Invacare MA65 Owners Manual - Page 30
Troubleshooting
View all Invacare MA65 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 30 highlights
microAIR® MA65 Series 7 Troubleshooting 7.1 Troubleshooting PROBLEM Mattress not inflating Not alternating properly No power 30 CAUSE SOLUTION Mattress hose disconnected Connect hose connectors and lock them in place Air hose kinked or split Unkink hose or replace split hose Major leak in the air cushions or overlay Replace leaking air cushions or overlay pad pad Kinked or split manifold Unkink manifold or replace split manifold Has power and fuse is good, control unit Send control unit back to the factory for does not come on repair Not alternating, solenoid malfunction Send control unit back to the factory for repair No air, pump malfunction Send unit for repair Control unit off Check power source and turn on unit Power cord disconnected Connect power cord to the power source No power in the power source Check power source has power and turn it On Power outage Wait till the power source has power Blown fuse Replace blown fuse with an equivalent fuse 1148137-H