NEC C751Q External Controls - Page 55

Asset Data write

Page 55 highlights

8.2 Asset Data write This command is used in order to write Asset Data. 1) The controller requests the monitor to write Asset data. Header Message SOH-'0'-Monitor ID'0'-'A'-N-N STX-'C'-'0'-'0'-'E'-'0'-'0'Data(0)-Data(1)---Data(N)-ETX Check code BCC Delimiter CR Header SOH (01h): Start of Header '0' (30h): Reserved Monitor ID: Specify the Monitor ID in which you want to write data. Ex.) If Monitor ID is '1', specify 'A'. '0' (30h): Message sender is the controller. 'A' (41h): Message type is "Command". N-N: Message length Note.) The maximum data length that can be written to the monitor at a time is 32bytes. Message STX (02h): Start of Message 'C'-'0'-'0'-'E' (43h, 30h, 30h, 45h): Asset Data writes command '0'-'0'(30h, 30h): Offset address from top of Asset data. 00h : Write data from top of the Asset data area. Data(0) -- Data(N): Asset data. The data must be ASCII characters strings. ETX (03h): End of Message Check code BCC: Block Check Code Refer to the section 4.3 "Check code" for a BCC calculation. Delimiter CR (0Dh): End of packet 2) The monitor replies a data for confirmation. Header Message SOH-'0'-'0'-Monitor ID'B'-N-N STX-'0'-'0'-'C'-'0'-'0'-'E'-'0'-'0'Data(0)-Data(1)---Data(N)-ETX Check code BCC Delimiter CR Header SOH (01h): Start of Header '0' (30h): Reserved '0' (30h): Message receiver is the controller. Monitor ID: Indicate a replying Monitor ID. Ex.) When this byte is set to 'A', the replying Monitor ID is '1'. 'B' (42h): Message type is "Command reply". N-N: Message length Note.) The maximum data length that can be written to the monitor at a time is 32bytes. Message STX (02h): Start of Message '0'-'0': Result code. No error. 'C'-'0'-'0'-'E' (43h, 30h, 30h, 45h): Asset Data write command '0'-'0'(30h, 30h): Offset address from top of Asset data. 00h : Write data into from top of the Asset data area. Data(0) -- Data(N): Asset data. The data must be ASCII characters strings. ETX (03h): End of Message Check code BCC: Block Check Code Refer to the section 4.3 "Check code" for a BCC calculation. (55/145)

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122
  • 123
  • 124
  • 125
  • 126
  • 127
  • 128
  • 129
  • 130
  • 131
  • 132
  • 133
  • 134
  • 135
  • 136
  • 137
  • 138
  • 139
  • 140
  • 141
  • 142
  • 143
  • 144
  • 145

(55/145)
8.2 Asset Data write
This command is used in order to write Asset Data.
1) The controller requests the monitor to write Asset data.
Header
Message
Check
code
Delimiter
SOH-'0'-Monitor ID-
'0'-'A'-N-N
STX-'C'-'0'-'0'-'E'-'0'-'0'-
Data(0)-Data(1)---Data(N)-ETX
BCC
CR
Header
SOH (01h): Start of Header
'0' (30h): Reserved
Monitor ID: Specify the Monitor ID in which you want to write data.
Ex.) If Monitor ID is '1', specify 'A'.
'0' (30h): Message sender is the controller.
'A' (41h): Message type is "Command".
N-N: Message length
Note.) The maximum data length that can be written to the monitor at a time is
32bytes.
Message
STX (02h): Start of Message
'C'-'0'-'0'-'E' (43h, 30h, 30h, 45h): Asset Data writes command
'0'-'0'(30h, 30h): Offset address from top of Asset data.
00h : Write data from top of the Asset data area.
Data(0)
- Data(N): Asset data. The data must be ASCII characters strings.
ETX (03h): End of Message
Check code
BCC: Block Check Code
Refer to the section 4.3
Check code
for a BCC calculation.
Delimiter
CR (0Dh): End of packet
2) The monitor replies a data for confirmation.
Header
Message
Check
code
Delimiter
SOH-'0'-'0'-Monitor ID-
'B'-N-N
STX-'0'-'0'-'C'-'0'-'0'-'E'-'0'-'0'-
Data(0)-Data(1)---Data(N)-ETX
BCC
CR
Header
SOH (01h): Start of Header
'0' (30h): Reserved
'0' (30h): Message receiver is the controller.
Monitor ID: Indicate a replying Monitor ID.
Ex.) When this byte is set to 'A', the replying Monitor ID is '1'.
'B' (42h): Message type is "Command reply".
N-N: Message length
Note.) The maximum data length that can be written to the monitor at a time is
32bytes.
Message
STX (02h): Start of Message
'0'-'0': Result code. No error.
'C'-'0'-'0'-'E' (43h, 30h, 30h, 45h): Asset Data write command
'0'-'0'(30h, 30h): Offset address from top of Asset data.
00h : Write data into from top of the Asset data area.
Data(0)
- Data(N): Asset data. The data must be ASCII characters strings.
ETX (03h): End of Message
Check code
BCC: Block Check Code
Refer to the section 4.3
Check code
for a BCC calculation.