Fluke 1552A-12 User Manual - Page 13

Operation, Immersion Depth and Usage

Page 13 highlights

Stik Thermometer Operation Operation Immersion Depth and Usage WCaution To prevent possible damage to the Product or to equipment under test, do not immerse the Probe beyond the maximum immersion depth. This may damage its internal electronics. Allow enough time for the Probe to stabilize before recording measurements. Use the Stability Display to determine when the Probe has stabilized. Probe sensor length is shown in Figure 1. 1 3 2 A Sheath Length: 1551A-9: 229 mm (9 in) 1551A-20: 508 mm (20 in) 1551A-12/1552A-12: 305 mm (12 in) B Sensor Length: 1551A: ≤10 mm (0.39 in) 1552A: ≤30 mm (1.18 in) C Sheath Diameter: 1551A-9: 4.8 mm (3/16 in) 1551A-12/1551A-20/1552A: 6.35 mm (1/4 in) Figure 1. Probe Measurements gjo004.eps 5

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22

Stik Thermometer
Operation
5
Operation
Immersion Depth and Usage
W
Caution
To prevent possible damage to the Product or to equipment
under test, do not immerse the Probe beyond the maximum
immersion depth. This may damage its internal electronics.
Allow enough time for the Probe to stabilize before recording measurements. Use the
Stability Display to determine when the Probe has stabilized.
Probe sensor length is shown in Figure 1.
3
2
1
gjo004.eps
A
Sheath Length:
1551A-9: 229 mm (9 in)
1551A-20: 508 mm (20 in)
1551A-12/1552A-12: 305 mm (12 in)
B
Sensor Length:
1551A:
10 mm (0.39 in)
1552A:
30 mm (1.18 in)
C
Sheath Diameter:
1551A-9: 4.8 mm (3/16 in)
1551A-12/1551A-20/1552A: 6.35 mm (1/4 in)
Figure 1. Probe Measurements