Fluke 1738 Calibration Manual - Page 28
Basic Instrument Setup for all Verifications, no voltage transformers used
View all Fluke 1738 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 28 highlights
1736/1738 Calibration Manual Table 9. Calibration and Verification in Excel Worksheet (cont.) Item Description Sensor selector - select items from the list for a verification of the accessory. Use DIRECT for the Fluke Logger verification. Calibration items - Select Voltage, AUX Input or Current input for calibration. For a Logger calibration all three items need to be calibrated sequentially. A specific order is not required. Calibrator Control setup - When the calibrator is connected to the PC using a RS232 cable select Automatic to control the calibrator. Use the drop-down list box to configure the COM port. Otherwise select Manual. Voltage Divider setup - Configure the resistor values, R1 and R2, of the voltage divider for current verification. Store the Excel workbook to keep the applied values for future use. Supported Calibrators: Fluke 5520A and 5522A Calibrator settings: Baud rate: 9600 Data bits: 8 Stop bit:1 Parity:None Stall: XON/XOFF EOL:CR/LF Basic Instrument Setup for all Verifications The Fluke173x-ExcelTool_Vxxx (ExcelTool) has built-in procedures to verify and adjust the Logger. The Verification uses an external divider. This divider, (see Verification Box Assembly) provides more accurate voltages than a direct connection to the 5520A. The 5520A uses a divider with a 50 Ω output impedance when sourcing