Fluke 175 FE 175,177,179 Users Manual - Page 16

Accuracy, Function, Range, Resolution, Model 175, For non-sinusoidal waveforms - true rms

Page 16 highlights

Models 175, 177 & 179 Users Manual Function Range 1 Resolution Accuracy ± ( [ % of Reading ] + [ Counts ] ) Model 175 Model 177 Model 179 AC Volts 2,3 600.0 mV 6.000 V 60.00 V 600.0 V 1000 V 0.1 mV 0.001 V 0.01 V 0.1 V 1 V 1.0 % + 3 (45 Hz to 500 Hz) 1.0 % + 3 (45 Hz to 500 Hz) 1.0 % + 3 (45 Hz to 500 Hz) 2.0 % + 3 (500 Hz to 1 kHz) 2.0 % + 3 (500 Hz to 1 kHz) 2.0 % + 3 (500 Hz to 1 kHz) DC mV 600.0 mV 0.1 mV 0.15 % + 2 0.09 % + 2 0.09 % + 2 DC Volts 6.000 V 60.00 V 600.0 V 0.001 V 0.01 V 0.1 V 0.15 % + 2 0.09 % + 2 0.09 % + 2 1000 V 1 V 0.15 % + 2 0.15 % + 2 0.15 % + 2 Continuity Ohms 600 Ω 600.0 Ω 6.000 kΩ 60.00 kΩ 600.0 kΩ 6.000 MΩ 50.00 MΩ 1 Ω 0.1 Ω 0.001 kΩ 0.01 kΩ 0.1 kΩ 0.001 MΩ 0.01 MΩ Meter beeps at < 25 Ω, beeper turns off at > 250 Ω; detects opens or shorts of 250 µs or longer. 0.9 % + 2 0.9 % + 1 0.9 % + 1 0.9 % + 1 0.9 % + 1 1.5 % + 3 0.9 % + 2 0.9 % + 1 0.9 % + 1 0.9 % + 1 0.9 % + 1 1.5 % + 3 0.9 % + 2 0.9 % + 1 0.9 % + 1 0.9 % + 1 0.9 % + 1 1.5 % + 3 Diode test 2.400 V 0.001 V 1 % + 2 Capacitance 1000 nF 10.00 µF 100.0 µF 9999 µF 4 1 nF 0.01 µF 0.1 µF 1 µF 1.2 % + 2 1.2 % + 2 1.2 % + 2 10 % typical AC Amps 5 (True RMS) (45 Hz to 1 kHz) 60.00 mA 400.0 mA 6.000 A 10.00 A 0.01 mA 0.1 mA 0.001 A 0.01 A 1.5 % + 3 1. All AC voltage and AC current ranges are specified from 5 % of range to 100 % of range. 2. Crest factor of ≤ 3 at full scale up to 500 V, decreasing linearly to crest factor ≤ 1.5 at 1000 V. 3. For non-sinusoidal waveforms, add -(2% reading + 2% full scale) typical, for crest factors up to 3. 4. In the 9999 µF range for measurements to 1000 µF, the measurement accuracy is 1.2 % + 2 for all models. 5. Amps input burden voltage (typical): 400 mA input 2 mV/mA, 10 A input 37 mV/A. 1.2 % + 2 1.2 % + 2 1.2 % + 2 10 % typical 1.5 % + 3 1.2 % + 2 1.2 % + 2 1.2 % + 2 10 % typical 1.5 % + 3 12

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18

Models 175, 177 & 179
Users Manual
12
Accuracy
± ( [ % of Reading ] + [ Counts ] )
Function
Range
1
Resolution
Model 175
Model 177
Model 179
AC Volts
2,3
600.0 mV
6.000 V
60.00 V
600.0 V
1000 V
0.1 mV
0.001 V
0.01 V
0.1 V
1 V
1.0 % + 3
(45 Hz to 500 Hz)
2.0 % + 3
(500 Hz to 1 kHz)
1.0 % + 3
(45 Hz to 500 Hz)
2.0 % + 3
(500 Hz to 1 kHz)
1.0 % + 3
(45 Hz to 500 Hz)
2.0 % + 3
(500 Hz to 1 kHz)
DC mV
600.0 mV
0.1 mV
0.15 % + 2
0.09 % + 2
0.09 % + 2
DC Volts
6.000 V
60.00 V
600.0 V
0.001 V
0.01 V
0.1 V
0.15 % + 2
0.09 % + 2
0.09 % + 2
1000 V
1 V
0.15 % + 2
0.15 % + 2
0.15 % + 2
Continuity
600
1
Meter beeps at < 25
, beeper turns off at > 250
; detects
opens or shorts of 250
μ
s or longer.
Ohms
600.0
6.000 k
60.00 k
600.0 k
6.000 M
50.00 M
0.1
0.001 k
0.01 k
0.1 k
0.001 M
0.01 M
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 3
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 3
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 3
Diode test
2.400 V
0.001 V
1 % + 2
Capacitance
1000 nF
10.00
μ
F
100.0
μ
F
9999
μ
F
4
1 nF
0.01
μ
F
0.1
μ
F
1
μ
F
1.2 % + 2
1.2 % + 2
1.2 % + 2
10 % typical
1.2 % + 2
1.2 % + 2
1.2 % + 2
10 % typical
1.2 % + 2
1.2 % + 2
1.2 % + 2
10 % typical
AC Amps
5
(True RMS)
(45 Hz to 1 kHz)
60.00 mA
400.0 mA
6.000 A
10.00 A
0.01 mA
0.1 mA
0.001 A
0.01 A
1.5 % + 3
1.5 % + 3
1.5 % + 3
1.
All AC voltage and AC current ranges are specified from 5 % of range to 100 % of range.
2.
Crest factor of
3 at full scale up to 500 V, decreasing linearly to crest factor
1.5 at 1000 V.
3.
For non-sinusoidal waveforms, add -(2% reading + 2% full scale) typical, for crest factors up to 3.
4.
In the 9999
μ
F range for measurements to 1000
μ
F, the measurement accuracy is 1.2 % + 2 for all models.
5.
Amps input burden voltage (typical): 400 mA input 2 mV/mA, 10 A input 37 mV/A.