Replacement and Kensington Lock Use 26 6. Replaceable Parts and Accessories 31 vii
  • Fluke 975 | FE 975 Users Manual - Page 10
    975 Users Manual viii
  • Fluke 975 | FE 975 Users Manual - Page 11
    975 AirMeter TM Test Tool Introduction WWarning Read "Safety Information" before using the 975 AirMeterTM Test Tool. The Fluke 975 AirMeterTM Test Tool ("the Meter") is a hand-held, five-sensor indoor air quality diagnostic instrument. It measures, calculates, and displays: • Temperature, dew
  • Fluke 975 | FE 975 Users Manual - Page 12
    975 Users Manual • Single-point and continuous data logging options • CO alarm • Detachable velocity probe (optional) • FlukeView Forms software with USB cable for downloading stored data • Hard shell carrying case • Protective holster • Calibration cap and tubing • Calibration certificate for NIST
  • Fluke 975 | FE 975 Users Manual - Page 13
    AirMeterTM Test Tool Symbols Symbols Symbols used in this manual and on the Meter are displayed in Table 1. Table 1. Symbols Symbol Meaning W Risk of danger. Important information. Refer to manual. X Hazardous voltage. Risk of electrical shock. P ; N1040 Conforms to relevant European
  • Fluke 975 | FE 975 Users Manual - Page 14
    975 Users Manual Shipping Contents The Meter ships with the items shown in Figure 1 and listed in Table 2. 2 1 4 F1 F1 F1 VELOCITY PROBE 9 4 3 6 5 12 10 7 11 Figure 1. 975 Shipping Contents 8 eba07f.eps
  • Fluke 975 | FE 975 Users Manual - Page 15
    Manuals CD containing the Users Manual in B AC Adapter/Charger with type A, C, G and I adapters to fit most countries. H English, French, German, Portuguese, Spanish, and Simplified Chinese C Three AA batteries (backup power supply) D USB cable I CO and CO2 calibration cap J Velocity probe
  • Fluke 975 | FE 975 Users Manual - Page 16
    975 Users Manual Using the Meter The following sections explain how to use the Meter and its features. Air- Activates the % Outside Air menu. See "% Outside Air". V Velocity-When used with the optional Velocity Probe, this activates the Air Velocity and Volume Flow Rate menus. See "Air Velocity
  • Fluke 975 | FE 975 Users Manual - Page 17
    line cord can be used. AirMeterTM Test Tool Using the Meter / Note This Meter contains a lithium battery. Do not mix with the solid waste stream. Spent batteries should be disposed of by a qualified recycler or hazardous materials handler. Contact your authorized Fluke Service Center for recycling
  • Fluke 975 | FE 975 Users Manual - Page 18
    975 Users Manual USB Port AC Adapter Input eba02f.eps Figure 3. AC Adapter Input and USB Port 8 Startup and Self Test To turn the Meter on, press and hold P for a few seconds and wait for the flashing red LED on the left side of the display. Press P for 2 seconds to turn the Meter OFF. Once
  • Fluke 975 | FE 975 Users Manual - Page 19
    Maintenance". Problem with the Velocity probe or the connection. Remove and reattach the probe. If the problem continues, have the probe serviced. Calibrate the Meter. Calibration date for the Meter has passed. See "Calibration". AirMeterTM Test Tool Using the Meter Note The Meter settling time
  • Fluke 975 | FE 975 Users Manual - Page 20
    975 Users Manual Multi-Language Interface The Meter's display supports five languages: English, French, German, Portuguese, and Spanish. Use the Setup menu to change the display language. See "The Setup Menu". Measurement Units The Meter supports both Metric and US measurement units. Select the
  • Fluke 975 | FE 975 Users Manual - Page 21
    Humidity (%RH) • Carbon monoxide (CO) levels • Carbon dioxide (CO2) levels • % Outside Air • Air Velocity (standard and actual) • Flow Rate Temperature, Relative Humidity, CO, CO2, Dew Point, and Wet Bulb Measurements When the Meter completes and passes its self-test, it is ready to take temperature
  • Fluke 975 | FE 975 Users Manual - Page 22
    975 Users Manual variables are known, the Meter computes the % outside air. % Outside Air (Temperature) To obtain % of outside air using temperature: 1. Place the Meter sensor perpendicular to the air stream being measured. 2. Press Q.The Meter switches to % outside air measuring mode and
  • Fluke 975 | FE 975 Users Manual - Page 23
    fails the self test, the velocity feature is disabled. The Meter alerts the user to either attach the probe or that there is a probe error. Note The velocity probe takes approximately one minute to warm up. Measuring Air Velocity Note To measure standard or actual air velocity the steps are the same
  • Fluke 975 | FE 975 Users Manual - Page 24
    is to multiply the area of the air-duct opening by the average velocity of the air. The Meter does the computation after the variables are entered. To measure the volume flow rate: 1. Attach the velocity probe. V 2. Press to access the velocity menu. The probe will initialize. 3. Press 1[Volume Flow
  • Fluke 975 | FE 975 Users Manual - Page 25
    Relative Humidity, Dew Point, Wet Bulb, and Air Velocity. To use Min Max mode, press M. Meter or can be uploaded to a PC running FlukeView Forms software. Refer to "Transferring Data to a Personal Computer" for more information. AirMeterTM Test Tool Min Max Avg Saving Single-Point Data The Meter
  • Fluke 975 | FE 975 Users Manual - Page 26
    975 Users Manual To save single-point data: 1. Take a desired measurement. 2. Press L. The Log/Save menu appears on the display. The measurement that was displayed when Lwas pressed is the measurement that will be saved. 3. Press 3[Save Data]. The Meter 2 • Air velocity (if the probe is connected)
  • Fluke 975 | FE 975 Users Manual - Page 27
    %. 6. The Meter requests user confirmation before starting Meter logs the sessions until stopped or the logging interval has expired. To stop logging, press 2[Stop]. All other keys are inoperative while logging. AirMeterTM Meter off. Logging Button Lockout The Meter has an automatic and manual
  • Fluke 975 | FE 975 Users Manual - Page 28
    975 Users Manual Recalling or Erasing Logged Data Use the Recall function to 2[Previous] to move the cursor to the previous or next detailed sample. 4. If the velocity probe is attached, press 3 to access the velocity section of the sample. To view logged sessions: 1. Press R to show the Recall
  • Fluke 975 | FE 975 Users Manual - Page 29
    USB port on the PC. 3. Power up the Meter if it is not already on. 4. Windows should show that it's found new hardware and will ask for new drivers. Point to the FlukeView Forms CD. 5. Open FlukeView Forms Documenting Software. AirMeterTM Test Tool Transferring Data to a Personal Computer 6. The
  • Fluke 975 | FE 975 Users Manual - Page 30
    from one to 365 days. When the Meter reaches its calibration due date, it alerts the user but will continue to operate. The Meter's CO and CO2 sensors can be calibrated by the user or returned to Fluke for service. See "Contacting Fluke". The recommended calibration interval is 1 month for CO and
  • Fluke 975 | FE 975 Users Manual - Page 31
    date. Use AirMeterTM Test Tool Calibration 1[DOWN] and 2[UP] to change the number and 3[ENTER]to store the number. Note When either CO or CO sensors are calibrated, 2 the calibration date is reset by the user. CO calibration is now complete and the Meter exits calibration mode. Calibrating the CO2
  • Fluke 975 | FE 975 Users Manual - Page 32
    975 Users Manual If one-point is chosen: a. Use 1 [DOWN] and 2[UP] to choose the CAL gas concentration. b. Press 3[ENTER] to enter the concentration. c. Apply the CAL gas and press 2[OK], press 3 [SKIP] to exit calibration mode. The Meter will now calibrate the CO sensor for 2 minutes. 2 press 3 [
  • Fluke 975 | FE 975 Users Manual - Page 33
    AirMeterTM Test Tool Calibration F1 F1 F1 VPERLOOBCEITY Figure 4. Calibration Set Up eba09f.eps 23
  • Fluke 975 | FE 975 Users Manual - Page 34
    975 Users Manual Maintenance XWWarning The Meter contains no user-serviceable parts. To avoid electric shock, injury, or damage to the Meter, do not open the case. For service and rechargeable battery replacement, the Meter must be sent to Fluke. See "Contacting Fluke". Cleaning the Meter
  • Fluke 975 | FE 975 Users Manual - Page 35
    • )RC - Rechargeable battery charge is at 10 % of capacity (flashing). AirMeterTM Test Tool Maintenance Charging the Battery and Using the Power Adapter / Note This Meter contains a rechargeable lithium battery that cannot be serviced by the user. Do not mix this battery with the solid waste stream
  • Fluke 975 | FE 975 Users Manual - Page 36
    975 Users Manual Figure 5. AA Battery Replacement and Kensington Lock Use eba08f.eps 26
  • Fluke 975 | FE 975 Users Manual - Page 37
    AirMeterTM Test Tool General /±1.98 °F from -20 °C to 5 °C ±2 %RH from 10 %RH to 90 %RH Hysteresis Spec 1% Air Velocity 50 to 3000 fpm 0.25 to 15 m/sec 1 fpm 0.005 m/sec 4 % or 4 fpm* 95 % RH) *Note: After the CO calibration seal is removed and the Meter is stored for an extended period in arid
  • Fluke 975 | FE 975 Users Manual - Page 38
    975 Users Manual Calculated Parameters Calculated Parameters Range Display Resolution Dew Point Temperature -44 to 57 °C -47 to 135 °F 0.1 °C 0.1 °F Wet Bulb Temperature -16 to 57 °C 3 to 135 °F 0.1 °C 0.1 °F Volume Flow Rate (in a duct) % outside air (Based on Temperature) % Outside Air (
  • Fluke 975 | FE 975 Users Manual - Page 39
    : Humidity: Altitude: Shock and Vibration: Physical Specifications Meter size: Meter weight: Hard-shell case size: Velocity probe size: Velocity probe weight: Impact resistance: Power supply: Backup power supply: Battery life: AirMeterTM Test Tool General Specifications -20 to 50 °C (-4 to
  • Fluke 975 | FE 975 Users Manual - Page 40
    975 Users Manual Agency Approvals, Certifications and Standards Compliance The Meter complies with the following standards and meets requirements for the following certifications: EMC: EN61326-1. Class B FCC Part 15 Class B Industry Canada: ICES-003 Class B UL
  • Fluke 975 | FE 975 Users Manual - Page 41
    1 2 3 4 AirMeterTM Test Tool Replaceable Parts and Accessories 5 6 7 9 10 8 12 13 14 15 11 Figure 6. Replaceable Parts and Accessories eba10f.eps 31
  • Fluke 975 | FE 975 Users Manual - Page 42
    975 Users Manual Item Number A B C D Part Number or Model Number 2514351 2514336 2514349 2664162 E 2664357 F 2526937 G Air Velocity Probe H Calibration Kit I 2679599 J 2665668 K 2679501 L TPAK80-4 M 1578384 N 2507224 O 2507213 Table 5. Replaceable Parts and Accessories
  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42

®
975
AirMeter
TM
Test Tool
Users Manual
PN 2507213
August 2006
© 2006 Fluke Corporation. All rights reserved. Printed in U.S.A.
All product names are trademarks of their respective companies.