Fluke 9142-A-156 Product Manual - Page 20

Base Unit Specifications, P Specifications

Page 20 highlights

914X Field Metrology Wells Specifications Base Unit Specifications 9142 9143 9144 Weight Power Requirements System Fuse Ratings 4-20 mA Fuse (-P model only) Computer Interface Safety 8.16 kg (18 lbs) 100 V to 115 V (± 10 %) 50/60 Hz, 635 W 230 V (± 10 %) 50/60 Hz, 575 W 115 V: 6.3 A T 250 V 230 V: 3.15 A T 250 V 7.3 kg (16 lbs) 7.7 kg (17 lbs) 100 V to 115 V (± 10 %), 50/60 Hz, 1400 W 230 V (± 10%), 50/60 Hz, 1800 W 115 V: 15 A F 250 V 230 V: 10 A F 250 V 50 mA F 250V RS-232 and 9930 Interface-it control software included EN 61010-1:2001, CAN/CSA C22.2 No. 61010.1-04 Table 3 -P Option Specifications -P Specifications Built-in Reference Thermometer Readout Accuracy (4-Wire Reference Probe)† Reference Resistance Range Reference Resistance Accuracy‡ Reference Characterizations Reference Measurement Capability Reference Probe Connection Built-in RTD Thermometer Readout Accuracy RTD Resistance Range Resistance Accuracy‡ RTD Characterizations RTD Measurement Capability RTD Connection ± 0.013 °C at -25 °C ± 0.015 °C at 0 °C ± 0.020 °C at 50 °C ± 0.025 °C at 150 °C ± 0.030 °C at 200 °C ± 0.040 °C at 350 °C ± 0.050 °C at 420 °C ± 0.070 °C at 660 °C 0 ohms to 400 ohms 0 ohms to 42 ohms: ±0.0025 ohms 42 ohms to 400 ohms: ±60 ppm of reading ITS-90, CVD, IEC-751, Resistance 4-wire 6 Pin Din with Infocon Technology NI-120: ± 0.015 °C at 0 °C PT-100 (385): ± 0.02 °C at 0 °C PT-100 (3926): ± 0.02 °C at 0 °C PT-100 (JIS): ± 0.02 °C at 0 °C 0 ohms to 400 ohms 0 ohms to 25 ohms: ±0.002 ohms 25 ohms to 400 ohms: ±80 ppm of reading PT-100 (385),(JIS),(3926), NI-120, Resistance 2-,3-,4-wire RTD w\ Jumpers only 4 terminal input 14 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42

914X Field Metrology Wells
Specifications
14
Base Unit Specifications
9142
9143
9144
Weight
8.16 kg (18 lbs)
7.3 kg (16 lbs)
7.7 kg (17 lbs)
Power Requirements
100 V to 115 V (± 10 %)
50/60 Hz, 635 W
230 V (± 10 %) 50/60 Hz,
575 W
100 V to 115 V (± 10 %), 50/60 Hz, 1400 W
230 V (± 10%), 50/60 Hz, 1800 W
System Fuse Ratings
115 V: 6.3 A T 250 V
230 V: 3.15 A T 250 V
115 V: 15 A F 250 V
230 V: 10 A F 250 V
4–20 mA Fuse (-P
model only)
50 mA F 250V
Computer Interface
RS-232 and 9930 Interface-it control software included
Safety
EN 61010-1:2001, CAN/CSA C22.2 No. 61010.1-04
-P Option Specifications
Table 3
-P Specifications
Built-in Reference Thermometer Readout
Accuracy
(4-Wire Reference Probe)
± 0.013 °C at -25 °C
± 0.015 °C at 0 °C
± 0.020 °C at 50 °C
± 0.025 °C at 150 °C
± 0.030 °C at 200 °C
± 0.040 °C at 350 °C
± 0.050 °C at 420 °C
± 0.070 °C at 660 °C
Reference Resistance Range
0 ohms to 400 ohms
Reference Resistance Accuracy‡
0 ohms to 42 ohms: ±0.0025 ohms
42 ohms to 400 ohms: ±60 ppm of reading
Reference Characterizations
ITS-90, CVD, IEC-751, Resistance
Reference Measurement Capability
4-wire
Reference Probe Connection
6 Pin Din with Infocon Technology
Built-in RTD Thermometer Readout Accuracy
NI-120: ± 0.015 °C at 0 °C
PT-100 (385): ± 0.02 °C at 0 °C
PT-100 (3926): ± 0.02 °C at 0 °C
PT-100 (JIS): ± 0.02 °C at 0 °C
RTD Resistance Range
0 ohms to 400 ohms
Resistance Accuracy
0 ohms to 25 ohms: ±0.002 ohms
25 ohms to 400 ohms: ±80 ppm of reading
RTD Characterizations
PT-100 (385),(JIS),(3926), NI-120, Resistance
RTD Measurement Capability
2-,3-,4-wire RTD w\ Jumpers only
RTD Connection
4 terminal input
Shop for Fluke products online at:
1.888.610.7664
www.
MyFlukeStore
.com