Page 18 highlights
3000 FC Calibration Manual Fuse Test To verify that the Product fuse is good: 1. Set the function switch to . 2. Connect a test lead to the jack as shown in Figure 1. 3. Touch the other end of the test lead to the mA jack. A good fuse will show a resistance of 12 Ω or less. Replace the fuse if the resistance is higher or OL is shown. See the "User-Replaceable Parts" section for the correct replacement fuse. 440 mA
3000 FC
Calibration Manual
10
Fuse Test
To verify that the Product fuse is good:
1. Set the function switch to
.
2.
Connect a test lead to the
jack as shown in Figure 1.
3.
Touch the other end of the test lead to the mA jack.
A good fuse will show a resistance of 12
Ω
or less. Replace the fuse if the
resistance is higher or
OL
is shown. See the “User-Replaceable Parts” section
for the correct replacement fuse.
440 mA
<12
OK
OK
gxr009.eps
Figure 1. Fuse Test