Invacare IRC5LX Owners Manual - Page 27
Troubleshooting Guide, Symptom, Probable Cause, Solution - 5 model
View all Invacare IRC5LX manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 27 highlights
SECTION 7-TROUBLESHOOTING GUIDE SYMPTOM: PROBABLE CAUSE: SOLUTION: YELLOW or RED 1. Low oxygen purity.* 1. Clean or Replace filters. LIGHT Illuminated. 2. Kinked or blocked tub- 2. Inspect for kinks or block- ing, cannula or humidifier.* ages. Correct, clean or replace Alarm: item. Once corrected, turn CONTINUOUS power OFF for 60 seconds and On red light only. then turn power back ON. * Only applies to 5LXO2 and 10LXO2 SensO2 models. 3. Flowmeter set too low: 3. Check flowmeter is set to 0.5 L/min (or less) for 1.0 L/min. or more. Platinum 5; 1.0 L/min (or less) for Platinum 10. 4. Unit overheating due to 4.a. Remove and clean cabinet blocked air intake. filters. 4b. Move concentrator at least three inches from walls, draperies, and furniture. 5. Insufficient power at outlet. 5. DO NOT use extension cords. Move to another electrical outlet or circuit. 6. Internal repairs required. 6. Call Service Provider. GREEN LIGHT with YELLOW Light Flashing. 1. Internal repairs required. 1. Call Service Provider. Part No 1118353 27 Platinum™ Series