Fluke 789 FE 789 Users Manual - Page 34
Auto Ramping the mA Output, Table 9. mA Stepping Pushbuttons, Pushbutton, Adjustment
View all Fluke 789 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 34 highlights
789 Users Manual Table 9. mA Stepping Pushbuttons Pushbutton X M % STEP % STEP G W F Span Check Span Check E Adjustment Adjusts up to the next higher 25 % step Adjusts down to the next lower 25 % step Sets to 100 % value Sets to 0 % value Step 0 % 25 % 50 % 75 % 100 % 125 % 120 % Table 10. mA Step Values Value (for each span setting) 4 to 20 mA 0 to 20 mA 4.000 mA 0.000 mA 8.000 mA 5.000 mA 12.000 mA 10.000 mA 16.000 mA 15.000 mA 20.000 mA 20.000 mA 24.000 mA 24.000 mA Auto Ramping the mA Output Auto ramping gives the ability to continuously apply a varying current stimulus from the meter to a transmitter, while hands remain free to test the response of the transmitter. Select either sourcing or simulating by choosing the SOURCE or SIMULATE jacks. 26