Changing the Auto Set Options, User Options, Auto Set Adjust, Signal > 1 Hz, Coupling, Unchanged
View all Fluke 196B/S manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 89 highlights
Changing the Auto Set Options With the next procedure you can choose how auto set behaves when you press the AUTO (auto set) key. 1 Display the USER key labels. 2 Open the User Options menu. 3 Open the Auto Set Adjust menu. If the frequency range is set to > 15 Hz, the Connect-andView function responds more quickly. The response is quicker because the test tool is instructed not to analyze low frequency signal components. However, when you measure frequencies lower than 15 Hz, the test tool must be instructed to analyze low frequency components for automatic triggering: 7 Tips Changing the Auto Set Options 4 Select Signal > 1 Hz, then jump to Coupling. With the coupling option you can choose how auto set behaves. When you press the AUTO (auto set) key, the coupling can either be set to dc or left unchanged: 5 Select Unchanged. Note The auto set option for the signal frequency is similar to the automatic trigger option for the signal frequency. (See Chapter 5: "Automatic Trigger Options"). However, the auto set option determines the behavior of the auto set function and shows only effect when you press the auto set key. 81