Fluke 374 Fluke 374, 375, 376 Users Manual - Page 24

Range, Resolution, and 375, Accuracy, digits 5 - 500 Hz, Crest Factor 50/60Hz

Page 24 highlights

374, 375, 376 Users Manual AC Current via Flexible Current Probe Range 2500 A Resolution 374 and 375 0.1 A (≤ 600 A) 1 A (≤ 2500 A) 376 0.1 A (≤ 999.9 A) 1 A (≤ 2500 A) Accuracy 3 % ±5 digits (5 - 500 Hz) Crest Factor (50/60Hz 3.0 at 1100 A (375 and 376 only) 2.5 at 1400 A 1.42 at 2500 A Add 2 % for C.F. > 2 20

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32

374, 375, 376
Users Manual
20
AC Current via Flexible Current Probe
Range
....................................................
2500 A
Resolution
374 and 375
........................................
0.1 A (
600 A)
1 A (
2500 A)
376
......................................................
0.1 A (
999.9 A)
1 A (
2500 A)
Accuracy
.................................................
3 %
±
5 digits (5 – 500 Hz)
Crest Factor (50/60Hz)
...........................
3.0 at 1100 A (375 and 376 only)
2.5 at 1400 A
1.42 at 2500 A
Add 2 % for C.F. > 2