Fuse Test, Fuse Test

Page 18 highlights

3000 FC Calibration Manual Fuse Test To verify that the Product fuse is good: 1. Set the function switch to . 2. Connect a test lead to the  jack as shown in Figure 1. 3. Touch the other end of the test lead to the mA jack. A good fuse will show a resistance of 12 Ω or less. Replace the fuse if the resistance is higher or OL is shown. See the "User-Replaceable Parts" section for the correct replacement fuse. 440 mA

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30

3000 FC
Calibration Manual
10
Fuse Test
To verify that the Product fuse is good:
1. Set the function switch to
.
2.
Connect a test lead to the
jack as shown in Figure 1.
3.
Touch the other end of the test lead to the mA jack.
A good fuse will show a resistance of 12
Ω
or less. Replace the fuse if the
resistance is higher or
OL
is shown. See the “User-Replaceable Parts” section
for the correct replacement fuse.
440 mA
<12
OK
OK
gxr009.eps
Figure 1. Fuse Test