116 Users Manual Measuring Frequency XWWarning To avoid electrical shock, disregard the bar graph for frequencies >1 kHz. If the frequency of the measured signal is >1 kHz, the bar graph and Z are unspecified. AC Voltage Frequency Hz Hz V The Meter measures the frequency of a signal by counting the number of times the signal crosses a trigger level each second. The trigger level is 0 V for all ranges. 1. Press g to turn the frequency measurement function on and off. Frequency works with ac volts only. 2. In frequency, the bar graph and range annunciator indicate the AC voltage present. 3. Select progressively lower ranges using manual ranging for a stable reading. Making Low Impedance Capacitance Measurements For making capacitance measurements on cables with ghost voltage, hold qwhile turning on the Meter to switch the Meter into LoZ, (low input impedance) Capacitance mode. In this mode, capacitance measurements will have a lower accuracy and lower dynamic range. This setting is not saved when the Meter is turned off or goes into sleep mode. edy09f.eps 14

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24

116
Users Manual
14
Measuring Frequency
XW
Warning
To avoid electrical shock, disregard
the bar graph for frequencies >1 kHz.
If the frequency of the measured
signal is >1 kHz, the bar graph
and
Z
are
unspecified.
AC Voltage Frequency
Hz
V
Hz
edy09f.eps
The Meter measures the frequency of a signal by counting
the number of times the signal crosses a trigger level each
second. The trigger level is 0 V for all ranges.
1.
Press
g
to turn the frequency measurement
function on and off. Frequency works with ac volts only.
2.
In frequency, the bar graph and range annunciator
indicate the AC voltage present.
3.
Select progressively lower ranges using manual ranging
for a stable reading.
Making Low Impedance Capacitance
Measurements
For making capacitance measurements on cables with
ghost voltage, hold
q
while turning on the Meter to
switch the Meter into LoZ, (low input impedance)
Capacitance mode. In this mode, capacitance
measurements will have a lower accuracy and lower
dynamic range. This setting is not saved when the Meter is
turned off or goes into sleep mode.