Page 118 highlights
741B/743B Users Manual DC Current Output Range/Mode Resolution Models 741B and 743B % of Output + % of Full Scale 1 Year 2 Year 22 mA/ Source mA 1 µA 0.01% + 0.015% 0.02% + 0.015% 22 mA/ Simulate Transmitter (Current Sink) 1 µA 0.02% + 0.03% 0.02% + 0.03% Maximum Burden Voltage: 24V Temperature Coefficient: (0.003% of output + 0.003% of f.s.)/°C in the ranges -10 to 18°C and 28 to 50°C Common Mode Error: 0.008% f.s/(Common Mode Volt) Maximum Input Voltage: 30V dc Specification applies for currents between 2 mA and 22 mA. For current below 2 mA, typical accuracy is 0.15% of full scale. 108
741B/743B
Users Manual
108
DC Current Output
Range/Mode
Resolution
Models 741B and 743B
% of Output + % of Full Scale
1 Year
2 Year
22 mA/ Source mA
1
μ
A
0.01% + 0.015%
0.02% + 0.015%
22 mA/ Simulate Transmitter
(Current Sink)
1
μ
A
0.02% + 0.03%
0.02% + 0.03%
Maximum Burden Voltage:
24V
Temperature Coefficient:
(0.003% of output + 0.003% of f.s.)/
°
C in the ranges -10 to 18
°
C and 28 to 50
°
C
Common Mode Error
: 0.008% f.s/(Common Mode Volt)
Maximum Input Voltage:
30V dc
Specification applies for currents between 2 mA and 22 mA. For current below 2 mA, typical accuracy is 0.15% of full
scale.