Fluke Ti90 Manual - Page 66

Detailed Specifications

Page 66 highlights

Ti90, Ti95, Ti100, Ti105, Ti110, Ti125, TiR105, TiR110, TiR125 Users Manual Detailed Specifications Temperature Measurements Temperature Range (not calibrated below -10 °C) Ti90, Ti95, Ti100, Ti105, Ti110 20 °C to +250 °C Ti125 20 °C to +350 °C TiR105, TiR110, TiR125 20 °C to +150 °C Accuracy 2 °C or 2 % (whichever is greater) at 25 °C ambient Measurement Modes Smooth Auto-Scaling and Manual Scaling On-screen Emissivity Correction all models Imaging Performance Field of View Ti100, Ti105, Ti110, Ti125, TiR110, TiR125 ...31 ° x 22.5 ° Ti95 26 ° x 26 ° Ti90 19.5 ° x 26 ° Spatial Resolution Ti100, Ti105, Ti110, Ti125, TiR110, TiR125 (IFOV 3.39 mRad Ti90, Ti95 (IFOV 5.6 mRad Minimum Focus Distance Ti100, Ti105, TiR105 122 cm (approx. 48 in) Ti90, Ti95 46 cm (approx. 18 in) Ti110, Ti125, TiR110, TiR125 15 cm (approx. 6 in) Focus Ti90, Ti95, Ti100, Ti105, TiR105 Focus-free Ti110, Ti125, TiR110, TiR125 IR-OptiFlex™ focus Image Capture or Refresh Rate Ti90, Ti95, Ti100, TiR105, TiR110, TiR125.....9 Hz Ti105, Ti110, Ti125 9 Hz or 30 Hz (factory set) Detector Type (Focal Plane Array, uncooled microbolometer) Ti100, Ti105, Ti110, Ti125, TiR110, TiR125 ...160 X 120 Ti95 80 x 80 Ti90 80 x 60 Thermal Sensitivity (NETD) Ti90 150 mK Ti95, Ti100, Ti105, Ti110, Ti125 100 mK (0.1 °C at 30 °C target temperature) TiR105, TiR110, TiR125 80 mK (0.08 °C at 30 °C target temperature) MyFlukeStore S6ho0p for Fluke products online at: www. .com 1.877.766.5412

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68

Ti90, Ti95, Ti100, Ti105, Ti110, Ti125, TiR105, TiR110, TiR125
Users Manual
60
Detailed Specifications
Temperature Measurements
Temperature Range (not calibrated below -10
°
C)
Ti90, Ti95, Ti100, Ti105, Ti110
.......................
-20
°
C to +250
°
C
Ti125
...............................................................
-20
°
C to +350
°
C
TiR105, TiR110, TiR125
..................................
-20
°
C to +150
°
C
Accuracy
..............................................................
±
2
°
C or 2 % (whichever is greater) at
25
°
C ambient
Measurement Modes
...........................................
Smooth Auto-Scaling and Manual
Scaling
On-screen Emissivity Correction
.........................
all models
Imaging Performance
Field of View
Ti100, Ti105, Ti110, Ti125, TiR110, TiR125 ...31
°
x 22.5
°
Ti95
.................................................................
26
°
x 26
°
Ti90
.................................................................
19.5
°
x 26
°
Spatial Resolution
Ti100, Ti105, Ti110, Ti125, TiR110,
TiR125 (IFOV)
.................................................
3.39 mRad
Ti90, Ti95 (IFOV)
.............................................
5.6 mRad
Minimum Focus Distance
Ti100, Ti105, TiR105
.......................................
122 cm (approx. 48 in)
Ti90, Ti95
........................................................
46 cm (approx. 18 in)
Ti110, Ti125, TiR110, TiR125
.........................
15 cm (approx. 6 in)
Focus
Ti90, Ti95, Ti100, Ti105, TiR105
.....................
Focus-free
Ti110, Ti125, TiR110, TiR125
.........................
IR-OptiFlex™ focus
Image Capture or Refresh Rate
Ti90, Ti95, Ti100, TiR105, TiR110, TiR125
.....
9 Hz
Ti105, Ti110, Ti125
.........................................
9 Hz or 30 Hz (factory set)
Detector Type (Focal Plane Array, uncooled microbolometer)
Ti100, Ti105, Ti110, Ti125, TiR110, TiR125 ...160 X 120
Ti95
.................................................................
80 x 80
Ti90
.................................................................
80 x 60
Thermal Sensitivity (NETD)
Ti90
.................................................................
150 mK
Ti95, Ti100, Ti105, Ti110, Ti125
.....................
100 mK (0.1
°
C at 30
°
C target
temperature)
TiR105, TiR110, TiR125
..................................
80 mK (0.08
°
C at 30
°
C target
temperature)