HoMedics AHBPA-060 User Manual - Page 15

Troubleshooting, Specifications

Page 15 highlights

TROUBLESHOOTING If any abnormality arises during use, please check the following points SYMPTOMS POSSIBLE CAUSES CORRECTION Unit does not turn on when the START/STOP button is pushed. Batteries have run down. Battery polarities have been positioned incorrectly. Replace them with four new AAA alkaline batteries. Re-insert the batteries in the correct positions. EE measurement error Cuff has been placed Wrap the cuff properly symbol shown on dis- incorrectly. so that it is positioned play or the blood pres- correctly. sure value is displayed excessively low (or high). Did you talk or move Measure again. Keep during measurement? arm steady during Shaking of the arm with measurement. the cuff on. E1 error symbol shown Air circuit abnormality. on display Cuff tube may not be plugged into monitor correctly. Check cuff connection. Measure again. E2 error symbol shown Inflation pressure Switch the unit off, then on display exceeding 300 mmHg. measure agaain. E3 error symbol shown Error determining on display measurement data. Measure again. CA Appears on display Calibration alert. Contact HoMedics Consumer Relations at 1-800-466-3342. Note: If the unit still does not work, contact HoMedics Consumer Relations. Under no circumstance should you disassemble or attempt to repair the unit by yourself. Contact information for HoMedics Consumer Relations Department can be found on the warranty page. 27 SPECIFICATIONS Power Source: Measurement Method: Four 1.5V DC (AAA) batteries or optional 6V DC AC adapter (not included) Oscillometric Measurement Range: Pressure: 40~280 millimeters mercury (mmHg) Pulse: 40~199 beats/minute Accuracy: Pressure: ±3 mmHg Pulse: ±5% of reading Pressure Sensor: Semi-conductor Inflation: Deflation: Memory Capacity: Pump driven Automatic Pressure release valve 50 memories for each User (100 total) Auto-shut-off: 1 minute after last button operation Operation Environment: Temperature: 50°F~104°F (10°C ~40°C) Humidity: Less Than 85% RH Storage Environment: Temperature: -4°F~140°F Humidity: Less Than 85% (-20°C~60°C) RH Weight: 0.65 lb (295 g) (without batteries) Arm Circumference: Standard size cuff: 9"-13" (23-33 cm) Large size cuff:13"-17" (33-43 cm) Dimensions: 5.6"(L) x 3.7"(W) x 2.8"(H) 142mm(L) x 95mm(W) x 71mm(H) Type BF Device and cuff are designed to provide special protection against electrical shocks. Note: These specifications are subject to change without notice. 28

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30

28
27
TROUBLESHOOTING
If any abnormality arises during use, please check the following points
Note:
If the unit still does not work, contact HoMedics Consumer Relations.
Under no circumstance should you disassemble or attempt to repair
the unit by yourself. Contact information for HoMedics Consumer
Relations Department can be found on the warranty page.
SYMPTOMS
POSSIBLE
CAUSES
CORRECTION
Unit does not turn on
when the
START/STOP
button
is pushed.
Batteries have
run down.
Replace them with
four new AAA
alkaline batteries.
Battery polarities
have been positioned
incorrectly.
Re-insert the batteries
in the correct positions.
EE measurement error
symbol shown on dis-
play or the blood pres-
sure value is displayed
excessively low
(or high).
Cuff has been placed
incorrectly.
Wrap the cuff properly
so that it is positioned
correctly.
Did you talk or move
during measurement?
Measure again. Keep
arm steady during
measurement.
Shaking of the arm with
the cuff on.
E1 error symbol shown
on display
Air circuit abnormality.
Cuff tube may not be
plugged into monitor
correctly.
Check cuff connection.
Measure again.
E2 error symbol shown
on display
Inflation pressure
exceeding 300 mmHg.
Switch the unit off, then
measure agaain.
E3 error symbol shown
on display
Error determining
measurement data.
Measure again.
CA Appears on
display
Calibration alert.
Contact HoMedics
Consumer Relations
at 1-800-466-3342.
Power Source:
Four 1.5V DC (AAA) batteries or
optional 6V DC AC adapter (not included)
Measurement
Method:
Oscillometric
Measurement Range:
Pressure: 40~280 millimeters mercury (mmHg)
Pulse: 40~199 beats/minute
Accuracy:
Pressure: ±3 mmHg
Pulse: ±5% of reading
Pressure Sensor:
Semi-conductor
Inflation:
Pump driven
Deflation:
Automatic Pressure release valve
Memory Capacity:
50 memories for each User (100 total)
Auto-shut-off:
1 minute after last button operation
Operation
Environment:
Temperature: 50°F~104°F (10°C ~40°C)
Humidity: Less Than 85% RH
Storage Environment:
Temperature: -4°F~140°F (-20°C~60°C)
Humidity: Less Than 85% RH
Weight:
0.65 lb (295 g) (without batteries)
Arm Circumference:
Standard size cuff: 9"–13" (23-33 cm)
Large size cuff:13"–17" (33-43 cm)
Dimensions:
5.6"(L) x 3.7"(W) x 2.8"(H)
142mm(L) x 95mm(W) x 71mm(H)
Type BF
Device and cuff are designed to provide
special protection against electrical shocks.
SPECIFICATIONS
Note:
These specifications are subject to change without notice.