. Printer Calibration Spectrophotometer-Based Calibration There are two methods provided for entering the calibration data values measured by a spectrophotometer. The first method allows you to use PhaserMatch or PhaserCal software to control and communicate with a spectrophotometer that is connected to your computer. The second method allows you to manually import LAB measured data. The spectrophotometers that can be controlled by PhaserMatch or PhaserCal software are: ■ GretagMacbeth Spectrolino/SpectroScan ■ GretagMacbeth Eye-One System ■ X-Rite DTP-41 ■ X-Rite DTP-70 ■ X-Rite PULSE ■ X-Rite Digital Swatchbook ■ SpectroStar SpectroCam Note: PhaserMatch and PhaserCal software do not support any densitometers for performing a color calibration procedure. PhaserMatch and PhaserCal software walks you through a series of steps that generate and download the calibration data to a printer. To perform a spectrophotometer-based calibration: 1. Launch the PhaserMatch or PhaserCal application. 2. Click the Calibrate Printer button in the main screen. 3. Click the Spectrophotometer-Based Calibration button in the main Calibrate Printer screen 4. Select a device from the list of supported spectrophotometers or select Manually Enter Data. 5. Select a printer, print quality, and paper type you use to do the calibration procedure. PhaserMatch and PhaserCal 4.0 User Guide 2-1

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35

PhaserMatch and PhaserCal 4.0 User Guide
2-1
.
Spectrophotometer-Based Calibration
There are two methods provided for entering the calibration data values measured by a
spectrophotometer. The first method allows you to use PhaserMatch or PhaserCal software to
control and communicate with a spectrophotometer that is connected to your computer. The
second method allows you to manually import LAB measured data.
The spectrophotometers that can be controlled by PhaserMatch or PhaserCal software are:
GretagMacbeth Spectrolino/SpectroScan
GretagMacbeth Eye-One System
X-Rite DTP-41
X-Rite DTP-70
X-Rite PULSE
X-Rite Digital Swatchbook
SpectroStar SpectroCam
Note:
PhaserMatch and PhaserCal software do not support any densitometers for performing
a color calibration procedure.
PhaserMatch and PhaserCal software walks you through a series of steps that generate and
download the calibration data to a printer.
To perform a spectrophotometer-based calibration:
1.
Launch the PhaserMatch or PhaserCal application.
2.
Click the
Calibrate Printer
button in the main screen.
3.
Click the
Spectrophotometer-Based Calibration
button in the main
Calibrate Printer
screen
4.
Select a device from the list of supported spectrophotometers or select
Manually Enter
Data
.
5.
Select a printer, print quality, and paper type you use to do the calibration procedure.
Printer Calibration