Fluke TiR29 Fluke TiR27 TiR29 TiR32 P3 Series Thermal Imager Product Datasheet - Page 2

Detailed specifications - thermal imager

Page 2 highlights

Detailed specifications TiR32 TiR29 TiR27 Temperature Temperature measurement range (not calibrated below -10 °C) -20 °C to +150 °C (-4 °F to +302 °F) Temperature measurement accuracy ± 2 °C or 2 % (at 25 °C nominal, whichever is greater) On-screen emissivity correction Yes On-screen reflected background temperature compensation Yes On-screen transmission correction Yes Imaging performance Image capture frequency 9 Hz refresh rate or 60 Hz refresh rate depending upon model variation Detector type Focal Plane Array, uncooled microbolometer, 320 x 240 pixels Focal Plan Array, uncooled microbolometer, 280 x 210 pixels Focal Plan Array, uncooled microbolometer, 240 x 180 pixels Thermal sensitivity (NETD) ≤ 0.04 °C at 30 °C target temp. (40 mK) ≤ 0.045 °C at 30 °C target temp. (45 mK) Total pixels 76,800 58,800 43,200 Infrared spectral band 7.5 µm to 14 µm (long wave) Visual (visible light) camera Industrial performance 2.0 megapixel Minimum focus distance 45 cm (approx. 18 in) Standard infrared lens type Field of view 23 ° x 17 ° Spatial resolution (IFOV) 1.25 mRad 1.43 mRad 1.67 mRad Minimum focus distance 15 cm (approx. 6 in) Optional telephoto infrared lens type Field of view 11.5 ° x 8.7 ° Spatial resolution (IFOV) 0.63 mRad 0.72 mRad 0.84 mRad Minimum focus distance 45 cm (approx. 18 in) Optional wide-angle infrared lens type Field of view 46 ° x 34 ° Spatial resolution (IFOV) 2.50 mRad 2.86 mRad 3.34 mRad Minimum focus distance 7.5 cm (approx. 3 in) Focus mechanism Manual, one-handed Smart Focus capability Image presentation Palettes Standard Ironbow, Blue-Red, High Contrast, Amber, Amber Inverted, Hot Metal, Grayscale, Grayscale Inverted Ultra Contrast™ Ironbow Ultra, Blue-Red Ultra, High Contrast Ultra, Amber Ultra, Amber Inverted Ultra, Hot Metal Ultra, Grayscale Ultra, Grayscale Inverted Ultra Level and span Smooth auto-scaling and manual scaling of level and span Fast auto toggle between manual and auto modes Yes Fast auto-rescale in manual mode Yes Minimum span (in manual mode) 2.0 °C (3.6 °F) Minimum span (in auto mode) 3.0 °C (5.4 °F) IR-Fusion® information Automatically aligned (parallax corrected) visual and IR blending Yes Picture-In-Picture (PIP) Three levels of on-screen IR blending displayed in center of LCD Full screen infrared Three levels of on-screen IR blending displayed on LCD Color alarms (temperature alarms) Dewpoint temperature alarm (user-selectable) Voice annotation 60 seconds maximum recording time per image; reviewable playback on imager Image capture and data storage The TiR32, TiR29 and Ti27 allows the user to adjust palette, blending, level, span, IR-Fusion® mode, emissivity, and reflected background temperature compensation, and transmission correction on a captured image before it is stored. Image capture, review, save mechanism One-handed image capture, review, and save capability Storage medium SD Memory Card (2 GB memory card will store at least 1200 fully radiometric (.is2) IR and linked visual images each with 60 seconds voice annotations, or 3000 basic bitmap (.bmp) images, or 3000 jpeg (.jpeg) images; transferrable to PC via included multi-format USB card reader File formats Non-radiometric (.bmp) or (.jpeg) or fully-radiometric (.is2) No analysis software required for non-radiometric (.bmp and .jpeg) files Export file formats w/SmartView® software BMP, DIB, GIF, JPE, JFIF, JPEG, JPG, PNG, TIF, and TIFF Memory review Thumbnail view navigation and review selection 2 Fluke Corporation TiR32, TiR29 and TiR27 Thermal Imagers

  • 1
  • 2
  • 3

2
Fluke Corporation
TiR32, TiR29 and TiR27 Thermal Imagers
Detailed specifications
TiR32
TiR29
TiR27
Temperature
Temperature measurement range
(not calibrated below -10 °C)
-20 °C to +150 °C (-4 °F to +302 °F)
Temperature measurement
accuracy
± 2 °C or 2 % (at 25 °C nominal, whichever is greater)
On-screen emissivity correction
Yes
On-screen reflected background
temperature compensation
Yes
On-screen transmission correction
Yes
Imaging performance
Image capture frequency
9 Hz refresh rate or 60 Hz refresh rate depending upon model variation
Detector type
Focal Plane Array, uncooled
microbolometer, 320 x 240 pixels
Focal Plan Array, uncooled
microbolometer, 280 x 210 pixels
Focal Plan Array, uncooled
microbolometer, 240 x 180 pixels
Thermal sensitivity (NETD)
≤ 0.04 °C at 30 °C target temp. (40 mK)
≤ 0.045 °C at 30 °C target temp. (45 mK)
Total pixels
76,800
58,800
43,200
Infrared spectral band
7.5 µm to 14 µm
(long wave)
Visual (visible light) camera
Industrial performance 2.0 megapixel
Minimum focus distance
45 cm (approx. 18 in)
Standard infrared lens type
Field of view
23 ° x 17 °
Spatial resolution (IFOV)
1.25 mRad
1.43 mRad
1.67 mRad
Minimum focus distance
15 cm (approx. 6 in)
Optional telephoto infrared lens type
Field of view
11.5 ° x 8.7 °
Spatial resolution (IFOV)
0.63 mRad
0.72 mRad
0.84 mRad
Minimum focus distance
45 cm (approx. 18 in)
Optional wide-angle infrared lens type
Field of view
46 ° x 34 °
Spatial resolution (IFOV)
2.50 mRad
2.86 mRad
3.34 mRad
Minimum focus distance
7.5 cm (approx. 3 in)
Focus mechanism
Manual, one-handed Smart Focus capability
Image presentation
Palettes
Standard
Ironbow, Blue-Red, High Contrast, Amber, Amber Inverted, Hot Metal, Grayscale, Grayscale Inverted
Ultra Contrast
Ironbow Ultra, Blue-Red Ultra, High Contrast Ultra, Amber Ultra, Amber Inverted Ultra, Hot Metal Ultra,
Grayscale Ultra, Grayscale Inverted Ultra
Level and span
Smooth auto-scaling and manual scaling of level and span
Fast auto toggle between manual
and auto modes
Yes
Fast auto-rescale in manual mode
Yes
Minimum span (in manual mode)
2.0 °C (3.6 °F)
Minimum span (in auto mode)
3.0 °C (5.4 °F)
IR-Fusion
®
information
Automatically aligned (parallax
corrected) visual and IR blending
Yes
Picture-In-Picture (PIP)
Three levels of on-screen IR blending displayed in center of LCD
Full screen infrared
Three levels of on-screen IR blending displayed on LCD
Color alarms (temperature alarms)
Dewpoint temperature alarm (user-selectable)
Voice annotation
60 seconds maximum recording time per image; reviewable playback on imager
Image capture and data storage
The TiR32, TiR29 and Ti27 allows the user to adjust palette, blending, level, span, IR-Fusion
®
mode, emissivity, and reflected
background temperature compensation, and transmission correction on a captured image before it is stored.
Image capture, review, save
mechanism
One-handed image capture, review, and save capability
Storage medium
SD Memory Card (2 GB memory card will store at least 1200 fully radiometric (.is2) IR and linked visual images
each with 60 seconds voice annotations, or 3000 basic bitmap (.bmp) images, or 3000 jpeg (.jpeg) images;
transferrable to PC via included multi-format USB card reader
File formats
Non-radiometric (.bmp) or (.jpeg) or fully-radiometric (.is2)
No analysis software required for non-radiometric (.bmp and .jpeg) files
Export file formats w/SmartView
®
software
BMP, DIB, GIF, JPE, JFIF, JPEG, JPG, PNG, TIF, and TIFF
Memory review