Fluke Calibration 1551A-12-DL Calibration Guide - Page 17
Adjustment at the Zero Calibration Point, Adjustment at the High Calibration Point
View all Fluke Calibration 1551A-12-DL manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 17 highlights
Stik Thermometer Calibration Adjustment at the Zero Calibration Point The next adjustment is made for the middle of the temperature range of the Product. Set the bath temperature to the Zero Adjustment Temperature given in Table 7. Make sure the temperatures of the bath, reference thermometer, and the Product are settled and stable before proceeding. Table 7. Zero Adjustment Temperature Zero Point 1551A Ex 0 °C 0 °C 1552A Ex To make an automatic adjustment at the Zero calibration point, issue the following command: CAL:USER:ZERO The parameter is the temperature of the bath as measured by the reference thermometer. The Product accepts a temperature value with up to three decimal places of resolution. The temperature value must be in the same unit in which the Product is measuring. The command will set the value of the CAL:USER:TEMP2 parameter to the given temperature and automatically adjust the CAL:USER:ADJ2 parameter to cause the Product to accurately measure the given temperature. This adjustment will not affect the measurements at the other Adjustment temperatures. Allow the readings of the Product to settle and then verify that the readings match the temperature of the reference thermometer well within the specified accuracy of the 1551A Ex/1552A Ex. Adjustment at the High Calibration Point The final adjustment is made for the high end of the temperature range of the Product. Set the bath temperature to the High Adjustment Temperature given in Table 8, depending on the model. Make sure the temperatures of the bath, reference thermometer, and Product are settled and stable before proceeding. Table 8. High Adjustment Temperature Point High 1551A Ex 157 °C 300 °C 1552A Ex To make an automatic adjustment at the High calibration point, issue the following command: CAL:USER:HIGH The parameter is the temperature of the bath as measured by the reference thermometer. The Product accepts a temperature value with up to three decimal places of resolution. The temperature value must be in the same unit in which the Product is measuring. 9