Fluke CO-220 Fluke Air Quality Measurement - Making the Number Add Up Understa - Page 4

Humidity, Airborne particles, Carbon dioxide - meter

Page 4 highlights

In terms of response time, both technologies perform well. But response time is also affected by the overall design of the test instrument. If the temperature sensor is enclosed in a massive housing, the surrounding material is liable to affect the speed with which air heats or cools the sensor, thus slowing response time. Minimal response time requires minimizing mass, and that means shrinking components. If carried too far, this could affect instrument durability. Stability and drift. The thermistor temperature sensor is very stable and does not drift like some other sensors. However, the temperature sensing function can drift over time due to other components used and the layout of the circuit design. In design and manufacture, care must be taken in the layout of the printed circuit board that connects with the thermistor. Ionic contamination from soldering flux residues can cause performance degradation. This increases as the impedance of the circuit increases. Good analog design is essential to achieving optimal temperature performance. User guidance. Allow for instrument 'settling time' when measuring temperature. The mass of a temperature tester may slow response time in rapidly changing conditions. In simple terms, it takes a while for the instrument to reach ambient temperature. In addition, the instrument's electronic circuitry will perform differently as temperatures vary, so the circuitry too must settle and stabilize. Humidity Sensor technologies. The device traditionally used to measure humidity is the sling psychrometer. This instrument consists of two bulb thermometers, one of which is surrounded by a wet cloth. When the device is slung through the air, moisture evaporates from the wet bulb and cools the thermometer. The dryer the air, the greater the temperature difference between the wet and dry thermometers. A more convenient alternative is a hand-held humidity meter that uses a capacitance sensor technology. A semi-permeable membrane in the sensor becomes more conductive as humidity increases and moisture penetrates the membrane. The meter interprets this change in terms of humidity level. Technology characteristics. The sling psychrometer is highly accurate when used correctly. It is also relatively slow and cumbersome to use. The hand-held meter is much faster and easier to use, but over time the sensor membrane will be affected by airborne contaminants that reduce its ability to absorb moisture. The sensors cannot be cleaned, so if they are accidentally contaminated or simply 'get old,' they must be replaced. Stability and drift. Users should be aware that the performance of humidity meters will decline over time. The rate of change will depend on how and where they are used. User guidance. Testing to determine accuracy and calibrate the instrument requires sophisticated and expensive test equipment beyond the means of the individual user. Instruments can be returned to the manufacturer or sent to an independent testing laboratory to verify their performance. If out of spec, the sensor of an expensive instrument may be replaced. In the case of a less costly instrument, the user will probably choose to replace the entire tool. Airborne particles Sensor technologies. The particle counter uses a pump to pull a sample of air into a space called the view volume, where particles intersect with a laser beam. The particles reflect differing amounts of light based upon their size. Photo detectors "see" these light flashes and convert each one to a millivolt signal. Larger particles reflect more light and create a stronger signal. Signals within a certain millivolt range are counted in one size "bin," particles in another range are put into another bin, and so on. Technology characteristics. Accuracy can be affected when the counter is used where extremely high levels of particles are present. Particles may collect within the intake passage and measurement chamber. If these particles are dislodged during a subsequent test, they can cause a misleading spike in the particle count. Stability and drift. Laser particle counter technology is generally stable over time, but the manufacturer's recommended calibration interval should be observed to maintain optimal performance. User guidance. Operating the air pump for a period of time before taking a reading will help to flush particles out of the instrument. Users can calibrate the instrument for a zero particle count by applying a HEPA (high efficiency particle air) filter over the air intake port. Carbon dioxide Sensor technologies. CO2 sensors use a non-dispersive infrared technology. Incandescent light is projected through a small sample cell called the "bench link". The CO2 present in the test sample will absorb a specific wavelength of the projected light. A filtered infrared detector at the other end of the chamber measures the amount of light at that wavelength that passes  Fluke Corporation Making the numbers add up: Understanding specifications and performance of IAQ test instruments

  • 1
  • 2
  • 3
  • 4
  • 5

4 Fluke Corporation
Making the numbers add up: Understanding specifications and performance of IAQ test instruments
In terms of response time, both
technologies perform well. But
response time is also affected
by the overall design of the test
instrument. If the temperature
sensor is enclosed in a mas-
sive housing, the surrounding
material is liable to affect the
speed with which air heats or
cools the sensor, thus slowing
response time. Minimal response
time requires minimizing mass,
and that means shrinking com-
ponents. If carried too far, this
could affect instrument durability.
Stability and drift.
The therm-
istor temperature sensor is very
stable and does not drift like
some other sensors. However,
the temperature sensing
function
can drift over time due to other
components used and the layout
of the circuit design. In design
and manufacture, care must
be taken in the layout of the
printed circuit board that con-
nects with the thermistor. Ionic
contamination from soldering flux
residues can cause performance
degradation. This increases as
the impedance of the circuit
increases. Good analog design
is essential to achieving optimal
temperature performance.
User guidance.
Allow for
instrument ‘settling time’ when
measuring temperature. The
mass of a temperature tester may
slow response time in rapidly
changing conditions. In simple
terms, it takes a while for the
instrument to reach ambient
temperature. In addition, the
instrument’s electronic circuitry
will perform differently as tem-
peratures vary, so the circuitry
too must settle and stabilize.
Humidity
Sensor technologies.
The device
traditionally used to measure
humidity is the sling psychrom-
eter. This instrument consists of
two bulb thermometers, one of
which is surrounded by a wet
cloth. When the device is slung
through the air, moisture evapo-
rates from the wet bulb and
cools the thermometer. The dryer
the air, the greater the tempera-
ture difference between the wet
and dry thermometers.
A more convenient alternative
is a hand-held humidity meter
that uses a capacitance sensor
technology. A semi-permeable
membrane in the sensor becomes
more conductive as humidity
increases and moisture pen-
etrates the membrane. The meter
interprets this change in terms of
humidity level.
Technology characteristics.
The sling psychrometer is highly
accurate when used correctly. It
is also relatively slow and cum-
bersome to use.
The hand-held meter is much
faster and easier to use, but over
time the sensor membrane will
be affected by airborne con-
taminants that reduce its ability
to absorb moisture. The sensors
cannot be cleaned, so if they
are accidentally contaminated
or simply ‘get old,’ they must be
replaced.
Stability and drift.
Users
should be aware that the perfor-
mance of humidity meters will
decline over time. The rate of
change will depend on how and
where they are used.
User guidance.
Testing to
determine accuracy and calibrate
the instrument requires sophisti-
cated and expensive test equip-
ment beyond the means of the
individual user. Instruments can
be returned to the manufacturer
or sent to an independent testing
laboratory to verify their perfor-
mance. If out of spec, the sensor
of an expensive instrument may
be replaced. In the case of a less
costly instrument, the user will
probably choose to replace the
entire tool.
Airborne particles
Sensor technologies.
The parti-
cle counter uses a pump to pull a
sample of air into a space called
the view volume, where particles
intersect with a laser beam. The
particles reflect differing amounts
of light based upon their size.
Photo detectors “see” these
light flashes and convert each
one to a millivolt signal. Larger
particles reflect more light and
create a stronger signal. Signals
within a certain millivolt range
are counted in one size “bin,”
particles in another range are put
into another bin, and so on.
Technology characteris-
tics.
Accuracy can be affected
when the counter is used where
extremely high levels of particles
are present. Particles may collect
within the intake passage and
measurement chamber. If these
particles are dislodged during a
subsequent test, they can cause
a misleading spike in the particle
count.
Stability and drift.
Laser
particle counter technology is
generally stable over time, but
the manufacturer’s recommended
calibration interval should be
observed to maintain optimal
performance.
User guidance.
Operating the
air pump for a period of time
before taking a reading will
help to flush particles out of the
instrument. Users can calibrate
the instrument for a zero particle
count by applying a HEPA (high
efficiency particle air) filter over
the air intake port.
Carbon dioxide
Sensor technologies.
CO
2
sen-
sors use a non-dispersive infra-
red technology. Incandescent
light is projected through a small
sample cell called the “bench
link”. The CO
2
present in the test
sample will absorb a specific
wavelength of the projected
light. A filtered infrared detector
at the other end of the chamber
measures the amount of light
at that wavelength that passes