Fluke T6-600 Product Manual - Page 11
Table 7. FieldSense Low Current Tests, Table 8. FieldSense High Current Tests, Calibrator Output,
View all Fluke T6-600 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 11 highlights
T6-1000/T6-600 Service Information 6. Verify that the Product reading is within the display limits. 7. Apply the test values for the remaining steps of Table 7. Verify the readings are within the display limits. Table 7. FieldSense Low Current Tests Step Calibrator Output (AUX HI) 1 0.5 A, 45 Hz 2 0.5 A, 60 Hz 3 0.5 A, 66 Hz T6-600/T6-1000 Display Limits 4.6 A to 5.4 A 4.6 A to 5.4 A 4.6 A to 5.4 A T6-1000 only Display Limits 43 Hz to 47 Hz 58 Hz to 62 Hz 64 Hz to 68 Hz 8. Set the calibrator to STANDBY. Change the connection from the calibrator AUX HI to the calibrator AUX 20A jack. 9. Set the calibrator to the values listed in Table 8, Step 1, and apply to the 10/50-turn coil. 10. Verify that the Tester reading is within the display limits. 11. Apply the test values shown in the remaining steps of Table 8 and verify the Tester readings are within the display limits. Table 8. FieldSense High Current Tests Step 1 2 3 4 5 6 Calibrator Output 10 A, 45 Hz 10 A, 60 Hz 10 A, 66 Hz 18 A, 45 Hz 18 A, 60 Hz 18 A, 66 Hz Display Limits 96.7 A to 103.3 A 96.7 A to 103.3 A 96.7 A to 103.3 A 174.3 A to 185.7 A 174.3 A to 185.7 A 174.3 A to 185.7 A MyFlukeStore 8 Shop for Fluke products online at: www. .com 1.888.610.7664