Cleaning Battery Terminals
View all Invacare TDXSI-2 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 90 highlights
7 SETUP/MAINTENANCE Cleaning Battery Terminals ƽ WARNING Most batteries are not sold with instructions. However, warnings are frequently noted on the cell caps. Read them carefully. DO NOT allow the liquid in the battery to come in contact with skin, clothes or other possessions. It is a form of acid and harmful or damaging burns may result. Should the liquid touch your skin, wash the area IMMEDIATELY and thoroughly with cool water. In serious cases or if eye contact is made, seek medical attention IMMEDIATELY. 1. Examine battery terminals for corrosion. 2. Verify the plastic caps are in place over battery cell holes. 3. Clean terminals by using a battery cleaning tool, wire brush, or medium grade sand paper. Upon completion, areas should be shiny, not dull. 4. Carefully dust off all metal particles. Invacare® TDX®SI Series 90 Part No 1154294