Invacare IRC5LXO2AW Owners Manual - Page 21
Troubleshooting
View all Invacare IRC5LXO2AW manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 21 highlights
TROUBLESHOOTING SYMPTOM: YELLOW or RED LIGHT Illuminated. Alarm: CONTINUOUS On red light only. * Only applies to SensO2 models. PROBABLE CAUSE: 1. Low oxygen purity.* 2. Kinked or blocked tubing, cannula or humidifier.* 3. Flowmeter set at 1/2 L/min. or less.* 4. Unit overheating due to blocked air intake. 5. Insufficient power at outlet. 6. Internal repairs required. SOLUTION: 1. Clean or replace filters. 2. Inspect for kinks or blockages. Correct, clean or replace item. Once corrected, turn power OFF for 60 seconds and then turn power back ON. 3. Check flowmeter is set to 1.0 L/min. or more. Refer to Recommended End-User Guidelines in the Operating Instructions of this manual. 4a. Remove and clean cabinet filters. 4b. Move concentrator at least 7.5 cm (3 in) from walls, draperies, and furniture. 5. Do not use extension cords. Move to another electrical outlet or circuit. 6. Call Service Provider. GREEN LIGHT with YELLOW Light Flashing. Alarm: RAPID Beep....Beep... Beep...Beep 1. Internal repairs required. 1. Call Service Provider. LOW FLOW ALARM 1a. Kinked or blocked tubing, cannula or humidifier. 1b. Flowmeter set at 0.5 L/min. or less. 1a. Inspect for kinks or blockages. Correct, clean or replace item. Once corrected, turn power OFF for 60 seconds and then turn power back ON. 1b. Reset flowmeter to prescribed flowrate. Flowrates less than 1L/min. are not recommended. 21