Invacare IRC5P Owners Manual - Page 27

Troubleshooting Guide, Symptom, Probable Cause, Solution - irc5po2

Page 27 highlights

SECTION 7-TROUBLESHOOTING GUIDE SYMPTOM: YELLOW or RED light Illuminated. Alarm: Continuous On RED light only. * Only applies to IRC5PO2. PROBABLE CAUSE: 1. Low oxygen purity.* 2. Kinked or blocked tubing, cannula or humidifier.* 3. Flowmeter set at 0.5 L/min or less.* 4. Unit overheating due to blocked air intake. 5. Insufficient power at outlet. 6. Internal repairs required. SOLUTION: 1. Clean or Replace filters. 2. Inspect for kinks or blockages. Correct, clean or replace item. Once corrected, turn power Off for 60 seconds and then turn power back On. 3. Check flowmeter is set to 1.0 L/min or more. 4a. Remove and clean cabinet filters. 4b. Move concentrator at least three inches from walls, draperies, and furniture. 5. DO NOT use extension cords. Move to another electrical outlet or circuit. 6. Call Service Provider. GREEN light with 1. Internal repairs YELLOW Light required. Flashing. 1. Call Service Provider. Part No 1143482 27 Perfecto2™Series

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32

SECTION 7—TROUBLESHOOTING GUIDE
Part No 1143482
27
Perfecto
2
â„¢
Series
YELLOW or RED
light Illuminated.
Alarm:
Continuous
On RED light
only.
* Only applies to
IRC5PO2.
1. Low oxygen purity.*
1. Clean or Replace filters.
2. Kinked or blocked
tubing, cannula or
humidifier.*
2. Inspect for kinks or
blockages. Correct, clean or
replace item. Once
corrected, turn power Off for
60 seconds and then turn
power back On.
3. Flowmeter set at 0.5
L/min or less.*
3. Check flowmeter is set to
1.0 L/min or more.
4. Unit overheating due
to blocked air intake.
4a. Remove and clean cabinet
filters.
4b. Move concentrator at
least three inches from walls,
draperies, and furniture.
5. Insufficient power at
outlet.
5. DO NOT use extension
cords. Move to another
electrical outlet or circuit.
6. Internal repairs
required.
6. Call Service Provider.
GREEN light with
YELLOW Light
Flashing.
1. Internal repairs
required.
1. Call Service Provider.
SYMPTOM:
PROBABLE CAUSE:
SOLUTION: