Infrared Thermometer Specifications Specifications Temperature Range Accuracy (Calibration geometry with ambient temperature 23 °C ±2 °C) Response Time (95 %) Spectral Response Emissivity Temperature Coefficient 62 MAX 62 MAX + -30 °C to 500 °C (-22 °F to 932 °F) -30 °C to 650 °C (-22 °F to 1202 °F) ≥0 °C: ±1.5 °C or ±1.5 % of reading, whichever is greater (≥32 °F: ±3 °F or ±1.5 % of reading, whichever is greater) ≥ -10 °C to

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22

Infrared Thermometer
Specifications
7
Specifications
62 MAX
62 MAX +
Temperature Range
-30
°
C to 500
°
C
(-22
°
F to 932
°
F)
-30
°
C to 650
°
C
(-22
°
F to 1202
°
F)
Accuracy
(Calibration geometry with
ambient temperature 23
°
C
±
2
°
C)
0
°
C:
±
1.5
°
C or
±
1.5 % of
reading, whichever is
greater
(
32
°
F:
±
3
°
F or
±
1.5 % of
reading, whichever is
greater)
-10
°
C to <0
°
C:
±
2
°
C
(
14
°
F to <32
°
F:
±
4
°
F)
< -10
°
C:
±
3
°
C
(<14
°
F:
±
6
°
F)
0
°
C:
±
1
°
C or
±
1 % of
reading, whichever is
greater (
32
°
F:
±
2
°
F or
±
1 % of reading, whichever
is greater)
-10
°
C to <0
°
C:
±
2
°
C
(
14
°
F to <32
°
F:
±
4
°
F)
< -10
°
C:
±
3
°
C
(<14
°
F:
±
6
°
F)
Response Time (95 %)
<500 ms (95 % of reading)
<300 ms (95 % of reading)
Spectral Response
8 to 14 microns
Emissivity
0.10 to 1.00
Temperature Coefficient
±
0.1
°
C/
°
C or
±
0.1 %/
°
C of reading (whichever is greater)