microAIR® MA80 Series and MA85 (with Alternating Pressure) Series Fowler Transmitter Battery Replacement (if equipped) The transmitter has an LED light to indicate the status of the battery. Green indicates good voltage and red indicates low or no voltage. Replace fowler if red indicator is present. Batteries may be replaced on fowlers where batteries are accessible. Sealed fowlers will need to be replaced as a whole part. • Battery must be 3.6V lithium AA battery. • When battery is first inserted the LED light will flash four times to indicate a 45° fowler. • Replace transmitter in mattress cover pocket in the orientation indicated on the transmitter. • Dispose of old battery according to local laws and statutes. 5.1.8 Max Inflate 1. Press the Max Inflate button H to select the max inflate mode and inflate the mattress rapidly to maximum firmness (pressurized to 35 ± 5 mmHg). A series of beeps sound every three minutes as a reminder that the Max Inflate mode is active. After 30 minutes, the Max Inflate mode deactivates and the control unit defaults to the previous setting. It takes 25-60 seconds for the mattress to inflate fully (inflation time depends on size of mattress). It is recommended that Max Inflate setting be used during patient ingress/egress, patient wound care, patient turning or patient cleaning. 2. Press the Max Inflate button again to manually disengage Max Inflate mode. This will deactivate the Max Inflate LED. 24 1148138-I