Problem, Symptom, Check This, Remedy

Page 15 highlights

EN Problem Symptom Check This Error message Err shows on the display. A measure- ment error occurred Err 1, 2, or 3 shows on the display The wrist cuff is not secure Err 4 shows on the display The monitor detected motion while measuring Err 5 shows on the display Err 6 shows on the display Pressure is over 37.3 kPa (280 mmHg) Deflation period was too long Remedy Relax for a moment and then measure again Refasten the cuff and then measure again Movement can affect the measurement. Relax for a moment and then measure again Relax for a moment and then measure again Movement can affect the measurement. Relax for a moment and then measure again Problem Symptom Check This Remedy Error message Err 7 or 8 shows on the display A calibration error occurred Retake the measurement. If the problem persists, contact the retailer or our customer service department for further assistance. Refer to the warranty for contact information and return instructions Settings are wrong Date and time are incorrect Measurement unit (kPa or mmHg) are incorrect The clock Reset the clock was not set or reset after installing new batteries The Reset the measurement measurement unit was not unit set or reset after installing new batteries 14

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20

EN
14
Problem
Symptom
Check This
Remedy
Error
message
Err
shows
on the
display.
A measure-
ment error
occurred
Relax for a
moment and
then measure
again
Err 1, 2,
or 3
shows
on the
display
The wrist
cuff is not
secure
Refasten the
cuff and then
measure again
Err 4
shows
on the
display
The monitor
detected
motion while
measuring
Movement can
affect the
measurement.
Relax for a
moment and
then measure
again
Err 5
shows
on the
display
Pressure is
over 37.3
kPa (280
mmHg)
Relax for a
moment and
then measure
again
Err 6
shows
on the
display
Deflation
period was
too long
Movement can
affect the
measurement.
Relax for a
moment and
then measure
again
Problem
Symptom
Check This
Remedy
Error
message
Err 7 or 8
shows
on the
display
A calibration
error
occurred
Retake the
measurement. If
the problem
persists, contact
the retailer or our
customer service
department for
further assistance.
Refer to the
warranty for
contact
information and
return instructions
Settings
are
wrong
Date and
time are
incorrect
The clock
was not set or
reset after
installing new
batteries
Reset the clock
Measure-
ment unit
(kPa or
mmHg)
are
incorrect
The
measurement
unit was not
set or reset
after installing
new batteries
Reset the
measurement
unit