Fluke 724 Fluke 712, 714, and 724 Pressure Calibrator Datasheet - Page 1

Fluke 724 Manual

Page 1 highlights

Fluke 712, 714 and 724 Temperature Calibrators Accurate and rugged, temperature calibration tools. The Fluke 710 Series temperature calibrators deliver outstanding performance, durability and reliability. Offered in the rugged Fluke 80 Series DMM package, the calibrators are compact, lightweight, and easy to carry. And with a push-button interface, similar to the multifunction Fluke 740 Series Documenting Process Calibrators, the 710s are easy to use. Each calibrator is EMI tolerant, dust- and splash-resistant, and features a removable battery door for quick battery changes. Temperature Calibration Data Sheet (Literature code 1560369) See how Fluke can help you calibrate, troubleshoot and verify the most common temperature devices, faster and easier. Fluke 712 RTD Calibrator • Compatible with pulsed current transmitters • Measure temperature from an RTD probe • Simulate RTD output • Operates with seven types of RTD • Measure additional RTDs using Ohms measurement function • Simulate additional RTDs using Ohms source function • °F or °C selectable • Four shrouded banana jacks Fluke 714 Thermocouple Calibrator Fluke 724 Temperature • Measure temperature from TC probes Calibrator • Simulate TC output • Source/measure TCs, RTDs, volts • Operable with nine types of thermocouples • Calibrate linear TC transmitter with mV source function and ohms • Measure mA while supplying • Selectable °F or °C • Thermocouple mini-jack termination loop power • 25 % and 100 % stepping, 26 • Available as accessories: Fluke 700TC1 and TC2 Thermocouple Mini-plug Kits auto-step and auto ramp • For specifications, see page 29 Specifications - 712 and 714 Calibrators Model Fluke-712 Function Measure/simulate RTD Range -200 °C to 800 °C (Pt 100-385) Resolution 0.1 °C, 0.1 °F Measure/simulate Resistance 15 Ω to 3200 Ω 0.1 Ω Accuracy 0.33 °C, 0.6 °F (Pt 100-385) 0.1 Ω to 1 Ω Notes Pt; 100 200 500 1000 (385); Pt 100 (392); Pt 100 (392) JIS; Ni 120 (672) Fluke-714 Measure/simulate Thermocouple Measure/simulate mV -200 °C to 1800 °C, depending on type (K, -200 °C to 1370 °C) 0.1 °C or °F (1 °C or °F; BRS) -10 to 75 mV 0.01 mV 0.3 °C + 10 µV 0.025 % + 1 count 9 TC types; J K T E R S B per NIST 175 and ITS-90 L U per DIN 43710 and PTS-68 General specifications for all Fluke 712 through 718 calibrators Maximum voltage: 30 V Non-operating temperature: -40 °C to 60 °C Operating temperature: -10 °C to 55 °C Relative humidity: 95 % (10 °C to 30 °C); 75 % (30 °C to 40 °C); 45 % (40 °C to 50 °C); 35 % (50 °C to 55 °C) Operating altitude: 3,000 m max Shock: 1m drop test Vibration: Random, 2 g, 5-500 Hz Safety: CSA C22.2 No. 1010.1:1992 EMC: EN50082-1:1992 and EN55022:1994 Class B Size/weight (HxWxD) (712-717): 187 mm x 87 mm x 32 mm (7.35 in x 3.41 in x 1.25 in) 330 g (12 oz) Size/weight (HxWxD) (712-717 with holster and FlexStand™): 201 mm x 98 mm x 52 mm (7.93 in x 3.86 in x 2.06 in) 600 g (21 oz) Size/weight (HxWxD) (718): 210 mm x 83 mm x 62 mm (8.25 in x 3.27 in x 2.44 in) 737 g (26 oz) Size/weight (HxWxD) (718 with holster): 216 mm x 94 mm x 66 mm (8.50 in x 3.72 in x 2.60 in) 992 g (35 oz) Power: 9 V battery ANSI/NEDA 1604A or IEC 6LR619V alkaline; two batteries in 718 Battery life: 4 to 20 hours, typical, depending on functions used Warranty: Three-years (one year on pressure pump in Fluke 718) Recommended accessories - 712, 714 and 724 Calibrators TL75 (712 and 714) C550 (712 and 714) TL220 (712 and 714) C25 (712 and 714) See page 39 See page 44 See page 39 See page 44 80PK-24 (714) See page 42 TL75 (724) See page 39 TL220 (724) See page 39 C100 (724) See page 44 80PK-3A (724) See page 42 80PK-IR (724) See page 42 Included accessories Each 71X calibrator includes: C81Y Protective Yellow Holster with Flex-Stand™, test leads and alligator clips (excluding model 714), single 9 V alkaline battery and instruction sheet (14 languages). Ordering information Fluke-712 RTD Calibrator Fluke-714 Themocouple Calibrator Fluke-724 Temperature Calibrator (also see page 29) For more information, go to wwww.fwlukwe..cMomy/teFmlpuckaleStore.com

  • 1

Recommended accessories – 712, 714 and 724 Calibrators
Specifications – 712 and 714 Calibrators
For more information, go to www.fluke.com/tempcal
Fluke 712, 714 and 724
Temperature Calibrators
Accurate and rugged,
temperature calibration tools.
Included accessories
Each 71X calibrator includes: C81Y
Protective Yellow Holster with
Flex-Stand
, test leads and alligator
clips (excluding model 714), single
9 V alkaline battery and instruction
sheet (14 languages).
Ordering information
Fluke-712
RTD Calibrator
Fluke-714
Themocouple Calibrator
Fluke-724
Temperature Calibrator
(also see page 29)
The Fluke 710 Series temperature calibrators deliver out-
standing performance, durability and reliability. Offered
in the rugged Fluke 80 Series DMM package, the calibra-
tors are compact, lightweight, and easy to carry. And
with a push-button interface, similar to the multifunction
Fluke 740 Series Documenting Process Calibrators, the
710s are easy to use. Each calibrator is EMI tolerant,
dust- and splash-resistant, and features a removable
battery door for quick battery changes.
Fluke 712 RTD Calibrator
Compatible with pulsed current transmitters
Measure temperature from an RTD probe
Simulate RTD output
Operates with seven types of RTD
Measure additional RTDs using Ohms measurement
function
Simulate additional RTDs using Ohms source function
°F or °C selectable
Four shrouded banana jacks
Fluke 714 Thermocouple Calibrator
Measure temperature from TC probes
Simulate TC output
Operable with nine types of thermocouples
Calibrate linear TC transmitter with mV source function
Selectable °F or °C
Thermocouple mini-jack termination
Available as accessories: Fluke 700TC1 and TC2
Thermocouple Mini-plug Kits
Model
Function
Range
Resolution
Accuracy
Notes
Fluke-712
Measure/simulate
-200 °C to 800 °C
0.1 °C, 0.1 °F
0.33 °C, 0.6 °F
Pt; 100 200 500 1000
RTD
(Pt 100-385)
(Pt 100-385)
(385); Pt 100 (392);
Pt 100 (392) JIS;
Ni 120 (672)
Measure/simulate
15
Ω
to 3200
Ω
0.1
Ω
0.1
Ω
to 1
Ω
Resistance
Fluke-714
Measure/simulate
-200 °C to 1800 °C,
0.1 °C or °F
0.3 °C + 10 μV
9 TC types;
Thermocouple
depending
(1 °C or °F; BRS)
J K T E R S B per NIST
on type
175 and ITS-90
(K, -200 °C to
L U per DIN 43710 and
1370 °C)
PTS-68
Measure/simulate
-10 to 75 mV
0.01 mV
0.025 %
mV
+ 1 count
General specifications for all Fluke 712 through 718 calibrators
Maximum voltage:
30 V
Non-operating temperature:
-40 °C to 60 °C
Operating temperature:
-10 °C to 55 °C
Relative humidity:
95 % (10 °C to 30 °C); 75 % (30 °C to 40 °C); 45 % (40 °C to 50 °C); 35 % (50 °C to 55 °C)
Operating altitude:
3,000 m max
Shock:
1m drop test
Vibration:
Random, 2 g, 5-500 Hz
Safety:
CSA C22.2 No. 1010.1:1992
EMC:
EN50082-1:1992 and EN55022:1994 Class B
Size/weight (HxWxD) (712-717):
187 mm x 87 mm x 32 mm (7.35 in x 3.41 in x 1.25 in) 330 g (12 oz)
Size/weight (HxWxD) (712-717 with holster and FlexStand
):
201mm x 98 mm x 52 mm (7.93 in x 3.86 in x 2.06 in) 600 g (21 oz)
Size/weight (HxWxD) (718):
210 mm x 83 mm x 62 mm (8.25 in x 3.27 in x 2.44 in) 737 g (26 oz)
Size/weight (HxWxD) (718 with holster):
216 mm x 94 mm x 66 mm (8.50 in x 3.72 in x 2.60 in) 992 g (35 oz)
Power:
9 V battery ANSI/NEDA 1604A or IEC 6LR619V alkaline; two batteries in 718
Battery life:
4 to 20 hours, typical, depending on functions used
Warranty:
Three-years (one year on pressure pump in Fluke 718)
TL75 (712 and 714)
C550 (712 and 714)
TL220 (712 and 714)
C25 (712 and 714)
80PK-24 (714)
See page 39
See page 44
See page 39
See page 44
See page 42
TL75 (724)
TL220 (724)
C100 (724)
80PK-3A (724)
80PK-IR (724)
See page 39
See page 39
See page 44
See page
42
See page 42
Fluke 724 Temperature
Calibrator
Source/measure TCs, RTDs, volts
and ohms
Measure mA while supplying
loop power
25 % and 100 % stepping,
auto-step and auto ramp
For specifications, see page 29
Temperature Calibration Data Sheet
(Literature code 1560369)
See how Fluke can help you
calibrate, troubleshoot and verify
the most common temperature
devices, faster and easier.
26