Fluke 8808A Fluke 8808A Digital Multimeter Datasheet - Page 3
Electrical Specifications - multimeter
View all Fluke 8808A manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 3 highlights
Electrical Specifications Accuracy specifications are valid for 5-½ digit mode and after at least a half-hour warm-up. DC Voltage Specifications Maximum Input 1000 V on any range. Common Mode Rejection 120 dB at 50 or 60 Hz ധ0.1㧑 (1 k˖ unbalance) Normal Mode Rejection 80 dB at Slow Rate A/D Nonlinearity 15 ppm of range Input Bias Current 30 pA at 25 rC Settling Considerations Measurement settling times are affected by source impedance, cable dielectric characteristics, and input signal changes Input Characteristics Range Full-Scale (5-1/2 Digits) Slow Resolution Medium Input Impedance Fast 200 mV 2 V 20 V 200 V 1000 V 199.999 mV 1.99999 V 19.9999 V 199.999 V 1000.00 V 1 µV 10 µV 100 µV 1 mV 10 mV 10 µV 100 µV 1000 µV 10 mV 100 mV 10 µV 100 µV 1000 µV 10 mV 100 mV >10 G [1] >10 G [1] 10 M ±1 % 10 M ±1 % 10 M ±1 % Notes: [1] At some dual display measurements, the input impedance of 200 mV and 2 V ranges may be changed to 10 M . Accuracy Accuracy [1] Range 90 days 23 °C ± 5 °C 200 mV 0.01 + 0.003 2 V 0.01 + 0.002 20 V 0.01 + 0.003 200 V 0.01 + 0.002 1000 V 0.01 + 0.002 Notes: [1] Accuracy given as ± (% of reading + % of range) 1 year 23 °C ± 5°C 0.015 + 0.004 0.015 + 0.003 0.015 + 0.004 0.015 + 0.003 0.015 + 0.003 Temperature Coefficient/°C Outside 18 - 28 °C 0.0015 + 0.0005 0.001 + 0.0005 0.0020 + 0.0005 0.0015 + 0.0005 0.0015 + 0.0005 2 Fluke Corporation 8808A Fluke 8808A Digital Multimeter Extended Specifications Fluke Corporation 3