Campbell Scientific HFP01 Open Path Eddy Covariance (OPEC) - Page 16
Sensor Configuration, CAUTION
View all Campbell Scientific HFP01 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 16 highlights
OPEC Open-Path Eddy-Covariance System North X Wind Vane Wind Vector Y FIGURE 3-2. Compass coordinate system, compass wind direction is 140 degrees 3.1.2 Sensor Configuration The IRGASON communicates with Campbell Scientific dataloggers using a digital communication protocol called Synchronous Device for Measurement (SDM). Each sensor connected to the SDM bus has a unique address. A maximum of 15 addresses are allowed by the protocol and each sensor must have a unique SDM address. CAUTION Do not use SDM address F (15) because it is reserved for use with the Group Trigger instruction. The IRGASON is shipped from the factory is set with a default SDM address of 1. This address is set in software and can be changed using the software DevConfig or ECMon and a USB cable. DevConfig is sensor/peripheral support software shipped with LoggerNet. ECMon is PC support software shipped with the IRGASON and is also available in the downloads section online at www.campbellsci.com. The program sets the IRGASON bandwidth to 20 Hz. The instrument delay is a function the bandwidth setting and is automatically computed by the datalogger program. To use a bandwidth other than 20 Hz, change the program constant "BANDWIDTH" to the appropriate value. For more information on IRGASON bandwidth and delay see the IRGASON manual. NOTE The settings are sent to the IRGASON each time it is powered on and only if connected to the datalogger via SDM. The settings are not saved onto the IRGASON's non-volatile memory. 8