Resistance measurement (R~) Measuring method Measuring voltage Short circuit current Measuring frequency Current and voltage measurement 20 V ac, square pulse > 250 mA ac 94, 105, 111, 128 Hz selected manually or automatically (AFC) Measuring range 0.020 7 to 300 k7 Display range 0.001 7 to 2.999 7 3.0 7 to 29.99 7 30 7 to 299.9 7 300 7 to 2999 7 3.0 k7 to 29.99 k7 30.0 k7 to 299.9 k7 Resolution 0.001 7 0.01 7 0.1 7 17 10 7 100 7 Intrinsic error ± (2 % of rdg + 2 digit) Measuring time Max. interference voltage Max overload typ. 6 sec. 24 V, with higher voltages measurement will not be started Urms max. = 250 V Operating errors ± (5 % of rdg + 5 digit) Resistance measurement (R ) Measuring method Open circuit voltage Short circuit current Formation of measured value current- voltage measurement as per IEC61557-4 possible 20 V dc 200 mA dc with 4-pole measurement wires on H, S, ES can be extended without additional error. Resistances > 1 7 in wire E can cause additional error of 5m 7/7. Measuring range 0.020 7 to 3 k7 Display range 0.001 7 to 2.999 7 3.0 7 to 29.99 7 30.0 7 to 299.9 7 300 7 to 2999 7 Resolution 0.001 7 0.01 7 0.1 7 17 Intrinsic error ± (2 % of rdg + 2 digit) Measuring sequence Measuring time Max. interference voltage Max inductivity Max. overload Approx. 2 measurements/s Typ. 4 sec. incl. reversal of polarity (2-pole or 4-pole) b 3 V ac or dc, with higher voltages measurement will not be started 2 Henry Urms = 250 V Operating error ± (5 % of rdg + 5 digit) Compensation of lead resistance (RK) Compensation of lead resistance (RK) can be switched on in functions RE 3-pole, RE 4-pole Formation of measured value R = R - R * display measured compensated , R~ , and R * Value of setpoint entry RK = 0.000 7, variable from 0.000 to 29.99 7 by means of measuring adjustment. 2-pole 9 Fluke Corporation 1623 and 1625 GEO Earth Ground Testers

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10

9
Fluke Corporation
1623 and 1625 GEO Earth Ground Testers
Resistance measurement (R~)
Measuring method
Current and voltage measurement
Measuring voltage
20 V ac, square pulse
Short circuit current
> 250 mA ac
Measuring frequency
94, 105, 111, 128Hz selected manually or automatically (AFC)
Measuring range
Display range
Resolution
Intrinsic error
Operating errors
0.020
to 300 k
0.001
to 2.999
0.001
± (2 % of rdg + 2 digit)
± (5 % of rdg + 5 digit)
3.0
to 29.99
0.01
30
to 299.9
0.1
300
to 2999
1
3.0 k
to 29.99 k
10
30.0 k
to 299.9 k
100
Measuring time
typ. 6 sec.
Max. interference voltage
24 V, with higher voltages measurement will not be started
Max overload
U
rms
max. = 250 V
Resistance measurement (R
)
Measuring method
current- voltage measurement as per IEC61557-4 possible
Open circuit voltage
20 V dc
Short circuit current
200 mA dc
Formation of measured value
with 4-pole measurement wires on H, S, ES can be extended without additional error.
Resistances > 1
in wire E can cause additional error of 5m
/
.
Measuring range
Display range
Resolution
Intrinsic error
Operating error
0.020
to 3 k
0.001
to 2.999
0.001
± (2 % of rdg + 2 digit)
± (5 % of rdg + 5 digit)
3.0
to 29.99
0.01
30.0
to 299.9
0.1
300
to 2999
1
Measuring sequence
Approx. 2 measurements/s
Measuring time
Typ. 4 sec. incl. reversal of polarity (2-pole or 4-pole)
Max. interference voltage
3 V ac or dc, with higher voltages measurement will not be started
Max inductivity
2 Henry
Max. overload
U
rms
= 250 V
Compensation of lead resistance (R
K
)
Compensation of lead resistance (R
K
) can be switched on in functions R
E
3-pole, R
E
4-pole
, R~ , and R
2-pole
Formation of measured value
R
display
= R
measured
- R
compensated
*
* Value of setpoint entry R
K
= 0.000
, variable from 0.000 to 29.99
by means of measuring adjustment.