Fluke 726 Fluke 726 Process Calibrator Datasheet - Page 2

Specifications - manual

Page 2 highlights

726 Specifications Function Measure or Source Voltage mA mV (TC terminals) Resistance Frequency Loop Supply Thermocouples Thermocouples RTDs Range or Type 0 to 100 mV 0 to 20 V (source) 0 to 30 V (measure) 0 to 24 -10.00 mV to +75.00 mV 5 Ω to 4000 Ω 2.0 to 1,000 CPM 1 to 1100 Hz 1.0 to 10.0 kHz 10.0 to 15.0 kHz 24 V dc J, K, T, E, L, N, U, C, BP, XK B, R, S Cu 10 Ni 120 (672) Pt 100, 200, 500, 1000 (385) Pt 100 (3916), Pt 100 (3926) Cu 10 Resolution 0.01 mV 0.01 V 0.01 V 0.001 mA .01 mV ≤ 0.01 Ω to 0.1 Ω 0.1 CPM 1 Hz 0.1 kHz 0.1 kHz N/A 0.1 ºC, 0.1 ºF 1 ºC, 1 ºF 0.1 °C, 0.1 °F 0.01 ºC, 0.01 °F Accuracy 0.01 % Rdg + 2 LSD Notes Max load, 1 mA 0.01 % Rdg + 2 LSD Max load, 1000 Ω 0.01 % of range + 1 LSD 0.015 % ± 0.05 % ± 0.05 % ± 0.25 % ± 0.5 % For frequency source; 1 V to 20 V p-p squarewave, -0.1 V offset 10 % to 0.2 ºC to 1.2 ºC to 1.8 °C to 0.15 ºC Simultaneous Function Capability 24.000 mA DC 24.000 mA DC with 24 V loop supply 100.00 mV DC 30.000 V DC measure 20.000 V DC measure, 20.000 VDC Source 5 to 4000 Ohms Thermocouple J, K, T, E, R, S, B, M, L, U, N, C, BP, XK RTD Ni120; Pt100 (392); Pt100 (JIS); Pt100, 200, 500, 1000 (385), Cu 10 Pressure (requires Fluke 700PXX Modules) Frequency; Squarewave, 1 CPM to10 kHz; fixed amplitude 5 V p-p M = Measure, S = Source/Simulate Channel A M M M M Channel B M or S M or S M or S M or S M or S M or S M used as S M or S General Specifications Storage temperature: -20 °C to 71 °C Operating temperature: -10 °C to 55 °C Relative humidity: 90 % (10 °C to 30 °C); 75 % (30 °C to 40 °C); 45 % (40 °C to 50 °C); 35 % (50 °C to 55 °C) Shock: 1 meter drop test Safety: CSA C22.2 No. 1010.1:1992 EMC: EN50082-1:1992 and EN55022:1994 Class B Size (HxWxD): 200 mm x 96 mm x 47 mm (7.9 in x 3.8 in x 1.9 in) Weight: 650 g (23 oz) Battery: Four AA alkaline batteries Battery life: 25 hours typical Warranty: Three-years Ordering Information Fluke-726 Precision Multifunction Process Calibrator Included TL75 Test Leads, AC72 Test Clips, one pair of stackable test leads, product overview manual (print) and user's manual (CD-ROM) in14 languages. 2 Fluke Corporation 726 Precision Multifunction Process Calibrator Fluke. Keeping your world up and running. Fluke Corporation PO Box 9090, Everett, WA USA 98206 Fluke Europe B.V. PO Box 1186, 5602 BD Eindhoven, The Netherlands For more information call: In the U.S.A. (800) 443-5853 or Fax (425) 446-5116 In Europe/M-East/Africa (31 40) 2 675 200 or Fax (31 40) 2 675 222 In Canada (800)-36-FLUKE or Fax (905) 890-6866 From other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116 Web access: http://www.fluke.com ©2005 Fluke Corporation. All rights reserved. Printed in U.S.A. 12/2005 2527590 D-EN-N Rev C

  • 1
  • 2

726 Specifications
2
Fluke Corporation
726 Precision Multifunction Process Calibrator
Fluke Corporation
PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa (31 40) 2 675 200 or
Fax (31 40) 2 675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
©2005 Fluke Corporation. All rights reserved.
Printed in U.S.A.
12/2005
2527590 D-EN-N Rev C
Fluke.
Keeping your world
up and running.
General Specifications
Storage temperature:
-20 °C to 71 °C
Operating temperature:
-10 °C to 55 °C
Relative humidity:
90 % (10 °C to 30 °C);
75 % (30 °C to 40 °C); 45 % (40 °C to 50
°C); 35 % (50 °C to 55 °C)
Shock:
1 meter drop test
Safety:
CSA C22.2 No. 1010.1:1992
EMC:
EN50082-1:1992 and
EN55022:1994 Class B
Size (HxWxD):
200 mm x 96 mm x 47 mm
(7.9 in x 3.8 in x 1.9 in)
Weight:
650 g (23 oz)
Battery:
Four AA alkaline batteries
Battery life:
25 hours typical
Warranty:
Three-years
Ordering Information
Fluke-726
Precision Multifunction
Process Calibrator
Included
TL75 Test Leads, AC72 Test Clips, one pair
of stackable test leads, product overview
manual (print) and user’s manual (CD-ROM)
in14 languages.
Simultaneous Function Capability
Channel A
Channel B
24.000 mA DC
M
M or S
24.000 mA DC with 24 V loop supply
M
100.00 mV DC
M or S
30.000 V DC measure
M
20.000 V DC measure, 20.000 VDC Source
M or S
5 to 4000 Ohms
M or S
Thermocouple J, K, T, E, R, S, B, M, L, U, N, C, BP, XK
M or S
RTD Ni120; Pt100 (392); Pt100 (JIS); Pt100, 200,
M or S
500, 1000 (385), Cu 10
Pressure (requires Fluke 700PXX Modules)
M
M used as S
Frequency; Squarewave, 1 CPM to10 kHz;
M or S
fixed amplitude 5 V p-p
M
= Measure,
S
= Source/Simulate
Function
Measure or Source
Range or Type
Resolution
Accuracy
Notes
Voltage
0 to 100 mV
0.01 mV
0.01 % Rdg + 2 LSD
Max load, 1 mA
0 to 20 V (source)
0.01 V
0 to 30 V (measure)
0.01 V
mA
0 to 24
0.001 mA
0.01 % Rdg + 2 LSD
Max load, 1000
Ω
mV (TC terminals)
-10.00 mV to +75.00 mV
.01 mV
0.01 % of range + 1 LSD
Resistance
5
Ω
to 4000
Ω
0.01
Ω
to 0.1
Ω
0.015 %
Frequency
2.0 to 1,000 CPM
0.1 CPM
± 0.05 %
For frequency source;
1 to 1100 Hz
1 Hz
± 0.05 %
1 V to 20 V p-p
1.0 to 10.0 kHz
0.1 kHz
± 0.25 %
squarewave, -0.1 V offset
10.0 to 15.0 kHz
0.1 kHz
± 0.5 %
Loop Supply
24 V dc
N/A
10 %
Thermocouples
J, K, T, E, L, N, U, C, BP, XK
0.1 ºC, 0.1 ºF
to 0.2 ºC
Thermocouples
B, R, S
1 ºC, 1 ºF
to 1.2 ºC
RTDs
Cu 10 Ni 120 (672)
0.1 °C, 0.1 °F
to 1.8 °C
Pt 100, 200, 500,
1000 (385)
Pt 100 (3916),
Pt 100 (3926) Cu 10
0.01 ºC, 0.01 °F
to 0.15 ºC