Fluke BT520 Product Manual - Page 41

Measure AC Voltage, Measurement Range, Save AC Voltage Readings

Page 41 highlights

3 Use the Product Make Measurements Measure AC Voltage The Product supplies two independent readings to show the rms and frequency of ac voltage. To measure ac voltage, turn the rotary switch to . See Figure 3-4 for connections. Vac UPS Figure 3-4. Measure AC Voltage hsn021.eps Measurement Range The ac voltage measurement has only one range: 600 V. The frequency measurement uses auto ranging. This cannot be changed. Save AC Voltage Readings In Meter mode, push the Save softkey to save the current ac voltage reading and the timestamp. All saved data is stored in chorological order. In Sequence mode, push the Save softkey to save the current ac voltage reading and the timestamp. The current battery number increases 1 automatically. The cell that corresponds to the tested point becomes full, and the cursors move forward. Note Up to 20 ac voltage readings can be saved in one profile. 3-15 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62

Use the Product
Make Measurements
3
3-15
Measure AC Voltage
The Product supplies two independent readings to show the rms and frequency of ac
voltage.
To measure ac voltage, turn the rotary switch to
. See Figure 3-4 for connections.
Vac
UPS
hsn021.eps
Figure 3-4. Measure AC Voltage
Measurement Range
The ac voltage measurement has only one range: 600 V. The frequency measurement
uses auto ranging. This cannot be changed.
Save AC Voltage Readings
In Meter mode, push the
Save
softkey to save the current ac voltage reading and the
timestamp. All saved data is stored in chorological order.
In Sequence mode, push the
Save
softkey to save the current ac voltage reading and
the timestamp. The current battery number increases 1 automatically. The cell that
corresponds to the tested point becomes full, and the cursors move forward.
Note
Up to 20 ac voltage readings can be saved in one profile.
Shop for Fluke products online at:
1.888.610.7664
www.
MyFlukeStore
.com