Fluke 721EX User Manual - Page 25

to scroll through the port choices P1 or P2., to open the %ERROR screen.

Page 25 highlights

Example: A pressure transmitter under test is 30 psi (2 Bar) FullScale and outputs a corresponding 4 mA to 20 mA signal. Program a 0 psi to 30 psi pressure span into the Product and the Product calculates and shows the deviation or %Error from the expected 4 mA to 20 mA output. This eliminates the need for manual calculations and helps when it is difficult to set an exact pressure. To use the %ERROR function: 1. Push  to access the menus. 2. Push  to step through the menu until %ERROR is shown on the menu bar. 3. Push  to open the %ERROR screen. 4. Push  to scroll through the port choices (P1 or P2). 5. Push  to configure the pressure range settings. 6. Use the arrow keys to set the 100 % point of the desired pressure range, select DONE SET when finished. 7. Use the arrow keys to set 0 % point and select DONE SET. The % ERROR mode is now ready to use. Pressure Calibrator Product Features Note The 0 % and 100 % point will be saved in nonvolatile memory until they are changed again by the user. hmq054.eps 17

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46

Pressure Calibrator
Product Features
17
Example:
A pressure transmitter under test is 30 psi (2 Bar) Full-
Scale and outputs a corresponding 4 mA to 20 mA signal.
Program a 0 psi to 30 psi pressure span into the Product
and the Product calculates and shows the deviation or
%Error from the expected 4 mA to 20 mA output. This
eliminates the need for manual calculations and helps
when it is difficult to set an exact pressure.
To use the %ERROR function:
1.
Push
to access the menus.
2.
Push
to step through the menu until
%ERROR
is
shown on the menu bar.
3.
Push
to open the %ERROR screen.
4.
Push
to scroll through the port choices (P1 or P2).
5.
Push
to configure the pressure range settings.
6.
Use the arrow keys to set the 100 % point of the
desired pressure range, select DONE SET when
finished.
7.
Use the arrow keys to set 0 % point and select
DONE SET. The % ERROR mode is now ready to
use.
Note
The 0 % and 100 % point will be saved in non-
volatile memory until they are changed again by
the user.
hmq054.eps