General specifications Data log function Measure functions: Voltage, current, resistance, frequency, temperature, pressure Reading rate: 1, 2, 5, 10, 20, 30, or 60 readings per minute Maximum record length: 8000 readings (7980 for 30 or 60 readings per minute) Ramp function Source functions: Voltage, current, resistance, frequency, temperature Rate: 4 steps/second Trip detect: Continuity* or voltage *Continuity detection not available when sourcing current Loop power function Voltage: 26 V Accuracy: 10 %, 18 V minimum at 22 mA Maximum current: 25 mA, short-circuit protected Maximum input voltage: 50 V dc Note: 250 Ω series resistance is automatically supplied whenever loop power is enabled on 754. HART modem interface (754 only) Maximum input voltage: 30 V dc Environmental specifications All calibrator specifications apply from +18 °C to +28 °C unless stated otherwise. Operating temperature: -10 °C to 50 °C Storage temperature: -20 °C to 60 °C Operating altitude: 3000 m above mean sea level (9842 ft) 90-day specifications: The standard specification intervals for the 750 Series are 1 and 2 years. Typical 90-day measurement and source accuracy can be estimated by dividing the one year "% of reading" or "% of output" specifications by 2. Floor specifications, expressed as "% of f.s." or "counts" or "ohms" remain constant. Ingress protection: IP-52 Power: Internal battery pack li-ion, 7.2 V, 4400 mAh, 30 Wh Battery life: Typical usage, >8 hours Dimensions: 136mm x 245mm x 63 mm (5.4 in x 9.6 in x 2.5 in) Weight: 1.2 kg (2.7 lb) Side port connections: • Pressure module connector • USB connector to interface to your PC • Digital instrument (HART) connector • Connection for optional battery charger/eliminator Safety: Complies with CAN/CSA C22.2 No 1010.1-92, ANSI/ISA S82.01-1994, UL3111, and EN610-1:1993. Data storage capacity: 1 week of calibration procedures and results Ordering information FLUKE-753 Documenting Process Calibrator FLUKE-754 Documenting Process Calibrator-HART Standard accessories include: Three sets of stackable test leads, three sets of TP220 test probes with three sets of "extended tooth" alligator clips, two sets AC280 hook clips, BP7240 Li-ion battery pack, BC7240 battery charger, C799 field soft case, USB communication cable, getting started guide, instruction manual on CDROM, NIST traceable certificate of calibration, DPC/TRACK 2 sample software that enables upload and printing of calibration records. Model Fluke-754 includes HART communication cable. Includes C799 Field Soft Case. Includes three year warranty. FLUKE-750SW DPC/TRACK 2 Software Included with DPC/TRACK software: Software media, Instruction Manual, USB Cable. FLUKE-700 Pxx Pressure Modules Included with each Fluke Pressure Module: BP-ISO Adapter(s) (except with P29 - P31), Instruction Sheet, NIST traceable calibration report and data, one-year warranty. Accessories Fluke-700PMP Fluke-700LTP-1 Fluke-700PTP-1 Fluke-700HTP-1 Fluke-700HTH-1 Fluke-700PRV-1 Fluke-700-IV Fluke-700PCK Fluke-700BCW Fluke-700TC1 Fluke-700TC2 Fluke-700TLK 754HCC BC7240 BP7240 C700 C781 C799 Pressure Pump; 100 psi/7 bar Low Pressure Test Pump Pneumatic Test Pump; 400 psi/40 bar Hydraulic Test Pump; 10,000 psi/700 bar Hydraulic Test Hose Pressure Relief Valve Kit for HTP Current Shunt (for mA/mA applications) Pressure Calibration Kit Bar Code Wand TC Mini-Plug Kit, 9 types TC Mini-Plug Kit, JKTERS Process Test lead kit Smart instrument communication cable Battery Charger Li-on Battery Pack Hard Carrying Case Soft Carrying Case Soft Field Case 11 Fluke Corporation

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11

General specifications
FLUKE-753 Documenting Process Calibrator
FLUKE-754 Documenting Process Calibrator-HART
Standard accessories include: Three sets of stackable
test leads, three sets of TP220 test probes with three
sets of “extended tooth” alligator clips, two sets AC280
hook clips, BP7240 Li-ion battery pack, BC7240 battery
charger, C799 field soft case, USB communication cable,
getting started guide, instruction manual on CDROM,
NIST traceable certificate of calibration, DPC/TRACK 2
sample software that enables upload and printing of
calibration records. Model Fluke-754 includes HART
communication cable. Includes C799 Field Soft Case.
Includes three year warranty.
FLUKE-750SW DPC/TRACK 2 Software
Included with DPC/TRACK software:
Software media, Instruction Manual, USB Cable.
FLUKE-700 Pxx Pressure Modules
Included with each Fluke Pressure Module:
BP-ISO Adapter(s) (except with P29 - P31), Instruction
Sheet, NIST traceable calibration report and data,
one-year warranty.
Accessories
Fluke-700PMP
Pressure Pump; 100 psi/7 bar
Fluke-700LTP-1
Low Pressure Test Pump
Fluke-700PTP-1
Pneumatic Test Pump;
400 psi/40 bar
Fluke-700HTP-1
Hydraulic Test Pump;
10,000 psi/700 bar
Fluke-700HTH-1
Hydraulic Test Hose
Fluke-700PRV-1
Pressure Relief Valve Kit for HTP
Fluke-700-IV
Current Shunt (for mA/mA
applications)
Fluke-700PCK
Pressure Calibration Kit
Fluke-700BCW
Bar Code Wand
Fluke-700TC1
TC Mini-Plug Kit, 9 types
Fluke-700TC2
TC Mini-Plug Kit, JKTERS
Fluke-700TLK
Process Test lead kit
754HCC
Smart instrument
communication cable
BC7240
Battery Charger
BP7240
Li-on Battery Pack
C700
Hard Carrying Case
C781
Soft Carrying Case
C799
Soft Field Case
Ordering information
Data log function
Measure functions:
Voltage, current, resistance,
frequency, temperature, pressure
Reading rate:
1, 2, 5, 10, 20, 30, or 60 readings
per minute
Maximum record length:
8000 readings
(7980 for 30 or 60 readings per minute)
Ramp function
Source functions:
Voltage, current, resistance,
frequency, temperature
Rate:
4 steps/second
Trip detect:
Continuity* or voltage
*Continuity detection not available when sourcing current
Loop power function
Voltage:
26 V
Accuracy:
10 %, 18 V minimum at 22 mA
Maximum current:
25 mA, short-circuit protected
Maximum input voltage:
50 V dc
Note: 250
Ω
series resistance is automatically supplied whenever loop power is
enabled on 754.
HART modem interface (754 only)
Maximum input voltage: 30 V dc
Environmental specifications
All calibrator specifications apply from +18 °C to
+28 °C unless stated otherwise.
Operating temperature:
-10 °C to 50 °C
Storage temperature:
-20 °C to 60 °C
Operating altitude:
3000 m above mean sea level
(9842 ft)
90-day specifications:
The standard specification
intervals for the 750 Series are 1 and 2 years.
Typical 90-day measurement and source accuracy can
be estimated by dividing the one year “% of reading”
or “% of output” specifications by 2. Floor specifications,
expressed as “% of f.s.” or “counts” or “ohms” remain
constant.
Ingress protection:
IP-52
Power:
Internal battery pack li-ion, 7.2 V,
4400 mAh, 30 Wh
Battery life:
Typical usage, >8 hours
Dimensions:
136mm x 245mm x 63 mm
(5.4 in x 9.6 in x 2.5 in)
Weight:
1.2 kg (2.7 lb)
Side port connections:
Pressure module connector
USB connector to interface to your PC
Digital instrument (HART) connector
Connection for optional battery charger/eliminator
Safety:
Complies with CAN/CSA C22.2 No 1010.1-92,
ANSI/ISA
S82.01-1994, UL3111, and EN610-1:1993.
Data storage capacity:
1 week of calibration procedures and results
11 Fluke Corporation